Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
View through CrossRef
Background and Objectives: To compare the biometry of eyes obtained with two swept-source optical coherence tomography-based biometers—Argos (A), using an individual refractive index, and IOLMaster 700 (IM), using an equivalent refractive index—for all structures. Materials and Methods: The biometry of 105 eyes of 105 patients before cataracts were analyzed in this study. Parameters such as axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) were compared from both devices. According to the axial length measurements, patients were divided into three groups, as follows: group 1—short eyes (AL < 22.5 mm), group 2—average eyes (22.5 ≤ AL ≤ 26.0 mm), and group 3—long eyes (AL > 26.0 mm). Results: The correlation coefficiency among all compared parameters varies from R = 0.92 to R = 1.00, indicating excellent reliability of IM and A. A statistical significance in axial length was indicated in the group of short eyes (n = 26)—mean AL (A) 21.90 mm (±0.59 mm) vs. AL (IM) 21.8 mm ± (0.61 mm) (p < 0.001)—and in the group of long eyes (n = 5)—mean AL (A) 27.95 mm (±2.62 mm) vs. mean AL (IM) 28.10 mm (±2.64) (p < 0.05). In the group of average eyes (n = 74), outcomes were similar—mean AL (A) 23.56 mm (±0.70 mm) vs. mean AL (IM) 23,56 mm (±0.71 mm) (p > 0.05). The anterior chamber depth measurements were higher when obtained with Argos than with IOLMaster 700—mean ACD (A) 3.06 mm (±0.48 mm) vs. mean ACD (IM) 2.92 mm (±0.46) p < 0.001. There was no statistical significance in mean LT—mean LT (A) 4.75 mm (±0.46 mm) vs. mean LT (IM) 4.72 mm (±0.44 mm) (p = 0.054). The biometry of one eye with dense cataracts could be measured only with Argos, using the Enhanced Retinal Visualization mode. Conclusions: Axial length measurements from both devices were different in the groups of short and long eyes, but were comparable in the group of average eyes. The anterior chamber depth values obtained with Argos were higher than the measurements acquired with IOLMaster 700. These differences may be particularly important when selecting IOLs for patients with extreme AL values.
Title: Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
Description:
Background and Objectives: To compare the biometry of eyes obtained with two swept-source optical coherence tomography-based biometers—Argos (A), using an individual refractive index, and IOLMaster 700 (IM), using an equivalent refractive index—for all structures.
Materials and Methods: The biometry of 105 eyes of 105 patients before cataracts were analyzed in this study.
Parameters such as axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) were compared from both devices.
According to the axial length measurements, patients were divided into three groups, as follows: group 1—short eyes (AL < 22.
5 mm), group 2—average eyes (22.
5 ≤ AL ≤ 26.
0 mm), and group 3—long eyes (AL > 26.
0 mm).
Results: The correlation coefficiency among all compared parameters varies from R = 0.
92 to R = 1.
00, indicating excellent reliability of IM and A.
A statistical significance in axial length was indicated in the group of short eyes (n = 26)—mean AL (A) 21.
90 mm (±0.
59 mm) vs.
AL (IM) 21.
8 mm ± (0.
61 mm) (p < 0.
001)—and in the group of long eyes (n = 5)—mean AL (A) 27.
95 mm (±2.
62 mm) vs.
mean AL (IM) 28.
10 mm (±2.
64) (p < 0.
05).
In the group of average eyes (n = 74), outcomes were similar—mean AL (A) 23.
56 mm (±0.
70 mm) vs.
mean AL (IM) 23,56 mm (±0.
71 mm) (p > 0.
05).
The anterior chamber depth measurements were higher when obtained with Argos than with IOLMaster 700—mean ACD (A) 3.
06 mm (±0.
48 mm) vs.
mean ACD (IM) 2.
92 mm (±0.
46) p < 0.
001.
There was no statistical significance in mean LT—mean LT (A) 4.
75 mm (±0.
46 mm) vs.
mean LT (IM) 4.
72 mm (±0.
44 mm) (p = 0.
054).
The biometry of one eye with dense cataracts could be measured only with Argos, using the Enhanced Retinal Visualization mode.
Conclusions: Axial length measurements from both devices were different in the groups of short and long eyes, but were comparable in the group of average eyes.
The anterior chamber depth values obtained with Argos were higher than the measurements acquired with IOLMaster 700.
These differences may be particularly important when selecting IOLs for patients with extreme AL values.
Related Results
Comparison of four formulas for the intraocular lens power prediction
Comparison of four formulas for the intraocular lens power prediction
Abstract
Purpose
To assess the accuracy of four formulas for intraocular lens (IOL) power prediction in cataractous eyes.
METHODS
In this prospective study, 51 eyes of 3...
Biometry Measurements Using a New Large Coherence Length Swept-Source Optical Coherence Tomography
Biometry Measurements Using a New Large Coherence Length Swept-Source Optical Coherence Tomography
Abstract
The swept-source OCT biometer (Argos) uses a 1060 nm wavelength and 20 nm bandwidth swept-source technology to collect 2-D OCT data of the full eye. The device m...
Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract
Comparison of ocular biometric parameters between two swept-source optical coherence tomography devices and Scheimpflug tomography in patients with cataract
AIM: To assess and compare the variations and agreements across different ocular biometric parameters using swept-source optical coherence tomography (SS-OCT) and Scheimpflug tomog...
Interpretation of Simulated Falloff Tests
Interpretation of Simulated Falloff Tests
Abstract
A study was made of the method for determination of swept volume and the proper average temperature to use for interpretation of combustion falloff data ...
The impact of cataract surgery on tear film physiology: signs and symptoms, progression and treatment
The impact of cataract surgery on tear film physiology: signs and symptoms, progression and treatment
PurposeThis study aimed to revise data published in the literature on the effects of cataract surgery on tear film characteristics, in relation to personal clinical surgical experi...
Traitement des signaux Argos 4
Traitement des signaux Argos 4
Cette thèse est dédié à l’étude de la problématique des interférences multi utilisateurs dans le système Argos et à la proposition des diverses techniques pour réduire les effets d...
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length).
Methods: This single-...
Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients
Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients
Purpose
To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), ...