USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
View through CrossRef
Context is due to the need to develop a methodology for optimising the parameters of pulse oximeter sensor settings based on the practical determination of the accuracy of heart rate and blood oxygen saturation measurements, which, unlike existing ones, do not require analysis of the amplitude of the LED current.A pulse oximeter is one of the sensors that monitor the patient’s vital signs, heart rate and blood oxygen saturation in particular. These indicators are determined based on the analysis of the values of the variable and constant components of the current of the red and infrared LEDs of the pulse oximeter sensor. Therefore, the accuracy of determining vital signs depends on the correct choice of the brightness and duration of the LEDs’ radiation. It is possible to select the current and duration of the LEDs’ radiation, as well as the ADC parameters of the sensor using software. In this case, the final conclusion regarding the correctness of the selected sensor settings is made based on the practical determination of the accuracy of heart rate and blood oxygen saturation measurements.The object is to develop a methodology for assessing the correctness of the pulse oximeter sensor settings based on the analysis of the stationarity of errors in heart rate and blood oxygen saturation measurements.Method. An experimental study of the accuracy of heart rate and blood oxygen saturation measurements by statistical analysis of measurement errors of the developed pulse oximeter model.Results. The practical application of the proposed methodology for determining the optimal parameters of the pulse oximetersensor settings was tested using the example of heart rate measurements.Conclusion. A methodology has been developed to assess the correctness of the choice of sensor setting parameters based on analysing the stationarity of errors in measuring heart rate and oxygen saturation in the patient’s blood. With the help of the developed methodology, the optimal settings parameters of the MAX30102 sensor of the pulse oximeter developed based on the ESP32 board were selected, which ensures the minimum error in measuring heart rate and blood oxygen saturation
National University "Zaporizhzhia Polytechnic"
Title: USE OF THE PARAMETERS OF THE LAW OF DISTRIBUTION OF THE MEASUREMENT ERRORS OF THE PULSE OXIMETER TO SELECT THE SENSOR SETTINGS
Description:
Context is due to the need to develop a methodology for optimising the parameters of pulse oximeter sensor settings based on the practical determination of the accuracy of heart rate and blood oxygen saturation measurements, which, unlike existing ones, do not require analysis of the amplitude of the LED current.
A pulse oximeter is one of the sensors that monitor the patient’s vital signs, heart rate and blood oxygen saturation in particular.
These indicators are determined based on the analysis of the values of the variable and constant components of the current of the red and infrared LEDs of the pulse oximeter sensor.
Therefore, the accuracy of determining vital signs depends on the correct choice of the brightness and duration of the LEDs’ radiation.
It is possible to select the current and duration of the LEDs’ radiation, as well as the ADC parameters of the sensor using software.
In this case, the final conclusion regarding the correctness of the selected sensor settings is made based on the practical determination of the accuracy of heart rate and blood oxygen saturation measurements.
The object is to develop a methodology for assessing the correctness of the pulse oximeter sensor settings based on the analysis of the stationarity of errors in heart rate and blood oxygen saturation measurements.
Method.
An experimental study of the accuracy of heart rate and blood oxygen saturation measurements by statistical analysis of measurement errors of the developed pulse oximeter model.
Results.
The practical application of the proposed methodology for determining the optimal parameters of the pulse oximetersensor settings was tested using the example of heart rate measurements.
Conclusion.
A methodology has been developed to assess the correctness of the choice of sensor setting parameters based on analysing the stationarity of errors in measuring heart rate and oxygen saturation in the patient’s blood.
With the help of the developed methodology, the optimal settings parameters of the MAX30102 sensor of the pulse oximeter developed based on the ESP32 board were selected, which ensures the minimum error in measuring heart rate and blood oxygen saturation.
Related Results
NICU Medication Errors: Describing the Cause and Nature of Medication Errors in a NICU in Qatar
NICU Medication Errors: Describing the Cause and Nature of Medication Errors in a NICU in Qatar
IntroductionA medication error can be defined as “any error occurring in the medication use process” and focuses on problems with the delivery of medication to a patient [1]. Medic...
Dynamic stochastic modeling for inertial sensors
Dynamic stochastic modeling for inertial sensors
Es ampliamente conocido que los modelos de error para sensores inerciales tienen dos componentes: El primero es un componente determinista que normalmente es calibrado por el fabri...
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
Evaluation of a New Smartphone Powered Low-cost Pulse Oximeter Device
BACKGROUND: Measurement of blood oxygen saturation is a vital part of monitoring coronavirus 2019 (COVID-19) patients. Pulse oximetry is commonly used to measure blood oxygen satur...
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
The Use of the Pulse Oximeter in Limb Ischemia: The Pulse Study
Purpose: Does a non-invasive pulse oximeter probe accurately indicate limb ischemia with inflation of a tourniquet? How are the pulse waveform, perfusion index and oxygen saturatio...
From Constitutional Comparison to Life in the Biosphere
From Constitutional Comparison to Life in the Biosphere
From Constitutional Comparison to Life in the Biosphere is a monograph that argues for a fundamental reorientation of constitutional law around the realities of biospheric interdep...
Técnicas de reconstrucción y compensación activa de frentes de onda complejos
Técnicas de reconstrucción y compensación activa de frentes de onda complejos
The continuous improvements of optical design tools and manufacturing technologies of free-form optical elements, allow the creation of new complex-shaped lenses that improve the p...
Design and Development of a Low Cost Pulse Oximeter
Design and Development of a Low Cost Pulse Oximeter
Abstract
There are various types of pulse oximeters (POs) including ‘reflectance POs’, which can be used on more sites like the forehead, chest or cheeks and work th...
Mezinárodní právo na prahu 21. století (dosažený stav, neúspěchy a perspektivy)
Mezinárodní právo na prahu 21. století (dosažený stav, neúspěchy a perspektivy)
The study deal with selected problems of international law at the time of change of the 20th and 21st centuries. Such a milestone gives an opportunity to review the achieved state ...