Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography

View through CrossRef
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray multimeter, and aluminum filter. The exposure process is carried out at certain kVp and mAs conditions after placing the detector on the examination table. The HVL test to determine the X-ray beam quality is carried out by placing the detector 4 cm from the edge of the chest wall on the patient's table in an upright position facing the tube. The resulting HVL values ​​are 0.37, 0.37, and 0.38. This study resulted in a p-value of 0.000 (p < 0.05). Statistically, there is a significant difference in HVL values ​​in 2019-2021. The resulting HVL value is by the tolerance limit recommended by the IAEA, which is 0.31 ≤ HVL ≤ 0.58. These values ​​indicate that mammography works well, and it is safe for the patient.
Title: Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
Description:
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021.
This measurement uses a ruler, X-ray multimeter, and aluminum filter.
The exposure process is carried out at certain kVp and mAs conditions after placing the detector on the examination table.
The HVL test to determine the X-ray beam quality is carried out by placing the detector 4 cm from the edge of the chest wall on the patient's table in an upright position facing the tube.
The resulting HVL values ​​are 0.
37, 0.
37, and 0.
38.
This study resulted in a p-value of 0.
000 (p < 0.
05).
Statistically, there is a significant difference in HVL values ​​in 2019-2021.
The resulting HVL value is by the tolerance limit recommended by the IAEA, which is 0.
31 ≤ HVL ≤ 0.
58.
These values ​​indicate that mammography works well, and it is safe for the patient.

Related Results

Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Abstract Background HVL in diagnostic X-ray machine is an important property that is used to define the penetrating ability of an X-ray beam and hence can be used to deter...
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Abstract In mammography imaging, the most common target material used in mammography machines are Molybdenum (Mo) target with Mo filters. In the publication of Inter...
Abstract IA15: Neighborhood factors and changes in mammography screening rates in response to promotora intervention in Latinas
Abstract IA15: Neighborhood factors and changes in mammography screening rates in response to promotora intervention in Latinas
Abstract Latinas have relatively low rates of mammography screening participation, which may contribute to later stage of breast cancer at diagnosis for Latinas, and...
Access Denied
Access Denied
Introduction As social-distancing mandates in response to COVID-19 restricted in-person data collection methods such as participant observation and interviews, researchers turned t...
Cost–benefit analysis of mammography screening: the perspective of Nigerian healthcare providers
Cost–benefit analysis of mammography screening: the perspective of Nigerian healthcare providers
Introduction There is no national breast cancer screening implemented in Nigeria. The National Cancer Control Plan (NCCP) has a goal of making screening services ...
Abstract P5-01-02: Glandular dose in contrast-enhanced dual-energy mammography
Abstract P5-01-02: Glandular dose in contrast-enhanced dual-energy mammography
Abstract INTRODUCTION Estimation of the Average Glandular Dose (AGD) absorbed by the breast during x-ray based examination is an established part of q...
Intervention to Improve Mammography Screening at a Student-Run Free Clinic
Intervention to Improve Mammography Screening at a Student-Run Free Clinic
Background: Most late-stage breast cancer occurs in those who have never been screened. Uninsured women have lower screening rates than insured women. Student-run free clinics (SRF...

Back to Top