MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
View through CrossRef
Purpose: To show that the nominal peak tube voltage potential (kVp) and measured half‐value layer (HVL) are sufficient to generate energy spectra and fluence profiles for fast and accurate machine‐specific dose computations in Computed Tomography (CT). Methods: Spatial variation of the x‐ray source spectrum was found by measuring HVL across the internal bowtie filter axis and using the nominal kVp settings and third‐party software Spektr to generate the spectra. The beam fluence was calculated by multiplying in‐air dose measurements along the filter axis with the integral product of the spectra and the in‐air NIST mass‐energy attenuation coefficients. Dose calculations were performed using a previously validated in‐house hybrid deterministic and stochastic kV x‐ray dose computation algorithm (kVDoseCalc). To ensure dose convergence while minimizing calculation time, we examined the sensitivity of kVDoseCalc to the number of photons seeded. We modeled the source of a Philips Brilliance Big Bore CT scanner for 90, 120, and 140 kVp settings. Doses measured using a Farmer‐type Capintec ion chamber (0.6 cc) placed in a cylindrical poly methyl methacrylate (PMMA) phantom were compared to those computed with kVDoseCalc. Results: The number of photons seeded required to keep the average statistical uncertainty in dose less than .1% was found to be 1.25 million. The average percent difference between calculation and measurement pooled over all 12 positions in the phantom was found to be 1.68%, 1.60%, and 1.25% for 90, 120, and 140 kVp, respectively. The maximum percent difference between calculation and measurement was less than 3.64% pooled over all energies and measurement positions. Thirty‐one out of a total of 36 simulation conditions were within the experimental uncertainties associated with measurement reproducibility and chamber volume effects. Conclusion: Our source characterization technique, which derives incident fluence and spectra from measurements of HVL across the bowtie profile, is sufficient for accurate machine‐specific CT dose computations.
Title: MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
Description:
Purpose: To show that the nominal peak tube voltage potential (kVp) and measured half‐value layer (HVL) are sufficient to generate energy spectra and fluence profiles for fast and accurate machine‐specific dose computations in Computed Tomography (CT).
Methods: Spatial variation of the x‐ray source spectrum was found by measuring HVL across the internal bowtie filter axis and using the nominal kVp settings and third‐party software Spektr to generate the spectra.
The beam fluence was calculated by multiplying in‐air dose measurements along the filter axis with the integral product of the spectra and the in‐air NIST mass‐energy attenuation coefficients.
Dose calculations were performed using a previously validated in‐house hybrid deterministic and stochastic kV x‐ray dose computation algorithm (kVDoseCalc).
To ensure dose convergence while minimizing calculation time, we examined the sensitivity of kVDoseCalc to the number of photons seeded.
We modeled the source of a Philips Brilliance Big Bore CT scanner for 90, 120, and 140 kVp settings.
Doses measured using a Farmer‐type Capintec ion chamber (0.
6 cc) placed in a cylindrical poly methyl methacrylate (PMMA) phantom were compared to those computed with kVDoseCalc.
Results: The number of photons seeded required to keep the average statistical uncertainty in dose less than .
1% was found to be 1.
25 million.
The average percent difference between calculation and measurement pooled over all 12 positions in the phantom was found to be 1.
68%, 1.
60%, and 1.
25% for 90, 120, and 140 kVp, respectively.
The maximum percent difference between calculation and measurement was less than 3.
64% pooled over all energies and measurement positions.
Thirty‐one out of a total of 36 simulation conditions were within the experimental uncertainties associated with measurement reproducibility and chamber volume effects.
Conclusion: Our source characterization technique, which derives incident fluence and spectra from measurements of HVL across the bowtie profile, is sufficient for accurate machine‐specific CT dose computations.
Related Results
Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Abstract
Background
HVL in diagnostic X-ray machine is an important property that is used to define the penetrating ability of an X-ray beam and hence can be used to deter...
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Abstract
In mammography imaging, the most common target material used in mammography machines are Molybdenum (Mo) target with Mo filters. In the publication of Inter...
Estimasi Resiko Radiasi Janin pada Pemeriksaan Radiografi Pelvis
Estimasi Resiko Radiasi Janin pada Pemeriksaan Radiografi Pelvis
Sinar-X merupakan sumber radiasi pengion yang paling banyak digunakan untuk pemeriksaan diagnostik dalam aplikasi klinik. Radiasi pengion memang memberikan manfaat yang besar dalam...
OPTIMIZATION OF RADIATION DOSE AND IMAGE QUALITY IN ABDOMINAL RADIOGRAPHY USING DIGITAL MOBILE X-RAY SYSTEM
OPTIMIZATION OF RADIATION DOSE AND IMAGE QUALITY IN ABDOMINAL RADIOGRAPHY USING DIGITAL MOBILE X-RAY SYSTEM
The abdominal radiography using digital mobile x-ray system is commonly requested for the immobilized patient. The selection of the exposure parameters needs the experience of the ...
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray mul...
Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique
Quantitative analysis of cerebral carotid CTA image quality and radiation dose at low kv and standard kv variations using bolus tracking technique
Introduction: Computed Tomography Angiography (CTA) of the cerebral carotid arteries is a key non-invasive modality for evaluating vascular abnormalities. Image quality and radiati...
Voltage Dependent Effect of Spiral Wound Plasma Discharge on DBC1.2 Cellular Integrity
Voltage Dependent Effect of Spiral Wound Plasma Discharge on DBC1.2 Cellular Integrity
Low temperature plasmas (LTPs) generated at atmospheric pressure and room temperature have gained increasing attention in biomedical research due to their ability to control cellul...
On the use of the absorbed depth‐dose measurements in the beam calibration of a surface electronic high‐dose‐rate brachytherapy unit, a Monte Carlo‐based study
On the use of the absorbed depth‐dose measurements in the beam calibration of a surface electronic high‐dose‐rate brachytherapy unit, a Monte Carlo‐based study
PurposeTo evaluate the use of the absorbed depth‐dose as a surrogate of the half‐value layer in the calibration of a high‐dose‐rate electronic brachytherapy (eBT) equipment. The ef...