Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL

View through CrossRef
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length). Methods: This single-center retrospective chart review analyzed the charts of 62 patients (n=93 eyes) who underwent cataract surgery with a toric IOL between January and April 2022. For all patients, preoperative biometry was measured using both the IOLMaster 700 and Pentacam AXL, followed by IOL calculations performed on the Barrett Toric Online Calculator. The suggested IOL power and toric were defined as spherical equivalent power and toric power closest to plano and minimal residual astigmatism, respectively. Surgeons used lens suggestions as per IOLMaster 700 measurements, with Pentacam AXL being used as a confirmatory test. Patients with a history of laser refractive surgery, corneal disease, or in whom cataract density precluded measurements with either device were excluded. The outcome measures compared between devices were power, toric, and alignment suggestions. Results: Power suggestions were within ± 0.5D in 94% of the eyes, and were identical in 54%. Toric suggestions differed by ± 1 toric step in 100% of eyes, and were identical in 67%. IOL alignment suggestions, however, were slightly more variable and differed by ± 5º in 38-55% of eyes. At one-month post-op, the mean absolute error in spherical equivalent was nearly identical between devices. Conclusion: Our results show that IOLMaster 700 and Pentacam AXL show generally similar TORIC calculations with respect to spherical equivalent and suggested toric power. Small differences in alignment axis are frequent between the two devices and warrant further study.
ASEAN Federation for Psychiatry and Mental Health
Title: Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Description:
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length).
Methods: This single-center retrospective chart review analyzed the charts of 62 patients (n=93 eyes) who underwent cataract surgery with a toric IOL between January and April 2022.
For all patients, preoperative biometry was measured using both the IOLMaster 700 and Pentacam AXL, followed by IOL calculations performed on the Barrett Toric Online Calculator.
The suggested IOL power and toric were defined as spherical equivalent power and toric power closest to plano and minimal residual astigmatism, respectively.
Surgeons used lens suggestions as per IOLMaster 700 measurements, with Pentacam AXL being used as a confirmatory test.
Patients with a history of laser refractive surgery, corneal disease, or in whom cataract density precluded measurements with either device were excluded.
The outcome measures compared between devices were power, toric, and alignment suggestions.
Results: Power suggestions were within ± 0.
5D in 94% of the eyes, and were identical in 54%.
Toric suggestions differed by ± 1 toric step in 100% of eyes, and were identical in 67%.
IOL alignment suggestions, however, were slightly more variable and differed by ± 5º in 38-55% of eyes.
At one-month post-op, the mean absolute error in spherical equivalent was nearly identical between devices.
Conclusion: Our results show that IOLMaster 700 and Pentacam AXL show generally similar TORIC calculations with respect to spherical equivalent and suggested toric power.
Small differences in alignment axis are frequent between the two devices and warrant further study.

Related Results

Abstract 406: Mechanism of AXL regulation in chemotherapy resistance
Abstract 406: Mechanism of AXL regulation in chemotherapy resistance
Abstract Constitutive activation of receptor tyrosine kinase AXL is an important phenomenon in chemotherapy resistance and is highly expressed in several malignancie...
Rotational stability of modified toric intraocular lens
Rotational stability of modified toric intraocular lens
We evaluated the rotational stability of a new toric intraocular lens (IOL), HOYA XY-1 toric IOL that is an improved version of HOYA 355 toric IOL, with longer overall length (13.0...
Development of an Axl/Mer Dual Inhibitor, ONO-9330547: Promising Single Agent Activity in an Acute Myeloid Leukemia (AML) Model
Development of an Axl/Mer Dual Inhibitor, ONO-9330547: Promising Single Agent Activity in an Acute Myeloid Leukemia (AML) Model
Abstract Purpose: Axl and Mer belong to the TAM (Tyro3, Axl, Mer) family of receptor tyrosine kinases and play a role in regulating cell proliferation, survival, mig...
Axl, A Receptor Tyrosine Kinase, Mediates Flow-Induced Vascular Remodeling
Axl, A Receptor Tyrosine Kinase, Mediates Flow-Induced Vascular Remodeling
Intima-media thickening (IMT) in response to hemodynamic stress is a physiological process that requires coordinated signaling among endothelial, inflammatory, and vascular smooth ...
Abstract 1572: Overexpression of Axl delays metastatic outgrowth of prostate cancer bone marrow disseminated tumor cells
Abstract 1572: Overexpression of Axl delays metastatic outgrowth of prostate cancer bone marrow disseminated tumor cells
Abstract The presence of detectable metastases in the bone marrow of prostate cancer patients represents lethal, incurable disease. These metastases that may emerge ...
Refractive Outcomes of Toric and Non-Toric IOLs in Post-Radial Keratotomy Eyes: A Comparative Study
Refractive Outcomes of Toric and Non-Toric IOLs in Post-Radial Keratotomy Eyes: A Comparative Study
Abstract Purpose To compare refractive outcomes in post-radial keratotomy (RK) eyes undergoing cataract surgery with toric vers...
Comparison of four formulas for the intraocular lens power prediction
Comparison of four formulas for the intraocular lens power prediction
Abstract Purpose To assess the accuracy of four formulas for intraocular lens (IOL) power prediction in cataractous eyes. METHODS In this prospective study, 51 eyes of 3...