Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams

View through CrossRef
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration process is required. To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile measured at different monitor units (MUs) is introduced and tested in a proton and photon beam.MethodsThe calibration procedure employs the dose ratio of film measurements of the same relative profile for different absolute dose values. Hence, the ratio of the dose is constant at any point of the profile, but the ratio of the net optical densities is not constant. The key idea of the method is to optimize the calibration function until the ratio of the calculated doses is constant. The proposed method was tested in the dose range between 0.25–12 and 1–6 Gy in a proton and photon beam, respectively. A radial symmetric profile and a rectangular profile were created, both having a central plateau region of about 3 cm diameter and a dose falloff of about 1.5 cm at larger distances. The dose falloff region was used as input for the optimization method and the central plateau region served as dose reference points. Only the plateau region of the highest dose entered the optimization as an additional objective. The measured data were randomly split into differently sized training and test sets. The optimization was repeated 1000 times with random start value initialization using the same start values for the standard and the gradient method. Finally, a proton plan with four dose levels was created, which were separated spatially, to test the possibility of a full calibration within a single measurement.ResultsParameter estimation was possible with as low as one dose ratio used for optimization in both the photon and the proton case, yet exhibiting a high sensitivity on the dose level. The root mean squared deviation (RMSD) of the dose was less than 1% when the dose ratio was in the order of 20, whereas the median RMSD of all optimizations was 1.7%. Using four dose levels for optimization resulted in a median RMSD of 1% when randomly selecting the dose levels. Having at least one dose ratio of about 20 included in the optimization considerably improved the RMSD of the calibration function. Using six or eight dose levels reduced the sensitivity on the dose level selection and the median RMSD was 0.8%. A full calibration was possible in a single measurement having four dose levels in one plan but spatially separated.ConclusionsThe number of measurements required to obtain an EBT3 film calibration function could be reduced using the proposed dose ratio method while maintaining the same accuracy as with the standard method.
Title: Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
Description:
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence.
To assure high precision and accuracy, a time‐consuming calibration process is required.
To improve the time efficiency, a novel calibration method utilizing the ratio of the same dose profile measured at different monitor units (MUs) is introduced and tested in a proton and photon beam.
MethodsThe calibration procedure employs the dose ratio of film measurements of the same relative profile for different absolute dose values.
Hence, the ratio of the dose is constant at any point of the profile, but the ratio of the net optical densities is not constant.
The key idea of the method is to optimize the calibration function until the ratio of the calculated doses is constant.
The proposed method was tested in the dose range between 0.
25–12 and 1–6 Gy in a proton and photon beam, respectively.
A radial symmetric profile and a rectangular profile were created, both having a central plateau region of about 3 cm diameter and a dose falloff of about 1.
5 cm at larger distances.
The dose falloff region was used as input for the optimization method and the central plateau region served as dose reference points.
Only the plateau region of the highest dose entered the optimization as an additional objective.
The measured data were randomly split into differently sized training and test sets.
The optimization was repeated 1000 times with random start value initialization using the same start values for the standard and the gradient method.
Finally, a proton plan with four dose levels was created, which were separated spatially, to test the possibility of a full calibration within a single measurement.
ResultsParameter estimation was possible with as low as one dose ratio used for optimization in both the photon and the proton case, yet exhibiting a high sensitivity on the dose level.
The root mean squared deviation (RMSD) of the dose was less than 1% when the dose ratio was in the order of 20, whereas the median RMSD of all optimizations was 1.
7%.
Using four dose levels for optimization resulted in a median RMSD of 1% when randomly selecting the dose levels.
Having at least one dose ratio of about 20 included in the optimization considerably improved the RMSD of the calibration function.
Using six or eight dose levels reduced the sensitivity on the dose level selection and the median RMSD was 0.
8%.
A full calibration was possible in a single measurement having four dose levels in one plan but spatially separated.
ConclusionsThe number of measurements required to obtain an EBT3 film calibration function could be reduced using the proposed dose ratio method while maintaining the same accuracy as with the standard method.

Related Results

Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Abstract Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure a...
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
It was always based on a teenage love story between the two kids. One is a sniffer and one is not. It was designed for Central Australia because we do write these kids off there. N...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...
An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties
An analytical formalism for the assessment of dose uncertainties due to positioning uncertainties
PurposeTo present an analytical formalism for the in depth assessment of uncertainties of field output factors in small fields related to detector positioning based on dose profile...
Ultraviolet radiation dosimetry using radiochromic film for extended dose range applications and estimating eye exposure
Ultraviolet radiation dosimetry using radiochromic film for extended dose range applications and estimating eye exposure
Abstract Radiochromic films are a reliable tool for quantifying ultraviolet exposure dose. When used without modification, these films enable a simple, low-cost m...
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Alternative Entrances: Phillip Noyce and Sydney’s Counterculture
Phillip Noyce is one of Australia’s most prominent film makers—a successful feature film director with both iconic Australian narratives and many a Hollywood blockbuster under his ...
Dose distribution in a radiosurgery treatment for an acoustic neuroma
Dose distribution in a radiosurgery treatment for an acoustic neuroma
Evaluation of the dose distribution delivered in a radiosurgery treatment of an acoustic neuroma using an adapted Alderson phantom and radiochromic film to assess absolute dose. A ...

Back to Top