Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
View through CrossRef
Abstract
In mammography imaging, the most common target material used in mammography machines are Molybdenum (Mo) target with Mo filters. In the publication of International Electrotechnical Commission (IEC) 61267, the establishment of mammography beam qualities assembly are given in 3 fixed parameters namely the target material which is Mo, the Mo filter thickness and the tube voltage. While, Mo target with rhodium (Rh) filters combination standard beam qualities were established by the Physikalish-Technische Bundesanstalt (PTB) primary standard Laboratory. The objective of this study is to measure the kVp and HVL ranging from 23-31 kV. Later compare the displayed kilovoltage peak (kVp) on the control panel with the kVp measured using a non-invasive kVp meter and validating them with the HVL measurement. From the results, the kVp dialed and kVp measured shows significant difference by 2-4 kV. The HVL measured for the kVp before and after corrected also shows similar pattern, where the HVL after kVp is correct is almost identical to the standard beam qualities with the difference of 1%. This verifies that the kVp measurement using a non-invasive kVp meter can be validated with just a simple measurement of its HVL for mammography standard beam qualities.
Title: Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Description:
Abstract
In mammography imaging, the most common target material used in mammography machines are Molybdenum (Mo) target with Mo filters.
In the publication of International Electrotechnical Commission (IEC) 61267, the establishment of mammography beam qualities assembly are given in 3 fixed parameters namely the target material which is Mo, the Mo filter thickness and the tube voltage.
While, Mo target with rhodium (Rh) filters combination standard beam qualities were established by the Physikalish-Technische Bundesanstalt (PTB) primary standard Laboratory.
The objective of this study is to measure the kVp and HVL ranging from 23-31 kV.
Later compare the displayed kilovoltage peak (kVp) on the control panel with the kVp measured using a non-invasive kVp meter and validating them with the HVL measurement.
From the results, the kVp dialed and kVp measured shows significant difference by 2-4 kV.
The HVL measured for the kVp before and after corrected also shows similar pattern, where the HVL after kVp is correct is almost identical to the standard beam qualities with the difference of 1%.
This verifies that the kVp measurement using a non-invasive kVp meter can be validated with just a simple measurement of its HVL for mammography standard beam qualities.
Related Results
Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Abstract
Background
HVL in diagnostic X-ray machine is an important property that is used to define the penetrating ability of an X-ray beam and hence can be used to deter...
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray mul...
MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
Purpose: To show that the nominal peak tube voltage potential (kVp) and measured half‐value layer (HVL) are sufficient to generate energy spectra and fluence profiles for fast and ...
Estimasi Resiko Radiasi Janin pada Pemeriksaan Radiografi Pelvis
Estimasi Resiko Radiasi Janin pada Pemeriksaan Radiografi Pelvis
Sinar-X merupakan sumber radiasi pengion yang paling banyak digunakan untuk pemeriksaan diagnostik dalam aplikasi klinik. Radiasi pengion memang memberikan manfaat yang besar dalam...
Abstract IA15: Neighborhood factors and changes in mammography screening rates in response to promotora intervention in Latinas
Abstract IA15: Neighborhood factors and changes in mammography screening rates in response to promotora intervention in Latinas
Abstract
Latinas have relatively low rates of mammography screening participation, which may contribute to later stage of breast cancer at diagnosis for Latinas, and...
Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique
Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique
Introduction: Computed Tomography Angiography (CTA) of the cerebral carotid arteries is a key non-invasive modality for evaluating vascular abnormalities. Image quality and radiati...
Cost–benefit analysis of mammography screening: the perspective of Nigerian healthcare providers
Cost–benefit analysis of mammography screening: the perspective of Nigerian healthcare providers
Introduction
There is no national breast cancer screening implemented in Nigeria. The National Cancer Control Plan (NCCP) has a goal of making screening services ...
Voltage Dependent Effect of Spiral Wound Plasma Discharge on DBC1.2 Cellular Integrity
Voltage Dependent Effect of Spiral Wound Plasma Discharge on DBC1.2 Cellular Integrity
Low temperature plasmas (LTPs) generated at atmospheric pressure and room temperature have gained increasing attention in biomedical research due to their ability to control cellul...