Intermittent pulse oximeter as a measurement of newborn oxygen: a cross-sectional study
View through CrossRef
Introduction: Intermittent monitoring using fingertip pulse oximeters is increasingly being used in guiding oxygen therapy and screening congenital heart defects in newborns. However, there are no evidence-based guidelines for using fingertip pulse oximeters for intermittent monitoring of newborns in low-resource settings. This research assesses the validity of the fingertip pulse oximeter to measure oxygen saturation in newborns against the continuous monitoring pulse oximeter as a standard method.
Method: A prospective study was conducted in the NICU and Neonatal Intermediate Unit of Dr. Soetomo Academic Hospital Surabaya. Newborns aged below 28 days using a respiratory device were included. Oxygen saturation measurements using a continuous monitoring pulse oximeter and intermittent pulse oximeter were done on the right hand and both feet. Paired t-test and Bland-Altman plot was performed to assess the validity.
Results: Eighty-five neonates were included. There is an agreement between Intermittent and continuous pulse oximeters in the normoxic group with a limit of agreement -1.29 (CI 95% -1.62 to -0.95) to 1.29 (CI 95% 0.95 to 1.62) and hypoxic group with a limit of agreement -3.36 (CI 95% -4.29 to -2.44) to 3.00 (CI 95% 2.08 to 3.92). Intermittent pulse oximeter measurement requires a longer time for maximum results compared to continuous pulse oximeter (p=0.001).
Conclusions: The intermittent pulse oximeter is available to be used alternately with a continuous pulse oximeter. However, it has limitations in hypoxemia settings, hence the application in real-time clinical practice needs further study.
Title: Intermittent pulse oximeter as a measurement of newborn oxygen: a cross-sectional study
Description:
Introduction: Intermittent monitoring using fingertip pulse oximeters is increasingly being used in guiding oxygen therapy and screening congenital heart defects in newborns.
However, there are no evidence-based guidelines for using fingertip pulse oximeters for intermittent monitoring of newborns in low-resource settings.
This research assesses the validity of the fingertip pulse oximeter to measure oxygen saturation in newborns against the continuous monitoring pulse oximeter as a standard method.
Method: A prospective study was conducted in the NICU and Neonatal Intermediate Unit of Dr.
Soetomo Academic Hospital Surabaya.
Newborns aged below 28 days using a respiratory device were included.
Oxygen saturation measurements using a continuous monitoring pulse oximeter and intermittent pulse oximeter were done on the right hand and both feet.
Paired t-test and Bland-Altman plot was performed to assess the validity.
Results: Eighty-five neonates were included.
There is an agreement between Intermittent and continuous pulse oximeters in the normoxic group with a limit of agreement -1.
29 (CI 95% -1.
62 to -0.
95) to 1.
29 (CI 95% 0.
95 to 1.
62) and hypoxic group with a limit of agreement -3.
36 (CI 95% -4.
29 to -2.
44) to 3.
00 (CI 95% 2.
08 to 3.
92).
Intermittent pulse oximeter measurement requires a longer time for maximum results compared to continuous pulse oximeter (p=0.
001).
Conclusions: The intermittent pulse oximeter is available to be used alternately with a continuous pulse oximeter.
However, it has limitations in hypoxemia settings, hence the application in real-time clinical practice needs further study.
Related Results
USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
Context is due to the need to develop a methodology for optimising the parameters of pulse oximeter sensor settings based on the practical determination of the accuracy of heart ra...
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
BACKGROUND: Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients. Pulse oximetry is commonly used to measure blood oxygen satur...
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
Purpose: Does a non-invasive pulse oximeter probe accurately indicate limb ischemia with inflation of a tourniquet? How are the pulse waveform, perfusion index and oxygen saturatio...
High Concentration Oxygen and Hypercapnia in Respiratory Disease
High Concentration Oxygen and Hypercapnia in Respiratory Disease
<p>Oxygen-induced elevations in arterial carbon dioxide tension have been demonstrated in patients with chronic obstructive pulmonary disease (COPD), asthma, pneumonia, obesi...
Survey of Pediatrician Practices in Retrieving Statewide Authorized Newborn Screening Results
Survey of Pediatrician Practices in Retrieving Statewide Authorized Newborn Screening Results
Objective.
Mandated state newborn screening programs for the approximately 4 million infants born each year in the United States involves the following 5 componen...
Accuracy of the Ohmeda 3700 Pulse Oximeter in Estimating Heart Rate during Exercise
Accuracy of the Ohmeda 3700 Pulse Oximeter in Estimating Heart Rate during Exercise
Pulse oximeters are often used as a means of estimating heart rate (HR) at rest and during exercise. However, at present, few data have been reported to support the validity of thi...
Pengaruh Puasa Intermiten Kering terhadap Kadar Kolesterol pada Mencit yang Diberi Pakan Tinggi Lemak
Pengaruh Puasa Intermiten Kering terhadap Kadar Kolesterol pada Mencit yang Diberi Pakan Tinggi Lemak
Abstract. Hypercholesterolemia is a condition when the total cholesterol level in the blood increases and the LDL cholesterol level in the blood exceeds the normal limit. High chol...
High acceptability of newborn screening for sickle cell disease among post-natal mothers in Western Kenya
High acceptability of newborn screening for sickle cell disease among post-natal mothers in Western Kenya
Sickle cell disease is a genetically inherited blood disorder that manifests early in life with resultant significant health complications. Globally, nearly three quarters of all a...