Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device

View through CrossRef
BACKGROUND: Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients. Pulse oximetry is commonly used to measure blood oxygen saturation and pulse rate for appropriate clinical intervention. But the majority of direct-to-consumer grade pulse oximeters did not pass through in-vivo testing, which results in their accuracy being questionable. Besides this, the ongoing COVID-19 pandemic exposed the limitations of the device in resource limited areas since independent monitoring is needed for COVID-19 patients. The purpose of this study was to perform an in-vivo evaluation of a newly developed smartphone powered low-cost pulse oximeter.METHODS: The new prototype of a smartphone powered pulse oximeter was evaluated against the standard pulse oximeter by taking measurements from fifteen healthy volunteers. The accuracy of measurement was evaluated by calculating the percentage error and standard deviation. A repeatability and reproducibility test were carried out using the ANOVA method.RESULTS: The average accuracy for measuring spot oxygen saturation (SPO2) and pulse rate (PR) was 99.18% with a standard deviation of 0.57 and 98.78% with a standard deviation of 0.61,respectively, when compared with the standard pulse oximeter device. The repeatability and reproducibility of SPO2 measurements were 0.28 and 0.86, respectively, which is in the acceptable range.CONCLUSION: The new prototype of smartphone powered pulse oximeter demonstrated better performance compared to the existing low-cost fingertip pulse oximeters. The device could be used for independent monitoring of COVID-19 patients at health institutions and also for home care.
Title: Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
Description:
BACKGROUND: Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients.
Pulse oximetry is commonly used to measure blood oxygen saturation and pulse rate for appropriate clinical intervention.
But the majority of direct-to-consumer grade pulse oximeters did not pass through in-vivo testing, which results in their accuracy being questionable.
Besides this, the ongoing COVID-19 pandemic exposed the limitations of the device in resource limited areas since independent monitoring is needed for COVID-19 patients.
The purpose of this study was to perform an in-vivo evaluation of a newly developed smartphone powered low-cost pulse oximeter.
METHODS: The new prototype of a smartphone powered pulse oximeter was evaluated against the standard pulse oximeter by taking measurements from fifteen healthy volunteers.
The accuracy of measurement was evaluated by calculating the percentage error and standard deviation.
A repeatability and reproducibility test were carried out using the ANOVA method.
RESULTS: The average accuracy for measuring spot oxygen saturation (SPO2) and pulse rate (PR) was 99.
18% with a standard deviation of 0.
57 and 98.
78% with a standard deviation of 0.
61,respectively, when compared with the standard pulse oximeter device.
The repeatability and reproducibility of SPO2 measurements were 0.
28 and 0.
86, respectively, which is in the acceptable range.
CONCLUSION: The new prototype of smartphone powered pulse oximeter demonstrated better performance compared to the existing low-cost fingertip pulse oximeters.
The device could be used for independent monitoring of COVID-19 patients at health institutions and also for home care.

Related Results

USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
Context is due to the need to develop a methodology for optimising the parameters of pulse oximeter sensor settings based on the practical determination of the accuracy of heart ra...
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
Purpose: Does a non-invasive pulse oximeter probe accurately indicate limb ischemia with inflation of a tourniquet? How are the pulse waveform, perfusion index and oxygen saturatio...
The Effects of Smartphone Use During Resistance Training
The Effects of Smartphone Use During Resistance Training
Several health risks are associated with sedentary behavior; therefore, it is important to better understand behaviors such as smartphone use and how it may influence physical acti...
Penentuan Strategi Bersaing pada Dua Brand Smartphone Menggunakan Teori Permainan
Penentuan Strategi Bersaing pada Dua Brand Smartphone Menggunakan Teori Permainan
Abstract. Smartphone competition in Indonesia, especially among students, is fiercely competitive. In choosing a smartphone, a lot of considerations are made by students. Each smar...
Design and Development of a Low Cost Pulse Oximeter
Design and Development of a Low Cost Pulse Oximeter
Abstract There are various types of pulse oximeters (POs) including ‘reflectance POs’, which can be used on more sites like the forehead, chest or cheeks and work th...
Correlation between dental crown’s color and pulse oximetry interpretation in teeth submitted to bleaching procedures
Correlation between dental crown’s color and pulse oximetry interpretation in teeth submitted to bleaching procedures
To evaluate the correlation between the dental crown’s color and the pulse oximeter result in healthy teeth before and after 30 days of the end bleaching procedures. Material and m...

Back to Top