Comparison of three newer generation freely available intraocular lens power calculation formulae across all axial lengths
View through CrossRef
Purpose:
The aim of this study is to evaluate the accuracy of three newer generation formulae (Barrett Universal II, EVO, Hill-RBF 2.0) for calculation of power of two standard IOLs—the Acrysof IQ and Tecnis ZCB00 across all axial lengths.
Methods:
In this retrospective series, 206 eyes of 206 patients, operated for cataract surgery with above two IOLs over the last 6 months, were included in the study. Preoperative biometry measurements were obtained from LenstarLS900. By using recommended lens constants, the mean error for each formula was calculated and compared. Then, the optimized IOL constants were calculated to reduce the mean error to zero. Mean and median absolute errors were calculated for all eyes and separately for short (AL<22.5 mm), medium (22.5–24.5 mm), and long eyes (>24.5 mm). Absolute errors and percentages of eyes within prediction errors of ±0.25 D, ±0.50 D, ±0.75 D, and ±1.00 D were compared.
Results:
Prediction error with using recommended lens constants was significantly lower in the Barrett Universal II formula as compared to the other two formulae. However, after optimizing lens constants, there were no significant differences in the absolute errors between the three formulae. The formulae ranked by mean absolute error were as follows: Barrett Universal II (0.304 D), EVO (0.317 D), and Hill-RBF (0.322) D. There were no significant differences between absolute errors in the three formulae in each of the short-, medium-, and long-eye subgroups.
Conclusion:
With proper lens constant optimization, the Barrett Universal II, EVO, and Hill-RBF 2.0 formulae were equally accurate in predicting IOL power across the entire range of axial lengths.
Title: Comparison of three newer generation freely available intraocular lens power calculation formulae across all axial lengths
Description:
Purpose:
The aim of this study is to evaluate the accuracy of three newer generation formulae (Barrett Universal II, EVO, Hill-RBF 2.
0) for calculation of power of two standard IOLs—the Acrysof IQ and Tecnis ZCB00 across all axial lengths.
Methods:
In this retrospective series, 206 eyes of 206 patients, operated for cataract surgery with above two IOLs over the last 6 months, were included in the study.
Preoperative biometry measurements were obtained from LenstarLS900.
By using recommended lens constants, the mean error for each formula was calculated and compared.
Then, the optimized IOL constants were calculated to reduce the mean error to zero.
Mean and median absolute errors were calculated for all eyes and separately for short (AL<22.
5 mm), medium (22.
5–24.
5 mm), and long eyes (>24.
5 mm).
Absolute errors and percentages of eyes within prediction errors of ±0.
25 D, ±0.
50 D, ±0.
75 D, and ±1.
00 D were compared.
Results:
Prediction error with using recommended lens constants was significantly lower in the Barrett Universal II formula as compared to the other two formulae.
However, after optimizing lens constants, there were no significant differences in the absolute errors between the three formulae.
The formulae ranked by mean absolute error were as follows: Barrett Universal II (0.
304 D), EVO (0.
317 D), and Hill-RBF (0.
322) D.
There were no significant differences between absolute errors in the three formulae in each of the short-, medium-, and long-eye subgroups.
Conclusion:
With proper lens constant optimization, the Barrett Universal II, EVO, and Hill-RBF 2.
0 formulae were equally accurate in predicting IOL power across the entire range of axial lengths.
Related Results
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
Effect of pupil dilation on IOLMaster® measurements in Pakistani Eyes
Effect of pupil dilation on IOLMaster® measurements in Pakistani Eyes
Objective: To analyse the effect of pupil dilation on intraocular lens instrument IOLMaster 500 biometric measurements, and to determine the effect of these measurements on intraoc...
Study on the myopia control effect of OK lens on children with different corneal curvature
Study on the myopia control effect of OK lens on children with different corneal curvature
Abstract
Objective: To explore the effect of OK lens on myopia control in children with different corneal curvature.
Method: A total of 178 myopic children admitted to our...
ANALYSIS OF THE OPERATION MODE OF THE SOLAR POWER PLANT
ANALYSIS OF THE OPERATION MODE OF THE SOLAR POWER PLANT
The article examines the load change schedule of the solar power plant in the Ukraine-Moldova energy union. The analysis of data averaged at minute and 15-minute intervals in the p...
Secondary Glaucoma in Intraocular Tumors
Secondary Glaucoma in Intraocular Tumors
The most common malignant intraocular tumors in adults include uveal melanoma, intraocular metastatic tumors, leukemic infiltrates, and intraocular lymphoma.
</br>Elevated in...
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Purpose:
To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL ...
Twenty-Seven-Gauge Needle-Assisted Technique for Repositioning of Late Luxated or Subluxated Capsular Bag-Intraocular Lens Complex: Inverted Scleral Tunnel for Flanged Polypropylene Suture Technique
Twenty-Seven-Gauge Needle-Assisted Technique for Repositioning of Late Luxated or Subluxated Capsular Bag-Intraocular Lens Complex: Inverted Scleral Tunnel for Flanged Polypropylene Suture Technique
Abstract
Purpose: To describe a minimally invasive and atraumatic technique for managing a polypropylene suture-assisted scleral fixation intraocular lens-capsular bag comp...
Intraocular and periocular lymphoma in dogs and cats: a retrospective review of 21 cases (2001–2012)
Intraocular and periocular lymphoma in dogs and cats: a retrospective review of 21 cases (2001–2012)
AbstractObjectiveTo immunologically phenotype and histologically classify canine and feline intraocular and periocular lymphomas.MethodsThe databases of four veterinary medical dia...