SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation

View through CrossRef
Purpose: To commissioningRadCalc program for secondary point dose check in conventional 2D and 3DCRT calculation techniques. Methods: Test plans for 6 MV and 10 MV photon beams with conventional and 3DCRT planwere created inRadCalc program.Field was varied with a range of field sizes and modifier techniques referred to IAEA‐TECDOC‐1583 with central axis and off central axis calculation points. The measured dose for each field was then compared to the calculated dose (RadCalc program and Eclipse treatment planning system). Results: In conventional field validation,RadCalc shows good agreement with measured doses.For open and almost all wedge fields, RadCalccalculationagree with dose measurement within +/−1%. Differences of over 1% were found in maximum wedge fields and some half fields. However, the dose difference between calculation and measurementwas within +/−2% in all tests. In 3DCRT plan validation, the central axis point dose between calculation and measurement in all tests are agree within the IAEA‐TECDOC‐1583 criteria. When measured/ calculated dose points were in non‐tissue equivalent such as lung and bone which were off central axis situation, dose difference is more than +/−10%. Unfortunately if measured/ calculated dose points areoutside the field (scatter dose), the result cannot be validated. Part of this large difference between two calculated and measured can be the volume effect of the chamber.In comparison between dose calculated by RadCalc and Eclipse, overall results show dose difference within +/−5%. Conclusions: RadCalc program can be used as a point dose calculation check in conventional 2D and 3DCRT techniques.
Title: SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation
Description:
Purpose: To commissioningRadCalc program for secondary point dose check in conventional 2D and 3DCRT calculation techniques.
Methods: Test plans for 6 MV and 10 MV photon beams with conventional and 3DCRT planwere created inRadCalc program.
Field was varied with a range of field sizes and modifier techniques referred to IAEA‐TECDOC‐1583 with central axis and off central axis calculation points.
The measured dose for each field was then compared to the calculated dose (RadCalc program and Eclipse treatment planning system).
Results: In conventional field validation,RadCalc shows good agreement with measured doses.
For open and almost all wedge fields, RadCalccalculationagree with dose measurement within +/−1%.
Differences of over 1% were found in maximum wedge fields and some half fields.
However, the dose difference between calculation and measurementwas within +/−2% in all tests.
In 3DCRT plan validation, the central axis point dose between calculation and measurement in all tests are agree within the IAEA‐TECDOC‐1583 criteria.
When measured/ calculated dose points were in non‐tissue equivalent such as lung and bone which were off central axis situation, dose difference is more than +/−10%.
Unfortunately if measured/ calculated dose points areoutside the field (scatter dose), the result cannot be validated.
Part of this large difference between two calculated and measured can be the volume effect of the chamber.
In comparison between dose calculated by RadCalc and Eclipse, overall results show dose difference within +/−5%.
Conclusions: RadCalc program can be used as a point dose calculation check in conventional 2D and 3DCRT techniques.

Related Results

Analysis of commissioning and procurement of adult social care in Scotland
Analysis of commissioning and procurement of adult social care in Scotland
Background: Strategic planning and commissioning of adult social care (ASC) involves forecasting population needs across a defined area, together with availability of services or r...
TUNING AND VALIDATION OF THE NEW RADCALC 3D MONTE CARLO BASED PRE-TREATMENT PLAN VERIFICATION TOOL
TUNING AND VALIDATION OF THE NEW RADCALC 3D MONTE CARLO BASED PRE-TREATMENT PLAN VERIFICATION TOOL
Verification procedures for patient-specific quality assurance (QA) in advanced radiotherapy are laborious and time-consuming. Moreover, it has been shown that these procedures can...
TU‐G‐116‐01: A Roadmap for Installing and Commissioning New Equipment
TU‐G‐116‐01: A Roadmap for Installing and Commissioning New Equipment
Learning Objectives:1. To learn how to build a new equipment commissioning project plan using efficient project management methods2. To identify equipment, personnel resources and ...
The photon blockade effect of a complete Buck-Sukumar model
The photon blockade effect of a complete Buck-Sukumar model
The Buck-Sukumar (BS) model, with a nonlinear coupling between the atom and the light field, is well defined only when its coupling strength is lower than a critical coupling. Its ...
Verification of High Speed on Chip with VIP using System Verilog
Verification of High Speed on Chip with VIP using System Verilog
Abstract - The exploration work is addressing verification of High speed on chips protocol; we've used the system Verilog grounded test bench structure. I developed a system Verilo...
Early startup methodologies in gas plant commissioning: An analysis of effective strategies and their outcomes
Early startup methodologies in gas plant commissioning: An analysis of effective strategies and their outcomes
Gas plant commissioning is a critical phase in the lifecycle of gas facilities, where the transition from construction to operational readiness occurs. Early startup methodologies ...
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration ...
Effective Approach for Enhancement of Rigless Well Commissioning
Effective Approach for Enhancement of Rigless Well Commissioning
Abstract Rigless well commissioning is an important stage of well completion to deliver a new drilled or worked-over wells (oil producer or water/gas injector) ready...

Back to Top