Intrasubject Repeatability and Interdevice Agreement of Anterior Chamber Depth Measurements by Orbscan and Pentacam in Different Grades of Keratoconus
View through CrossRef
Objectives:
To evaluate the repeatability of anterior chamber depth (ACD) measurements by Orbscan and Pentacam imaging devices in different grades of keratoconus.
Methods:
One examiner performed 3 consecutive ACD measurements with both devices on 74 eyes of 42 keratoconus patient. Repeatability was assessed using intrasession test–retest variability. Within-subject SD was determined for repeatability, and the coefficient of variation was calculated for each measurement. The intraclass correlation coefficient (ICC) was also determined to assess the variance of repeated data.
Results:
Overall, the three ACD measurements were not significantly different either with Pentacam or Orbscan. The ICC index values were greater than 90% with both devices, and it significantly reduced at higher grades of keratoconus with Pentacam. Interdevice differences were statistically significant. The interdevice agreement with ACD measurements was 0.981, and the agreement was lower at higher grades of keratoconus. The 95% limits of agreement between the 2 devices for ACD was from −0.19 to 0.04.
Conclusions:
Our results indicate acceptable repeatability for ACD measurements with both Orbscan and Pentacam in keratoconus patients. The progression of keratoconus has no significant effect on repeatability results of these devices. The good agreement between them allows their interchangeable use.
Ovid Technologies (Wolters Kluwer Health)
Title: Intrasubject Repeatability and Interdevice Agreement of Anterior Chamber Depth Measurements by Orbscan and Pentacam in Different Grades of Keratoconus
Description:
Objectives:
To evaluate the repeatability of anterior chamber depth (ACD) measurements by Orbscan and Pentacam imaging devices in different grades of keratoconus.
Methods:
One examiner performed 3 consecutive ACD measurements with both devices on 74 eyes of 42 keratoconus patient.
Repeatability was assessed using intrasession test–retest variability.
Within-subject SD was determined for repeatability, and the coefficient of variation was calculated for each measurement.
The intraclass correlation coefficient (ICC) was also determined to assess the variance of repeated data.
Results:
Overall, the three ACD measurements were not significantly different either with Pentacam or Orbscan.
The ICC index values were greater than 90% with both devices, and it significantly reduced at higher grades of keratoconus with Pentacam.
Interdevice differences were statistically significant.
The interdevice agreement with ACD measurements was 0.
981, and the agreement was lower at higher grades of keratoconus.
The 95% limits of agreement between the 2 devices for ACD was from −0.
19 to 0.
04.
Conclusions:
Our results indicate acceptable repeatability for ACD measurements with both Orbscan and Pentacam in keratoconus patients.
The progression of keratoconus has no significant effect on repeatability results of these devices.
The good agreement between them allows their interchangeable use.
Related Results
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Abstract
Introduction
The exact manner in which large language models (LLMs) will be integrated into pathology is not yet fully comprehended. This study examines the accuracy, bene...
Repeatability and reproducibility of Orbscan II
Repeatability and reproducibility of Orbscan II
This study aimed to determine the repeatability and reproducibility of Orbscan for anterior and posterior best fit sphere (BFS), simulated keratometry (Sim-K), and central (CCT) an...
Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Abstract
Purpose
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
...
Topographic Keratoconus Incidence in Greece Diagnosed in Routine Consecutive Cataract Procedures: A Consecutive Case Series of 1250 Cases over 5 Years
Topographic Keratoconus Incidence in Greece Diagnosed in Routine Consecutive Cataract Procedures: A Consecutive Case Series of 1250 Cases over 5 Years
Background: Scheimpflug tomography has for many years been an integral part of our pre-operative assessment in cataract extraction. We retrospectively reviewed the incidence of top...
The comparison of white-to-white via triple person-times caliper measuring and machine- measuring in V4c implantable collamer lens implantation
The comparison of white-to-white via triple person-times caliper measuring and machine- measuring in V4c implantable collamer lens implantation
Abstract
Background: This study aimed to compare the differences in white-to-white (WTW) values obtained using the triple-person-times caliper measurement, IOL-Master 700, ...
Comparison of Anterior Chamber Parameters between Normal and Keratoconus Eyes using Scheimpflug Photography
Comparison of Anterior Chamber Parameters between Normal and Keratoconus Eyes using Scheimpflug Photography
ABSTRACT
Aim
To evaluate anterior chamber indices including anterior chamber depth (ACD) and anterior chamber volume (ACV) with the progression of keratoconus using a rotating Sch...
The comparison of white-to-white via triple person-times caliper measuring and machine-measuring in V4c implantable collamer lens implantation
The comparison of white-to-white via triple person-times caliper measuring and machine-measuring in V4c implantable collamer lens implantation
Abstract
This study aimed to compare the differences and characteristics of white-to-white (WTW) values obtained before V4c implantation using triple person-times...
Correlation of the KISA% index and Scheimpflug tomography in ‘normal’, ‘subclinical’, ‘keratoconus‐suspect’ and ‘clinically manifest’ keratoconus eyes
Correlation of the KISA% index and Scheimpflug tomography in ‘normal’, ‘subclinical’, ‘keratoconus‐suspect’ and ‘clinically manifest’ keratoconus eyes
AbstractPurposeTo analyse tomographic changes in eyes classified as ‘normal’, ‘keratoconus‐suspect’ and ‘clinically manifest keratoconus’ based on the established KISA% definition ...