SU‐FF‐T‐448: Validation of a New Photon Dose Calculation Model—Analytical Anisotropic Algorithm

View through CrossRef
Purpose: To validate a new photon dose calculation model Analytical Anisotropic Algorithm (AAA) on Eclipse™ treatment planning system (TPS). Comparison of AAA dose calculation was performed with measurements and other two conventional algorithms, Pencil Beam Convolution (PBC) algorithm on Eclipse™ and Collapsed Cone Convolution/Superposition (CCC) algorithm on Pinnacle3.0 TPS.Method and Materials: Four phantoms were CT scanned and the image set was imported into both TPS for dose computation and analysis. The four phantoms were: 1) homogenous tissue equivalent phantom, 2) tissue equivalent phantom with infinite lung heterogeneity, 3) tissue equivalent phantom with finite lung, 4) IMRT dose verification phantom. Measurements were made by exposing the phantom using Varian Linac. Point measurement and film measurements were compared with calculated results from the three algorithms. Dose responses for high and low energy photon beams were investigated for several different depths and PDD curves were compared in the phantom for various field sizes. The IMRT plans were generated by both TPS and were performed on the IMRT phantom to compare fluence maps. Results: AAA dose prediction fits the film measurements well except that there is up to ±6% discrepancy for dose profile perpendicular to the interface of tissue and lung. Point measurements support the AAA algorithm calculations. AAA also accurately predicts the decrease in PDD curves due to the lung inhomogeneity for 6MV energy. For the high energy photon beam and very small field size (2cm*2cm) in lung region, AAA prediction is up to 8% lower than the measurements. Conclusion: AAA algorithm accounts for attenuation corrections and electron transport, and models the deposited dose in the lung with greater accuracy than PBC. It is also faster than CCC algorithm. AAA algorithm can not accurately model the lateral scattering in tissue heterogeneity, but it can still give a reasonably close (within ±6%) prediction.
Title: SU‐FF‐T‐448: Validation of a New Photon Dose Calculation Model—Analytical Anisotropic Algorithm
Description:
Purpose: To validate a new photon dose calculation model Analytical Anisotropic Algorithm (AAA) on Eclipse™ treatment planning system (TPS).
Comparison of AAA dose calculation was performed with measurements and other two conventional algorithms, Pencil Beam Convolution (PBC) algorithm on Eclipse™ and Collapsed Cone Convolution/Superposition (CCC) algorithm on Pinnacle3.
0 TPS.
Method and Materials: Four phantoms were CT scanned and the image set was imported into both TPS for dose computation and analysis.
The four phantoms were: 1) homogenous tissue equivalent phantom, 2) tissue equivalent phantom with infinite lung heterogeneity, 3) tissue equivalent phantom with finite lung, 4) IMRT dose verification phantom.
Measurements were made by exposing the phantom using Varian Linac.
Point measurement and film measurements were compared with calculated results from the three algorithms.
Dose responses for high and low energy photon beams were investigated for several different depths and PDD curves were compared in the phantom for various field sizes.
The IMRT plans were generated by both TPS and were performed on the IMRT phantom to compare fluence maps.
Results: AAA dose prediction fits the film measurements well except that there is up to ±6% discrepancy for dose profile perpendicular to the interface of tissue and lung.
Point measurements support the AAA algorithm calculations.
AAA also accurately predicts the decrease in PDD curves due to the lung inhomogeneity for 6MV energy.
For the high energy photon beam and very small field size (2cm*2cm) in lung region, AAA prediction is up to 8% lower than the measurements.
Conclusion: AAA algorithm accounts for attenuation corrections and electron transport, and models the deposited dose in the lung with greater accuracy than PBC.
It is also faster than CCC algorithm.
AAA algorithm can not accurately model the lateral scattering in tissue heterogeneity, but it can still give a reasonably close (within ±6%) prediction.

Related Results

The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...
Effect of glycolysis inhibition by miR-448 on glioma radiosensitivity
Effect of glycolysis inhibition by miR-448 on glioma radiosensitivity
OBJECTIVEAlthough glucose metabolism reengineering is a typical feature of various tumors, including glioma, key regulators of glycolytic reprogramming are still poorly understood....
Shadow and photon ring of black hole in asymptotically safe gravity
Shadow and photon ring of black hole in asymptotically safe gravity
In this paper, we focus on discussing the influence of thin disk accretion and asymptotically safe (AS) gravity correction parameter on the shadow and photon ring of black holes. F...
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration ...
Multi‐Photon Microscopy
Multi‐Photon Microscopy
AbstractIn this series of papers on light microscopy imaging, we have covered the fundamentals of microscopy, super‐resolution microscopy, and lightsheet microscopy. This last revi...
SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation
SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation
Purpose: To commissioningRadCalc program for secondary point dose check in conventional 2D and 3DCRT calculation techniques. Methods: Test plans for 6 MV and 10 MV photon beams wit...

Back to Top