SU‐GG‐T‐173: Dosimetric Verification of IMRT Fields Using GafChromic® EBT‐2 Film

View through CrossRef
Purpose: Film based dosimetry often has advantages over detector based dosimetry when measurements with high spatial resolution are needed. We present comparisons of dosimetric verification measurements using GafChromic® EBT‐2 film with IBA Dosimetry® I'mRT MatriXX with MultiCube. Methods and Materials: Dosimetric measurements of IMRT treatment fields for 5 patients were performed using GafChromic® EBT‐2 film and compared to measurements performed using IBA Dosimetry® MatriXX. The measurement results from both systems were analyzed using IBA Dosimetry® OmniPro‐I'mRT software. GafChromic® films (8”×10”) were scanned both pre‐ and post‐irradiation using an Epson® 10000XL flatbed scanner. The pre‐irradiation films were used for background subtraction in the OmniPro software. The films were placed in a solid water phantom for irradiation. The MatriXX system uses a 32×32 array of 0.08 cc ion chambers with a center to center spacing of approximately 0.76 cm to measure absolute dose distributions within a solid phantom. Verification plans were created in Pinnacle treatment planning software for both the MatriXX and solid water phantoms. The results were analyzed by comparison of the measured central axis dose, gamma analysis (3 % and 3mm agreement), and evaluation of dose profiles and isodose distributions. Results: Better agreement with the planned central axis dose is seen for the measurements with the MatriXX system over those measured with GafChromic® EBT‐2 film. However, the film‐based measurements show improved agreement in the dose profiles and isodose distributions. These improvements result in better gamma analysis agreement, increasing the average gamma from 97.39% to 99.49%. Conclusion: GafChromic® EBT‐2 can effectively be used for film‐based dosimetry and may provide superior results for small fields, areas with steep gradients, or beam delivery geometries where detector based measurements are limited. Some dosimetry software, such as IBA Dosimetry® OmniPro‐I'mRT, may allow users to analyze GafChromic® EBT‐2 film without purchasing additional software.
Title: SU‐GG‐T‐173: Dosimetric Verification of IMRT Fields Using GafChromic® EBT‐2 Film
Description:
Purpose: Film based dosimetry often has advantages over detector based dosimetry when measurements with high spatial resolution are needed.
We present comparisons of dosimetric verification measurements using GafChromic® EBT‐2 film with IBA Dosimetry® I'mRT MatriXX with MultiCube.
Methods and Materials: Dosimetric measurements of IMRT treatment fields for 5 patients were performed using GafChromic® EBT‐2 film and compared to measurements performed using IBA Dosimetry® MatriXX.
The measurement results from both systems were analyzed using IBA Dosimetry® OmniPro‐I'mRT software.
GafChromic® films (8”×10”) were scanned both pre‐ and post‐irradiation using an Epson® 10000XL flatbed scanner.
The pre‐irradiation films were used for background subtraction in the OmniPro software.
The films were placed in a solid water phantom for irradiation.
The MatriXX system uses a 32×32 array of 0.
08 cc ion chambers with a center to center spacing of approximately 0.
76 cm to measure absolute dose distributions within a solid phantom.
Verification plans were created in Pinnacle treatment planning software for both the MatriXX and solid water phantoms.
The results were analyzed by comparison of the measured central axis dose, gamma analysis (3 % and 3mm agreement), and evaluation of dose profiles and isodose distributions.
Results: Better agreement with the planned central axis dose is seen for the measurements with the MatriXX system over those measured with GafChromic® EBT‐2 film.
However, the film‐based measurements show improved agreement in the dose profiles and isodose distributions.
These improvements result in better gamma analysis agreement, increasing the average gamma from 97.
39% to 99.
49%.
Conclusion: GafChromic® EBT‐2 can effectively be used for film‐based dosimetry and may provide superior results for small fields, areas with steep gradients, or beam delivery geometries where detector based measurements are limited.
Some dosimetry software, such as IBA Dosimetry® OmniPro‐I'mRT, may allow users to analyze GafChromic® EBT‐2 film without purchasing additional software.

Related Results

Alkion "Puukkojunkkareista"
Alkion "Puukkojunkkareista"
Bemerkungen über Santeri Alkios Werk "Puukkojunkkarit" (saksa)Kielenainekset*loika (kieli: suomi, sivulla: 172)apajus (kieli: suomi, sivulla: 171)asettaa (kieli: suomi, sivulla: 17...
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams
Abstract Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure a...
Composition Measurements of Uranus’ Atmosphere
Composition Measurements of Uranus’ Atmosphere
Knowing the composition of the giant planets is important in understanding their forma­tion and evolution history. The abundances of heavy elements, of noble gases, and is...
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
Dosimetric comparison of 3DCRT and IMRT in radical chemoradiotherapy of squamous cell carcinoma esophagus
ABSTRACT Background: Radical chemoradiation is the standard of treatment for locally advanced squamous cell carcinoma of esophagus and for patien...
Heittää verbin semantiikkaa
Heittää verbin semantiikkaa
Des significations du verbe heittää (ranska)Kielenaineksetheipata (kieli: suomi, sivulla: 173)heitellä (kieli: suomi, sivulla: 165)heitlema (kieli: viro, sivulla: 165)heitma (kieli...
Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic® EBT3 film
Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic® EBT3 film
AbstractPurposeThis study aimed to examine the dosimetric properties of Gafchromic® EBT3 film and intensity-modulated radiation therapy quality assurance (IMRT QA).Materials and me...
SU‐GG‐T‐211: An Independent MU Calculation Program for Dynamic IMRT
SU‐GG‐T‐211: An Independent MU Calculation Program for Dynamic IMRT
Purpose: The performance of an independent monitor unit calculation program is evaluated to verify the monitor unit (MU) for dynamic IMRT fields. The MU calculation program runs on...

Back to Top