SU‐D‐220‐06: Non‐Contact (Air‐Coupled) Ultrasound Applied to Cortical Bone Phantoms

View through CrossRef
Purpose: The purpose was to take the first steps towards applying Non‐Contact Ultrasound (NCU) to the in‐vivo monitoring of osteoporosis and to skeletal quantitative ultrasound imaging (QUS) using cortical bone. The goals of this study focused on the possible advantages of NCU such as its lack of reliance on a technologist and a layer of acoustical coupling gel, its ability to use methods similar to those in x‐ray imaging, and the likely reduction in statistical and systematic errors. This project pursued additional applications of NCU beyond its usage in assessing the severity of third degree burns. Methods: An NCU imaging system, a pair of broadband, 1.5 MHz, non‐contact transducers, and cortical bone phantoms were used to determine bone mineral density (BMD), speed of sound (SOS), integrated acoustical response (IR), and ultrasonic transmittance. Air gaps of greater than 3 cm, two transmission and reflection paths, and a digital signal processor were used to collect data from phantoms of nominal mass density and (BMD) from 1.17 g/cm3 to 2.25 g/cm3 and from 0 g/cm3 to 1.7 g/cm3. Results: Good correlations between known BMD and measured SOS, IR, and transmittance were obtained for all 14 phantoms. Repeated measurements of the speed of sound (SOS), thickness, and IR for the phantom set show a small variation of plus or minus 1–2 %. These NCU data were shown to be in agreement with similar results from contact ultrasound to within 1–2%. Attenuation images of cortical bone phantoms showed discernible differences as a function of BMD. Conclusions: The results suggest that NCU might find additional applications beyond burn assessment. Since the NCU and conventional gel‐coupled measurements yield similar results, our air‐coupled methods might be able to reach levels of accuracy close to those attained with gel, while avoiding the disadvantages of contact.The authors would like to express their gratitude to the National Institutes of Health who supported this research through Grant Number GM 08156‐22 and its supplements which included funding for obtaining an NCU system and other ultrasound hardware used in this study.
Title: SU‐D‐220‐06: Non‐Contact (Air‐Coupled) Ultrasound Applied to Cortical Bone Phantoms
Description:
Purpose: The purpose was to take the first steps towards applying Non‐Contact Ultrasound (NCU) to the in‐vivo monitoring of osteoporosis and to skeletal quantitative ultrasound imaging (QUS) using cortical bone.
The goals of this study focused on the possible advantages of NCU such as its lack of reliance on a technologist and a layer of acoustical coupling gel, its ability to use methods similar to those in x‐ray imaging, and the likely reduction in statistical and systematic errors.
This project pursued additional applications of NCU beyond its usage in assessing the severity of third degree burns.
Methods: An NCU imaging system, a pair of broadband, 1.
5 MHz, non‐contact transducers, and cortical bone phantoms were used to determine bone mineral density (BMD), speed of sound (SOS), integrated acoustical response (IR), and ultrasonic transmittance.
Air gaps of greater than 3 cm, two transmission and reflection paths, and a digital signal processor were used to collect data from phantoms of nominal mass density and (BMD) from 1.
17 g/cm3 to 2.
25 g/cm3 and from 0 g/cm3 to 1.
7 g/cm3.
Results: Good correlations between known BMD and measured SOS, IR, and transmittance were obtained for all 14 phantoms.
Repeated measurements of the speed of sound (SOS), thickness, and IR for the phantom set show a small variation of plus or minus 1–2 %.
These NCU data were shown to be in agreement with similar results from contact ultrasound to within 1–2%.
Attenuation images of cortical bone phantoms showed discernible differences as a function of BMD.
Conclusions: The results suggest that NCU might find additional applications beyond burn assessment.
Since the NCU and conventional gel‐coupled measurements yield similar results, our air‐coupled methods might be able to reach levels of accuracy close to those attained with gel, while avoiding the disadvantages of contact.
The authors would like to express their gratitude to the National Institutes of Health who supported this research through Grant Number GM 08156‐22 and its supplements which included funding for obtaining an NCU system and other ultrasound hardware used in this study.

Related Results

Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Frequency of Common Chromosomal Abnormalities in Patients with Idiopathic Acquired Aplastic Anemia
Objective: To determine the frequency of common chromosomal aberrations in local population idiopathic determine the frequency of common chromosomal aberrations in local population...
Kauppis-Heikin humoristisista tyylikeinoista
Kauppis-Heikin humoristisista tyylikeinoista
Kielenaineksetarkintähti (kieli: suomi, sivulla: 221)asessyöri (kieli: suomi, sivulla: 221)filosovia (kieli: suomi, sivulla: 221)humpauttaa (kieli: suomi, sivulla: 220)hurpattaa (k...
Noncontact ultrasound imaging applied to cortical bone phantoms
Noncontact ultrasound imaging applied to cortical bone phantoms
Purpose:The purpose of this paper was to take the first steps toward applying noncontact ultrasound (NCU) to the tasks of monitoring osteoporosis and quantitative ultrasound imagin...
Poster 107: The Use of Coacervate Sustained Release System to Identify the Most Potent BMP for Bone Regeneration
Poster 107: The Use of Coacervate Sustained Release System to Identify the Most Potent BMP for Bone Regeneration
Objectives: Bone morphogenetic proteins (BMPs) belong to the transforming growth factor superfamily that were first discovered by Marshall Urist. There are 14 B...
SU‐FF‐I‐26: Tissue Equivalent Phantoms for the Evaluation of Tube Current Modulated CT Dose and Image Quality
SU‐FF‐I‐26: Tissue Equivalent Phantoms for the Evaluation of Tube Current Modulated CT Dose and Image Quality
Purpose: To develop and test new, flexible, tissue‐equivalent phantoms for the evaluation of tube current modulation dose reduction and image quality in CT. The developed phantom m...
Photothermal Response of Tissue Phantoms Containing Multi-Walled Carbon Nanotubes
Photothermal Response of Tissue Phantoms Containing Multi-Walled Carbon Nanotubes
Inclusion of multi-walled carbon nanotubes (MWNTs) into tissue prior to laser therapy has the potential to enhance the selectivity and effectiveness of cancer therapy by providing ...

Back to Top