Validity of non-invasive hemoglobin measured by pulse co-oximeter in neonates - An observational study
View through CrossRef
Background: Hemoglobin is a frequently ordered investigation in neonatal intensive care units. There is a need for hemoglobin estimation by point of care methods to reduce iatrogenic anemia and to alleviate pain associated with repeated venous sampling. Pulse co-oximeters have been developed to measure hemoglobin non-invasively based on spectrophotometric method. We compared hemoglobin measured by pulse co-oximeter with reference venous hemoglobin in neonates admitted to a tertiary care newborn unit. Design: This was an observational study. Duration: The study duration was from November 2016 to December 2016. Setting: Department of Neonatology, Institute of Child Health and Hospital for Children, Egmore. Methods: Neonates admitted in nursery who warranted hemoglobin estimation underwent both invasive venous hemoglobin estimation by automated hematology analyzer and non-invasive hemoglobin estimation by pulse co-oximeter (Masimo SET radical 7). Results: Of 158 newborns enrolled, the device failure rate was 12.5%. The bias between transcutaneous and venous hemoglobin was 1.66±2.26 g/dl (mean ± standard deviation). Transcutaneous and venous hemoglobin showed moderate agreement on Bland Altman plot with intraclass correlation coefficient of 0.56. At lower levels of hemoglobin, we noted higher bias. It was 2.69±1.87 g/dl at hemoglobin <13 g/dl and 3.29±1.86 at hemoglobin ?10 g/dl. On regression analysis, only the level of hemoglobin influenced bias and device failure rate. Conclusion: Non-invasive hemoglobin measured by pulse co-oximeter shows only a moderate agreement with reference venous hemoglobin in neonates admitted to nursery. We report a high device failure rate of 12.6%. Level of hemoglobin is the single most determinant of device failure and degree of agreement. With high device failure rates and poor agreement at low hemoglobin levels, the clinical utility appears negligible.
Mansa STM Publishers
Title: Validity of non-invasive hemoglobin measured by pulse co-oximeter in neonates - An observational study
Description:
Background: Hemoglobin is a frequently ordered investigation in neonatal intensive care units.
There is a need for hemoglobin estimation by point of care methods to reduce iatrogenic anemia and to alleviate pain associated with repeated venous sampling.
Pulse co-oximeters have been developed to measure hemoglobin non-invasively based on spectrophotometric method.
We compared hemoglobin measured by pulse co-oximeter with reference venous hemoglobin in neonates admitted to a tertiary care newborn unit.
Design: This was an observational study.
Duration: The study duration was from November 2016 to December 2016.
Setting: Department of Neonatology, Institute of Child Health and Hospital for Children, Egmore.
Methods: Neonates admitted in nursery who warranted hemoglobin estimation underwent both invasive venous hemoglobin estimation by automated hematology analyzer and non-invasive hemoglobin estimation by pulse co-oximeter (Masimo SET radical 7).
Results: Of 158 newborns enrolled, the device failure rate was 12.
5%.
The bias between transcutaneous and venous hemoglobin was 1.
66±2.
26 g/dl (mean ± standard deviation).
Transcutaneous and venous hemoglobin showed moderate agreement on Bland Altman plot with intraclass correlation coefficient of 0.
56.
At lower levels of hemoglobin, we noted higher bias.
It was 2.
69±1.
87 g/dl at hemoglobin <13 g/dl and 3.
29±1.
86 at hemoglobin ?10 g/dl.
On regression analysis, only the level of hemoglobin influenced bias and device failure rate.
Conclusion: Non-invasive hemoglobin measured by pulse co-oximeter shows only a moderate agreement with reference venous hemoglobin in neonates admitted to nursery.
We report a high device failure rate of 12.
6%.
Level of hemoglobin is the single most determinant of device failure and degree of agreement.
With high device failure rates and poor agreement at low hemoglobin levels, the clinical utility appears negligible.
Related Results
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
BACKGROUND: Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients. Pulse oximetry is commonly used to measure blood oxygen satur...
USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
Context is due to the need to develop a methodology for optimising the parameters of pulse oximeter sensor settings based on the practical determination of the accuracy of heart ra...
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
Purpose: Does a non-invasive pulse oximeter probe accurately indicate limb ischemia with inflation of a tourniquet? How are the pulse waveform, perfusion index and oxygen saturatio...
KADAR HEMOGLOBIN PADA ORANG DEWASA YANG TINGGAL DI DATARAN TINGGI DENGAN KETINGGIAN YANG BERBEDA
KADAR HEMOGLOBIN PADA ORANG DEWASA YANG TINGGAL DI DATARAN TINGGI DENGAN KETINGGIAN YANG BERBEDA
Abstract: Background: High altitude has an impact on the levels of hemoglobin in an individual. Altitude will cause hypoxia due to the partial pressure of oxygen decreases and body...
COMPARISON OF BLOOD GLUCOSE LEVEL IN PRETERM AND FULL-TERM BABIES IN 1ST 48 HOURS OF LIFE IN A TERTIARY CARE HOSPITAL
COMPARISON OF BLOOD GLUCOSE LEVEL IN PRETERM AND FULL-TERM BABIES IN 1ST 48 HOURS OF LIFE IN A TERTIARY CARE HOSPITAL
Background: Neonatal glucose regulation is crucial for early metabolic adaptation, particularly in the first 48 hours of life when the transition from intrauterine to extrauterine ...
Does Gender Affect Levels of Hyperbilirubinemia in Term Neonates
Does Gender Affect Levels of Hyperbilirubinemia in Term Neonates
Introduction: Hyperbilirubinemia is a common & in most1cases, benign problem in1first month of1life which is often1physiologic & intervention is not1usually1necessary. In t...
Umbilical serum concentrations of chemokines (RANTES and MGSA/GRO‐α) in preterm and term neonates
Umbilical serum concentrations of chemokines (RANTES and MGSA/GRO‐α) in preterm and term neonates
AbstractBackground: The objective of this study was to explore the relationship between labor (preterm and term) and umbilical blood serum regulated on activation, normal T cell e...
Assessment of vascular indices by abdominal aortic ultrasonography in preterm neonates with bronchopulmonary dysplasia
Assessment of vascular indices by abdominal aortic ultrasonography in preterm neonates with bronchopulmonary dysplasia
Abstract
Background
Preterm infants with bronchopulmonary dysplasia (BPD) often experience systemic hypertension, but the exact cause is not yet known. Since there have be...