Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
View through CrossRef
Abstract
Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer. This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.4 MeV and 148.2 MeV proton beams and 15–100 kV x-rays. Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated. Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak. EBT-XD films showed the same characteristic on film darkening as EBT3. The effects between portrait and landscape orientation were reduced by 3.1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy. The LRA is reduced for EBT-XD films for all investigated dose ranges. The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy. In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency. A quenching effect of 10% was found for both film types. The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams. Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.
Title: Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Description:
Abstract
Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy.
The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer.
This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.
4 MeV and 148.
2 MeV proton beams and 15–100 kV x-rays.
Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated.
Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak.
EBT-XD films showed the same characteristic on film darkening as EBT3.
The effects between portrait and landscape orientation were reduced by 3.
1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy.
The LRA is reduced for EBT-XD films for all investigated dose ranges.
The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy.
In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency.
A quenching effect of 10% was found for both film types.
The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams.
Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.
Related Results
Characterization of EBT3 radiochromic films for dosimetry of proton beams in the presence of magnetic fields
Characterization of EBT3 radiochromic films for dosimetry of proton beams in the presence of magnetic fields
PurposeRadiochromic film dosimetry is extensively used for quality assurance in photon and proton beam therapy. So far, GafchromicTM EBT3 film appears as a strong candidate to be u...
Measurements of Radiochromic Films (EBT3 and EBT-XD) Using Three Portable Colorimeters
Measurements of Radiochromic Films (EBT3 and EBT-XD) Using Three Portable Colorimeters
The authors recently proposed a method for on-site dosimetry by a combination of radiochromic film and portable colorimeter. To test and expand the applicability of this method, do...
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi dosis radiasi teknik penyinaran Three-Dimensional Conformal Radiation Therapy  (3D-CRT) pada pasien kanker payudara telah dilakukan menggunakan film EBT3. Verifikasi do...
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses...
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
Reclaiming the Wasteland: Samson and Delilah and the Historical Perception and Construction of Indigenous Knowledges in Australian Cinema
It was always based on a teenage love story between the two kids. One is a sniffer and one is not. It was designed for Central Australia because we do write these kids off there. N...
Adsorptive Removal of Hazardous Eriochrome Black T and Its Metal Complexes from Aqueous Media Using Spent Coffee Grounds
Adsorptive Removal of Hazardous Eriochrome Black T and Its Metal Complexes from Aqueous Media Using Spent Coffee Grounds
Abstract
An adsorbent was prepared from the spent coffee grounds (SCG) waste by simple washing and tested for its effective use as a low-cost adsorbent for removing eriochr...
A phantom for 2D dose measurements in the vicinity of metal implants for photon and proton beams
A phantom for 2D dose measurements in the vicinity of metal implants for photon and proton beams
This work aimed to characterize a dedicated phantom for assessing the dose near metal implants for radiotherapy with photons and protons. A dosimetry audit phantom was redesigned t...
Odd version Mathieu-Gaussian beam based on Green function
Odd version Mathieu-Gaussian beam based on Green function
Like the theoretical pattern of non-diffracting Bessel beams, ideal non-diffracting Mathieu beams also carry infinite energy, but cannot be generated as a physically realizable ent...