Biometry Measurements Using a New Large Coherence Length Swept-Source Optical Coherence Tomography
View through CrossRef
Abstract
The swept-source OCT biometer (Argos) uses a 1060 nm wavelength and 20 nm bandwidth swept-source technology to collect 2-D OCT data of the full eye. The device measures the axial length and the anterior chamber depth, the central corneal thickness, aqueous depth, lens thickness, pupil size, the corneal diameter, and the corneal power in less than a second. The repeatability and reproducibility of the Argos measurements have been tested by means of variation analysis study and demonstrated that the new OCT biometer produces precise and reproducible measurements. Measurements made with the swept-source OCT biometer show a high acquisition rate compared with the IOLMaster 500 and the Lenstar biometers, especially in dense cataracts. The Argos swept-source optical coherence tomography measures the optical path length of each segment of the eye and uses a specific refractive index for each of these segments that has led to an improvement in IOL power calculation especially in long and short eyes.
Title: Biometry Measurements Using a New Large Coherence Length Swept-Source Optical Coherence Tomography
Description:
Abstract
The swept-source OCT biometer (Argos) uses a 1060 nm wavelength and 20 nm bandwidth swept-source technology to collect 2-D OCT data of the full eye.
The device measures the axial length and the anterior chamber depth, the central corneal thickness, aqueous depth, lens thickness, pupil size, the corneal diameter, and the corneal power in less than a second.
The repeatability and reproducibility of the Argos measurements have been tested by means of variation analysis study and demonstrated that the new OCT biometer produces precise and reproducible measurements.
Measurements made with the swept-source OCT biometer show a high acquisition rate compared with the IOLMaster 500 and the Lenstar biometers, especially in dense cataracts.
The Argos swept-source optical coherence tomography measures the optical path length of each segment of the eye and uses a specific refractive index for each of these segments that has led to an improvement in IOL power calculation especially in long and short eyes.
Related Results
Interpretation of Simulated Falloff Tests
Interpretation of Simulated Falloff Tests
Abstract
A study was made of the method for determination of swept volume and the proper average temperature to use for interpretation of combustion falloff data ...
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
Aerodynamic Characteristics of Forward Swept Wing in Subsonic Speed
Aerodynamic Characteristics of Forward Swept Wing in Subsonic Speed
A forward swept wing was designed to use for a supersonic aircraft. Its aerodynamic characteristics were studied through experiments and numerical simulations which excluded the su...
EN FACE IMAGING OF RETINAL ARTERY MACROANEURYSMS USING SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY
EN FACE IMAGING OF RETINAL ARTERY MACROANEURYSMS USING SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY
Purpose:
To describe the advantages of en face view with swept-source optical coherence tomography in assessing the morphologic features of retinal arterial macroaneury...
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Background: Accurate biometry is critical for optimal visual outcomes in cataract surgery. Both optical and ultrasound biometry are widely used, but their comparative effectiveness...
STARDUST SIGN AND RETINAL TEAR DETECTION ON SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY
STARDUST SIGN AND RETINAL TEAR DETECTION ON SWEPT SOURCE OPTICAL COHERENCE TOMOGRAPHY
Purpose:
The causes of floaters include posterior vitreous detachment and fundus hemorrhage, both of which are risk factors for retinal tears. We observed the vitreous ...
Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
Evaluation of Selected Biometric Parameters in Cataract Patients—A Comparison between Argos® and IOLMaster 700®: Two Swept-Source Optical Coherence Tomography-Based Biometers
Background and Objectives: To compare the biometry of eyes obtained with two swept-source optical coherence tomography-based biometers—Argos (A), using an individual refractive ind...
OPTICAL COHERENCE TOMOGRAPHY-ANGIOGRAPHY, A NEW DIAGNOSTIC AND FOLLOW-UP TOOL FOR GLAUCOMA
OPTICAL COHERENCE TOMOGRAPHY-ANGIOGRAPHY, A NEW DIAGNOSTIC AND FOLLOW-UP TOOL FOR GLAUCOMA
Glaucoma is an optic neuropathy and is one of the leading causes of irreversible blindness worldwide. There are studies on the role of vascular dysfunction in the pathogenesis of g...