TUNING AND VALIDATION OF THE NEW RADCALC 3D MONTE CARLO BASED PRE-TREATMENT PLAN VERIFICATION TOOL
View through CrossRef
Verification procedures for patient-specific quality assurance (QA) in advanced radiotherapy are laborious and time-consuming. Moreover, it has been shown that these procedures cannot detect some inaccuracies for some particular complex cases due to tissue inhomogeneity and highly modulated plans. Secondary dose calculation verification of radiotherapy plans is an important aspect in patient-specific QA. A suitably optimized software, RadCalc equipped with 3D Monte Carlo Module (MC), was used to dosimetrically verify radiotherapy treatment plans where the measured dose distributions can be inaccurate due to the TPS dose calculation algorithm and/or treatment unit delivery uncertainties. MC Models were built using specific commissioning measurements and then the Additional Radiation to Light Field Offset parameter (Dosimetric Leaf Gap parameter) was tuned to achieve the best dose comparison agreement with phantom patient-specific QA measurements. The results showed a good agreement between the TPS and the simulations. RadCalc MC also allows to better estimate the plan doses in lung cancer patients and to detect the presence of possible inaccuracies due to tissue inhomogeneity, which is not estimable with a homogeneous phantoms.
Title: TUNING AND VALIDATION OF THE NEW RADCALC 3D MONTE CARLO BASED PRE-TREATMENT PLAN VERIFICATION TOOL
Description:
Verification procedures for patient-specific quality assurance (QA) in advanced radiotherapy are laborious and time-consuming.
Moreover, it has been shown that these procedures cannot detect some inaccuracies for some particular complex cases due to tissue inhomogeneity and highly modulated plans.
Secondary dose calculation verification of radiotherapy plans is an important aspect in patient-specific QA.
A suitably optimized software, RadCalc equipped with 3D Monte Carlo Module (MC), was used to dosimetrically verify radiotherapy treatment plans where the measured dose distributions can be inaccurate due to the TPS dose calculation algorithm and/or treatment unit delivery uncertainties.
MC Models were built using specific commissioning measurements and then the Additional Radiation to Light Field Offset parameter (Dosimetric Leaf Gap parameter) was tuned to achieve the best dose comparison agreement with phantom patient-specific QA measurements.
The results showed a good agreement between the TPS and the simulations.
RadCalc MC also allows to better estimate the plan doses in lung cancer patients and to detect the presence of possible inaccuracies due to tissue inhomogeneity, which is not estimable with a homogeneous phantoms.
Related Results
Optimising tool wear and workpiece condition monitoring via cyber-physical systems for smart manufacturing
Optimising tool wear and workpiece condition monitoring via cyber-physical systems for smart manufacturing
Smart manufacturing has been developed since the introduction of Industry 4.0. It consists of resource sharing and networking, predictive engineering, and material and data analyti...
SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation
SU‐E‐T‐348: Commissioning of An Independent Monitor Unit Verification Program (RadCalc Version 6.1) in Photon Point Dose Calculation
Purpose: To commissioningRadCalc program for secondary point dose check in conventional 2D and 3DCRT calculation techniques. Methods: Test plans for 6 MV and 10 MV photon beams wit...
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
Monte Carlo methods: barrier option pricing with stable Greeks and multilevel Monte Carlo learning
For discretely observed barrier options, there exists no closed solution under the Black-Scholes model. Thus, it is often helpful to use Monte Carlo simulations, which are easily a...
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Validation in Doctoral Education: Exploring PhD Students’ Perceptions of Belonging to Scaffold Doctoral Identity Work
Aim/Purpose: The aim of this article is to make a case of the role of validation in doctoral education. The purpose is to detail findings from three studies which explore PhD stude...
Research on Multi-Group Monte Carlo Calculations Based on Group Constants Generated by RMC
Research on Multi-Group Monte Carlo Calculations Based on Group Constants Generated by RMC
Abstract
Nowadays, deterministic two-step or Monte Carlo methods are commonly used in core physics calculations. However, with the development of reactor core design, tradi...
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Shenzi 16-Inch Oil Export SCR CVA Verification
Shenzi 16-Inch Oil Export SCR CVA Verification
Abstract
In 2006 Enterprise developed a 16-inch oil export system from Shenzi field located in Green Canyon Block 653 in the Gulf of Mexico, approximately 120 nau...
Absolute quantification in brain SPECT imaging
Absolute quantification in brain SPECT imaging
Certes malalties neurològiques estan associades amb problemes en els sistemes de neurotransmissió. Una aproximació a l'estudi d'aquests sistemes és la tomografia d'emissió SPECT (S...