Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
View through CrossRef
Abstract
Purpose
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
Patients and methods
A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18–50 years and a corrected Snellen’s distance visual acuity (CDVA) of 0.8 decimal or better. Angle Kappa is assessed directly using Orbscan® III software version 1.8.165.1. (Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.21r.65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.2.1.60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.
Results
Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (− 0.18° ± 1.8), and it was statistically insignificant (p value = 0.1294). Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (p value = 0.0004 and < 0.0001 consecutively). Bland Altman analysis showed a 95% confidence interval between Orbscan III and Pentacam of − 3.76 to 3.4 and between Orbscan III and Sirius of − 3.79 to 2.26.
Conclusion
Pentacam parameters can be used as a reliable method to calculate angle kappa indirectly, without usage of any additional measurements from other machine. Sirius device parameters could also be used, but with less accurate results. A simple modification to those devices’ software to calculate it, and incorporate it in the printout is possible, and highly recommended.
Springer Science and Business Media LLC
Title: Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Description:
Abstract
Purpose
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
Patients and methods
A prospective randomized cohort study, conducted on 47 eyes of 47 healthy orthotropic individuals, with an age range of 18–50 years and a corrected Snellen’s distance visual acuity (CDVA) of 0.
8 decimal or better.
Angle Kappa is assessed directly using Orbscan® III software version 1.
8.
165.
1.
(Bausch and Lomb Rochester, New York, United States), while Pentacam® HR 1.
21r.
65 (Oculus Optikgeräte GmbH, Wetzlar, Germany) and Sirius device (CSO, version 3.
2.
1.
60, Costruzione Strumenti Oftalmici, Florence, Italy) were used to calculate angle kappa indirectly.
Results
Least mean difference of estimated angle Kappa was between Orbscan and Pentacam devices (− 0.
18° ± 1.
8), and it was statistically insignificant (p value = 0.
1294).
Differences between both Orbscan and Sirius, and Pentacam and Sirius were statistically significant (p value = 0.
0004 and < 0.
0001 consecutively).
Bland Altman analysis showed a 95% confidence interval between Orbscan III and Pentacam of − 3.
76 to 3.
4 and between Orbscan III and Sirius of − 3.
79 to 2.
26.
Conclusion
Pentacam parameters can be used as a reliable method to calculate angle kappa indirectly, without usage of any additional measurements from other machine.
Sirius device parameters could also be used, but with less accurate results.
A simple modification to those devices’ software to calculate it, and incorporate it in the printout is possible, and highly recommended.
Related Results
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Exploring Large Language Models Integration in the Histopathologic Diagnosis of Skin Diseases: A Comparative Study
Abstract
Introduction
The exact manner in which large language models (LLMs) will be integrated into pathology is not yet fully comprehended. This study examines the accuracy, bene...
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
North Syrian Mortaria and Other Late Roman Personal and Utility Objects Bearing Inscriptions of Good Luck
<span style="font-size: 11pt; color: black; font-family: 'Times New Roman','serif'">ΠΗΛΙΝΑ ΙΓ&Delta...
Un manoscritto equivocato del copista santo Theophilos († 1548)
Un manoscritto equivocato del copista santo Theophilos († 1548)
<p><font size="3"><span class="A1"><span style="font-family: 'Times New Roman','serif'">ΕΝΑ ΛΑΝ&...
Restricted kappa chain expression in early ontogeny: biased utilization of V kappa exons and preferential V kappa-J kappa recombinations.
Restricted kappa chain expression in early ontogeny: biased utilization of V kappa exons and preferential V kappa-J kappa recombinations.
To determine the extent of kappa chain diversity in the preimmune repertoire early in development, kappa cDNA libraries were analyzed from 15-d old fetal omentum, 18-d-old fetal li...
Correlation study of changes in visual acuity, pupil diameter, Kappa angle, and Alpha angle after phacoemulsification surgery combined with multifocal IOL implantation
Correlation study of changes in visual acuity, pupil diameter, Kappa angle, and Alpha angle after phacoemulsification surgery combined with multifocal IOL implantation
Abstract
Purpose To study the changes and correlation of visual acuity, pupil size, kappa angle and Alpha angle after phacoemulsification combined with multifocal intraocul...
Double-slit, Comprehensive-Double-Slit and Irregular-Double-Slit Experiments Showing That Light is Photons and It is Photons That Produce both Non-Interference Patterns and Interference Patterns in Same Experiment
Double-slit, Comprehensive-Double-Slit and Irregular-Double-Slit Experiments Showing That Light is Photons and It is Photons That Produce both Non-Interference Patterns and Interference Patterns in Same Experiment
Young’s double slit experiments have been descripted by wave theories, i.e., before and after passing through a double slit, the light is waves. To test the wave description, we ex...
The Improvement in Symptoms After Sublingual Immunotherapy for Japanese Cedar Pollinosis Coincided With the Reduction in Nasal Metachromatic and Eosinophilic Cells
The Improvement in Symptoms After Sublingual Immunotherapy for Japanese Cedar Pollinosis Coincided With the Reduction in Nasal Metachromatic and Eosinophilic Cells
Background and objective
Previously, it was reported that droplet SLIT in Japanese cedar pollinosis reduced metachromatic cell and eosinophil counts in nasal sw...
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Purpose:
The purpose of this study was to assess the repeatability and agreement of horizontal white-to-white (WTW) measurements using 4 different imaging modalities in...