Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
View through CrossRef
Abstract
Background
HVL in diagnostic X-ray machine is an important property that is used to define the penetrating ability of an X-ray beam and hence can be used to determine the component of beam hardness.
Purpose
The purpose of the study was to assess the effect of HVL on the dose output of diagnostic X-ray machine and to determine the different corrective actions for the cases which did not comply the established dose limits.
Material and Methods
The study utilized the findings of the radiation safety inspections conducted between 2021–2022 by Atomic Energy Council to investigate different Half value layer (HVL) ranges for diagnostic X-ray machines. A number of 64 X-ray machines were selected for the study based on a set criterion.
Results
A total of 26 X-ray machines failed the HVL test though produced different descriptions of measured dose output. Fifteen (15) X-ray machines produced doses within the permissible range of 0.025 mGy/mAs – 0.080 mGy/mAs, eight (08) X-ray machines produced doses below the lower limit of the dose range of 0.025 mGy/mAs, while three (03) X-ray machines produce doses above the upper limit of the dose range of 0.080 mGy/mAs. The other 38 X-ray machines passed the HVL test but failed the dose output test. These were classified in different HVL groups that exceeded the recommended regulatory limit, that is, moderate (5), high (19), very high (10) and extreme high (4) X-ray machines. The corrective action made was either adding or removing filter plates to enhance or minimize the filtration for machines that failed the HVL test and the ones that passed HVL test but failed the dose output test respectively. However, this should only be after a conclusive investigation of checking the accuracy of tube current (mA), timer (s) and tube potential (kVp) parameters.
Conclusion
Therefore, the corrective action for HVL should not be done in isolation from other machine dose contributing parameters like tube current potential parameters and exposure time. The regulatory body recommends that HVL tests should be part of the acceptance and commissioning tests for the new machines and done routinely for the machines in use as specified in the quality control program for each facility.
Title: Effect of varying HVL values on dose output of plain X-ray machines at a fixed kV of 80.
Description:
Abstract
Background
HVL in diagnostic X-ray machine is an important property that is used to define the penetrating ability of an X-ray beam and hence can be used to determine the component of beam hardness.
Purpose
The purpose of the study was to assess the effect of HVL on the dose output of diagnostic X-ray machine and to determine the different corrective actions for the cases which did not comply the established dose limits.
Material and Methods
The study utilized the findings of the radiation safety inspections conducted between 2021–2022 by Atomic Energy Council to investigate different Half value layer (HVL) ranges for diagnostic X-ray machines.
A number of 64 X-ray machines were selected for the study based on a set criterion.
Results
A total of 26 X-ray machines failed the HVL test though produced different descriptions of measured dose output.
Fifteen (15) X-ray machines produced doses within the permissible range of 0.
025 mGy/mAs – 0.
080 mGy/mAs, eight (08) X-ray machines produced doses below the lower limit of the dose range of 0.
025 mGy/mAs, while three (03) X-ray machines produce doses above the upper limit of the dose range of 0.
080 mGy/mAs.
The other 38 X-ray machines passed the HVL test but failed the dose output test.
These were classified in different HVL groups that exceeded the recommended regulatory limit, that is, moderate (5), high (19), very high (10) and extreme high (4) X-ray machines.
The corrective action made was either adding or removing filter plates to enhance or minimize the filtration for machines that failed the HVL test and the ones that passed HVL test but failed the dose output test respectively.
However, this should only be after a conclusive investigation of checking the accuracy of tube current (mA), timer (s) and tube potential (kVp) parameters.
Conclusion
Therefore, the corrective action for HVL should not be done in isolation from other machine dose contributing parameters like tube current potential parameters and exposure time.
The regulatory body recommends that HVL tests should be part of the acceptance and commissioning tests for the new machines and done routinely for the machines in use as specified in the quality control program for each facility.
Related Results
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray mul...
On the use of the absorbed depth‐dose measurements in the beam calibration of a surface electronic high‐dose‐rate brachytherapy unit, a Monte Carlo‐based study
On the use of the absorbed depth‐dose measurements in the beam calibration of a surface electronic high‐dose‐rate brachytherapy unit, a Monte Carlo‐based study
PurposeTo evaluate the use of the absorbed depth‐dose as a surrogate of the half‐value layer in the calibration of a high‐dose‐rate electronic brachytherapy (eBT) equipment. The ef...
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Validation of kVp measurement using the HVL approach for standard beam qualities for mammography
Abstract
In mammography imaging, the most common target material used in mammography machines are Molybdenum (Mo) target with Mo filters. In the publication of Inter...
MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
MO‐D‐134‐10: Using HVL and KVp to Portray An X‐Ray Source for Dose Calculations in CT
Purpose: To show that the nominal peak tube voltage potential (kVp) and measured half‐value layer (HVL) are sufficient to generate energy spectra and fluence profiles for fast and ...
Optimizing Hydroxyurea Dosing in Sickle Cell Anemia: The Uganda MTD Study
Optimizing Hydroxyurea Dosing in Sickle Cell Anemia: The Uganda MTD Study
Introduction. Hydroxyurea treatment has proven safety, feasibility, and efficacy for children with sickle cell anemia living in sub-Saharan Africa. Even in malaria endemic regions,...
P-598 Dose accuracy of the follitropin-alfa, follitropin-alfa/lutropin-alfa and choriogonadotropin-alfa pen injectors used for fertility treatment
P-598 Dose accuracy of the follitropin-alfa, follitropin-alfa/lutropin-alfa and choriogonadotropin-alfa pen injectors used for fertility treatment
Abstract
Study question
Do the Merck family of fertility pen injectors (follitropin-alfa, follitropin-alfa/lutropin-alfa, chorio...
The Plain Packaging of Tobacco Products
The Plain Packaging of Tobacco Products
The Plain Packaging of Tobacco ProductsThis special edition of the QUT Law Review considers the international debate over the introduction of Australia's pioneering plain packaging...
INFINITESIMAL FIELD SIZE HVL OF ISO 4037 NARROW SPECTRUM X-RAYS: SIMULATED RESULTS
INFINITESIMAL FIELD SIZE HVL OF ISO 4037 NARROW SPECTRUM X-RAYS: SIMULATED RESULTS
Half-value layer (HVL) measurement is the method most often used to validate an X-ray reference field as defined in ISO 4037. This work presents Particle and Heavy Ion Transport co...