A phantom for 2D dose measurements in the vicinity of metal implants for photon and proton beams
View through CrossRef
This work aimed to characterize a dedicated phantom for assessing the dose near metal implants for radiotherapy with photons and protons. A dosimetry audit phantom was redesigned to position a Gafchromic EBT-3 film within a bisected titanium pedicle screw (6.5 mm diameter). The mass density and the water equivalent thickness (WET) of the phantom material were determined. The phantom was irradiated using a photon arc and a horizontal proton beam in combination with a couch rotation of 20°, with three repeated measurements each. Treatment plans utilizing a single field covering the screw and the EBT-3 film were optimized to deliver a physical dose of 2 Gy using a collapsed cone and Monte Carlo dose engine for photons and protons, respectively. The mass density and the WET of the phantom were determined as (1.033 ± 0.010) gcm−3 and (1.022 ± 0.013), respectively. Ionisation chamber measurements agreed within 1% (photons) and 0.5% (protons) with the calculated dose values. Relative photon dosimetry measurements using EBT-3 films revealed an agreement between measured and calculated horizontal profiles within the confidence interval for areas beyond 5 mm from the center. For photon plans, significant deviations of more than 10% were found at the interfaces between phantom material and screw. The proton measurements showed a gradual decrease of 3% across both profiles. In contrast to photon plans, no dose increase was measured within the screw, but significant dose fluctuations (>5%) in the beam’s exit region. This study showed that the behavior of dose engine is affected by metal implants and thus dosimetric measurements are highly recommended. The presented phantom can serve as foundation for dedicated end-2-end phantoms.
Frontiers Media SA
Title: A phantom for 2D dose measurements in the vicinity of metal implants for photon and proton beams
Description:
This work aimed to characterize a dedicated phantom for assessing the dose near metal implants for radiotherapy with photons and protons.
A dosimetry audit phantom was redesigned to position a Gafchromic EBT-3 film within a bisected titanium pedicle screw (6.
5 mm diameter).
The mass density and the water equivalent thickness (WET) of the phantom material were determined.
The phantom was irradiated using a photon arc and a horizontal proton beam in combination with a couch rotation of 20°, with three repeated measurements each.
Treatment plans utilizing a single field covering the screw and the EBT-3 film were optimized to deliver a physical dose of 2 Gy using a collapsed cone and Monte Carlo dose engine for photons and protons, respectively.
The mass density and the WET of the phantom were determined as (1.
033 ± 0.
010) gcm−3 and (1.
022 ± 0.
013), respectively.
Ionisation chamber measurements agreed within 1% (photons) and 0.
5% (protons) with the calculated dose values.
Relative photon dosimetry measurements using EBT-3 films revealed an agreement between measured and calculated horizontal profiles within the confidence interval for areas beyond 5 mm from the center.
For photon plans, significant deviations of more than 10% were found at the interfaces between phantom material and screw.
The proton measurements showed a gradual decrease of 3% across both profiles.
In contrast to photon plans, no dose increase was measured within the screw, but significant dose fluctuations (>5%) in the beam’s exit region.
This study showed that the behavior of dose engine is affected by metal implants and thus dosimetric measurements are highly recommended.
The presented phantom can serve as foundation for dedicated end-2-end phantoms.
Related Results
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration ...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Abstract
Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure a...
Clonogenic survival measurement of human glioblastoma (U87) cells cultured in 3Dbio-phantom after proton and X-ray irradiations
Clonogenic survival measurement of human glioblastoma (U87) cells cultured in 3Dbio-phantom after proton and X-ray irradiations
Abstract
Background: Glioblastoma multiforme (GBM) is the most malignant brain tumor and is resistant to conventional radiotherapy. Proton radiotherapy utilizes accelerated...
Construction of Bi-Pearcey beams and their mathematical mechanism
Construction of Bi-Pearcey beams and their mathematical mechanism
We present a theoretical expression in the form of the Pearcey function by deducing the Fresnel diffraction distribution of an elliptic line. Then, we numerically simulate and expe...
In silico analysis of optimum photon energy spectra and beam parameters for iodine nanoparticle–aided orthovoltage radiation therapy of brain tumors
In silico analysis of optimum photon energy spectra and beam parameters for iodine nanoparticle–aided orthovoltage radiation therapy of brain tumors
In this study, the effect of photon energy spectra on the dose enhancement factor (DEF) and other influencing parameters such as skull–tumor dose ratios were estimated for a confin...
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams
Dosimetric characteristics of radiophotoluminescent glass dosimeters for proton beams
Abstract
Background: The study of radiophotoluminescent glass dosimeters (RGDs) in the clinical usage of proton beams is limited. The aim of this study was to investigate t...