Comparing Total Keratometry Measurement on the IOLMaster 700 With Goggin Nomogram Adjusted Anterior Keratometry
View through CrossRef
PURPOSE:
To compare “total keratometry” measurements with Goggin nomogram adjusted anterior keratometry (GNAK) for toric intraocular lens calculation.
METHODS:
Routine biometry provided measured total keratometry values from which astigmatism was derived. The Goggin nomogram was applied to anterior keratometry values on the same eyes to provide nomogram-adjusted anterior keratometry values (GNAK) that estimate total corneal astigmatism. The agreement between total keratometry and GNAK was analyzed.
RESULTS:
Overall, in 46 eyes there was no statistically significant difference between median GNAK and total keratometry power values (
P
= .746). No statistically significant difference remained in against-the-rule and oblique subgroup analyses. Absolute and signed steep axis of astigmatism was statistically significantly different for GNAK and total keratometry in the overall analysis (
P
< .001 and = .029, respectively) and for against-the-rule and oblique subgroup analyses. The with-the-rule subgroup showed a statistically significant difference in astigmatic power and no significant signed steep axis difference between GNAK and total keratometry.
CONCLUSIONS:
Total keratometry appears able to measure total corneal astigmatism to match closely (clinically and statistically) GNAK estimation of that parameter. This indicates that it would be safe and reasonable to use total keratometry data for planning of cataract surgery with toric IOLs.
[
J Refract Surg.
2018;34(8):521–526.]
Title: Comparing Total Keratometry Measurement on the IOLMaster 700 With Goggin Nomogram Adjusted Anterior Keratometry
Description:
PURPOSE:
To compare “total keratometry” measurements with Goggin nomogram adjusted anterior keratometry (GNAK) for toric intraocular lens calculation.
METHODS:
Routine biometry provided measured total keratometry values from which astigmatism was derived.
The Goggin nomogram was applied to anterior keratometry values on the same eyes to provide nomogram-adjusted anterior keratometry values (GNAK) that estimate total corneal astigmatism.
The agreement between total keratometry and GNAK was analyzed.
RESULTS:
Overall, in 46 eyes there was no statistically significant difference between median GNAK and total keratometry power values (
P
= .
746).
No statistically significant difference remained in against-the-rule and oblique subgroup analyses.
Absolute and signed steep axis of astigmatism was statistically significantly different for GNAK and total keratometry in the overall analysis (
P
< .
001 and = .
029, respectively) and for against-the-rule and oblique subgroup analyses.
The with-the-rule subgroup showed a statistically significant difference in astigmatic power and no significant signed steep axis difference between GNAK and total keratometry.
CONCLUSIONS:
Total keratometry appears able to measure total corneal astigmatism to match closely (clinically and statistically) GNAK estimation of that parameter.
This indicates that it would be safe and reasonable to use total keratometry data for planning of cataract surgery with toric IOLs.
[
J Refract Surg.
2018;34(8):521–526.
].
Related Results
Accuracy of the Goggin Nomogram for 0.50 D Steps in Toric IOL Cylinder Power Calculation
Accuracy of the Goggin Nomogram for 0.50 D Steps in Toric IOL Cylinder Power Calculation
PURPOSE:
To examine the accuracy of the cylinder power choice for toric intraocular lenses (IOLs) using the Goggin Nomogram, which adjusts anterior keratometric astigma...
Comparison of four formulas for the intraocular lens power prediction
Comparison of four formulas for the intraocular lens power prediction
Abstract
Purpose
To assess the accuracy of four formulas for intraocular lens (IOL) power prediction in cataractous eyes.
METHODS
In this prospective study, 51 eyes of 3...
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length).
Methods: This single-...
Predischarge Screening of Neonates with Transcutaneous Bilirubinometer
Predischarge Screening of Neonates with Transcutaneous Bilirubinometer
Objective:
To compare the diagnostic accuracy of various transcutaneous bilirubin (TcB) nomograms for predischarge screening.
...
A specific nomogram to predict subsequent brain metastasis in metastatic triple-negative breast cancer patients
A specific nomogram to predict subsequent brain metastasis in metastatic triple-negative breast cancer patients
1028 Background: The development of brain metastases is usually a fatal event in the history of breast cancer patients with metastatic disease other than brain (MDOB). We previous...
Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract
Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract
AIM: To assess and compare the variations and agreements across different ocular biometric parameters using swept-source optical coherence tomography (SS-OCT) and Scheimpflug tomog...
Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
Background and Objectives: To compare the biometry of eyes obtained with two swept-source optical coherence tomography-based biometers—Argos (A), using an individual refractive ind...