Accuracy of the Goggin Nomogram for 0.50 D Steps in Toric IOL Cylinder Power Calculation

View through CrossRef
PURPOSE: To examine the accuracy of the cylinder power choice for toric intraocular lenses (IOLs) using the Goggin Nomogram, which adjusts anterior keratometric astigmatic power values for the likely effect of posterior corneal and non-corneal, non-lenticular astigmatism. METHODS: A consecutive, retrospective case series was based at the Queen Elizabeth Hospital and Ashford Advanced Eye Care in Adelaide, Australia. A total of 586 consecutive eyes of 586 patients underwent phacoemulsification surgery with implantation of a Zeiss AT TORBI 709MP or AT LISA Tri Toric 939 MP toric IOL, calculated using the Goggin Nomogram. The median absolute magnitude of error and geometric mean astigmatic correction index in consecutive eyes with toric IOL cylinder powers of 1.00 to 3.00 diopters (D) were analyzed. RESULTS: Overall, all eyes receiving IOL cylinder powers of 1.00 to 3.00 D inclusive had a median magnitude of error value of 0.19 D (IQR: 0.31) and astigmatic correction index value of 1.03 (IQR: 0.33). For eyes with with-the-rule, against-the-rule, and oblique astigmatism, the median magnitude of error was 0.18 D (interquartile range [IQR]: 0.29), 0.19 D (IQR: 0.31), and 0.17 D (IQR: 0.39), respectively, and the astigmatic correction index was 1.06 (IQR: 0.28), 1.01 (IQR: 0.35) and 1.08 (IQR: 0.32), respectively. CONCLUSIONS: Goggin Nomogram adjusted keratometry provided optimal refractive astigmatic outcome in IOL cylinder powers of 1.00 to 3.00 D in eyes with with-the-rule, against-the-rule, and oblique astigmatism. Goggin Nomogram adjusted keratometry compensates for both posterior corneal astigmatism and any other source of ocular astigmatism. [ J Refract Surg . 2022;38(5):298–303.]
Title: Accuracy of the Goggin Nomogram for 0.50 D Steps in Toric IOL Cylinder Power Calculation
Description:
PURPOSE: To examine the accuracy of the cylinder power choice for toric intraocular lenses (IOLs) using the Goggin Nomogram, which adjusts anterior keratometric astigmatic power values for the likely effect of posterior corneal and non-corneal, non-lenticular astigmatism.
METHODS: A consecutive, retrospective case series was based at the Queen Elizabeth Hospital and Ashford Advanced Eye Care in Adelaide, Australia.
A total of 586 consecutive eyes of 586 patients underwent phacoemulsification surgery with implantation of a Zeiss AT TORBI 709MP or AT LISA Tri Toric 939 MP toric IOL, calculated using the Goggin Nomogram.
The median absolute magnitude of error and geometric mean astigmatic correction index in consecutive eyes with toric IOL cylinder powers of 1.
00 to 3.
00 diopters (D) were analyzed.
RESULTS: Overall, all eyes receiving IOL cylinder powers of 1.
00 to 3.
00 D inclusive had a median magnitude of error value of 0.
19 D (IQR: 0.
31) and astigmatic correction index value of 1.
03 (IQR: 0.
33).
For eyes with with-the-rule, against-the-rule, and oblique astigmatism, the median magnitude of error was 0.
18 D (interquartile range [IQR]: 0.
29), 0.
19 D (IQR: 0.
31), and 0.
17 D (IQR: 0.
39), respectively, and the astigmatic correction index was 1.
06 (IQR: 0.
28), 1.
01 (IQR: 0.
35) and 1.
08 (IQR: 0.
32), respectively.
CONCLUSIONS: Goggin Nomogram adjusted keratometry provided optimal refractive astigmatic outcome in IOL cylinder powers of 1.
00 to 3.
00 D in eyes with with-the-rule, against-the-rule, and oblique astigmatism.
Goggin Nomogram adjusted keratometry compensates for both posterior corneal astigmatism and any other source of ocular astigmatism.
[ J Refract Surg .
2022;38(5):298–303.
].

Related Results

Rotational stability of modified toric intraocular lens
Rotational stability of modified toric intraocular lens
We evaluated the rotational stability of a new toric intraocular lens (IOL), HOYA XY-1 toric IOL that is an improved version of HOYA 355 toric IOL, with longer overall length (13.0...
Intraocular lens tilt and decentration in congenital ectopia lentis: baseline characteristics and first-year report
Intraocular lens tilt and decentration in congenital ectopia lentis: baseline characteristics and first-year report
AIM: To investigate the characteristics and associated factors of intraocular lens (IOL) tilt and decentration after transscleral suture-fixated IOL surgery in congenital ectopia l...
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length). Methods: This single-...
Iron Overload after Hematopoietic Stem Cell Transplantation
Iron Overload after Hematopoietic Stem Cell Transplantation
Abstract Introduction Iron overload (IOL) is a common complication after HSCT, mainly due to iterative red blood cell (RBC) transfusions with other mechanisms as ine...
Comparative estimation of toric intraocular lenses optic power calculation acurary
Comparative estimation of toric intraocular lenses optic power calculation acurary
Purpose. To assess the refractive error of toric IOL calculation using Holladay IOL Consultant program with Holladay II formula, Verion navigation system using Holladay II and Barr...
Refractive Outcomes of Toric and Non-Toric IOLs in Post-Radial Keratotomy Eyes: A Comparative Study
Refractive Outcomes of Toric and Non-Toric IOLs in Post-Radial Keratotomy Eyes: A Comparative Study
Abstract Purpose To compare refractive outcomes in post-radial keratotomy (RK) eyes undergoing cataract surgery with toric vers...
Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
Biometry and Intraocular Lens Power Calculation by Combined Scheimpflug-Placido Disc versus Optical Interferometry Devices
Purpose: To compare the results of the current gold standard, laser interferometry, and keratometry by the IOL-Master, with a newly developed Galilei G6 using raytracing software O...

Back to Top