Effect of pupil dilation on IOLMaster® measurements in Pakistani Eyes
View through CrossRef
Objective: To analyse the effect of pupil dilation on intraocular lens instrument IOLMaster 500 biometric measurements, and to determine the effect of these measurements on intraocular lens power calculations in Asian eyes.
Methods: The prospective study was conducted at the Aga Khan University Hospital, Karachi, between January and April 2017, and comprised all patients scheduled for cataract surgery who underwent scanning with IOLMaster 500. For each patient, pre-dilation and post-dilation measurements were taken. The intraocular lens power was determined through Sanders/Retzlaff/Kraff Theoretical, Holladay, and Hoffer Q formulae. Data was analysed using SPSS 24.
Results: There were 276 eyes of 138 participants who had a mean age of 59.7±11.1 years. Anterior chamber depth changed significantly with pupil dilation (p=0.001). No significant changes were observed in the axial length (p=0.410), keratometry measurements (p=0.931), and intraocular lens power calculations (p>0.05).
Conclusions: The change in anterior chamber depth, though significant, was perhaps clinically non-significant. intraocular lens power did not exhibit any significant differences post-mydriasis.
Key Words: Mydriasis, Lenses, Intraocular, Lens, Implantation, Phakic intraocular lenses, Biometry.
Pakistan Medical Association
Title: Effect of pupil dilation on IOLMaster® measurements in Pakistani Eyes
Description:
Objective: To analyse the effect of pupil dilation on intraocular lens instrument IOLMaster 500 biometric measurements, and to determine the effect of these measurements on intraocular lens power calculations in Asian eyes.
Methods: The prospective study was conducted at the Aga Khan University Hospital, Karachi, between January and April 2017, and comprised all patients scheduled for cataract surgery who underwent scanning with IOLMaster 500.
For each patient, pre-dilation and post-dilation measurements were taken.
The intraocular lens power was determined through Sanders/Retzlaff/Kraff Theoretical, Holladay, and Hoffer Q formulae.
Data was analysed using SPSS 24.
Results: There were 276 eyes of 138 participants who had a mean age of 59.
7±11.
1 years.
Anterior chamber depth changed significantly with pupil dilation (p=0.
001).
No significant changes were observed in the axial length (p=0.
410), keratometry measurements (p=0.
931), and intraocular lens power calculations (p>0.
05).
Conclusions: The change in anterior chamber depth, though significant, was perhaps clinically non-significant.
intraocular lens power did not exhibit any significant differences post-mydriasis.
Key Words: Mydriasis, Lenses, Intraocular, Lens, Implantation, Phakic intraocular lenses, Biometry.
Related Results
Comparison of four formulas for the intraocular lens power prediction
Comparison of four formulas for the intraocular lens power prediction
Abstract
Purpose
To assess the accuracy of four formulas for intraocular lens (IOL) power prediction in cataractous eyes.
METHODS
In this prospective study, 51 eyes of 3...
Examining pupil characteristics and relative afferent pupil defect prevalence in Syrian patients
Examining pupil characteristics and relative afferent pupil defect prevalence in Syrian patients
Purpose:
This study aims to explore pupil characteristics in a cohort of Syrian patients across different age groups and under varying luminance conditions, bot...
Exploring Language Features of Male and Female Speakers in Pakistani TEDx Talks: A Corpus-based Comparative Analysis
Exploring Language Features of Male and Female Speakers in Pakistani TEDx Talks: A Corpus-based Comparative Analysis
The study explores the linguistic patterns in Pakistani TEDx Talks. It is based on gender-based language use. It consists of ten talks selected from YouTube and applies both quanti...
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Comparison of Toric Intraocular Lens Calculations Between the IOLMaster 700 and Pentacam AXL
Purpose: To compare toric IOL (Intra Ocular Lenses) suggestions for power, toric, and IOL alignment between the IOLMaster 700 and Pentacam AXL (Axial Length).
Methods: This single-...
Sympathetic Nervous System Does Not Mediate Reflex Pupillary Dilation during Desflurane Anesthesia
Sympathetic Nervous System Does Not Mediate Reflex Pupillary Dilation during Desflurane Anesthesia
Background
Pupil size is determined by an interaction between the sympathetic and parasympathetic divisions of the autonomic nervous system. Noxious stimulation dilates...
Pupil diameter tracked during motor adaptation in humans
Pupil diameter tracked during motor adaptation in humans
Abstract
Pupil diameter, under constant illumination, is known to reflect individuals’ internal states, such as surprise about observation and environmental uncerta...
Pupil responses to implied motion in figurative and abstract paintings
Pupil responses to implied motion in figurative and abstract paintings
Motion can be perceived in static images, such as photos and figurative paintings, representing realistic subjects in motion, with or without directional information (e.g., motion ...
Resistance exercising on unstable surface leads to Pupil Dilation
Resistance exercising on unstable surface leads to Pupil Dilation
Abstract
Background
Chronic resistance training and acute resistance exercises improve physical performance and can enhance cognitive performance. H...