The comparison of white-to-white via triple person-times caliper measuring and machine-measuring in V4c implantable collamer lens implantation

View through CrossRef
Abstract This study aimed to compare the differences and characteristics of white-to-white (WTW) values obtained before V4c implantation using triple person-times caliper, IOL-Master 700, Pentacam HR, and UBM, and to assess their correlation with vaulting. A total of 930 myopia patients (1842 eyes) who were interested in undergoing ICL surgery were assessed before the procedure using various instruments. The WTW measurements were obtained using a triple person-times caliper, Pentacam HR, and IOL-Master 700, whereas the angle-to-angle (ATA) measurements were obtained using UBM. The size of the ICL was subsequently calculated using triple person-times caliper measurements. The vault of the ICL was assessed using Pentacam HR three months after the surgery. The WTW was determined to be 11.30 ± 0.29 mm, 11.43 ± 0.29 mm, and11.86 ± 0.38 mm, respectively, using the triple person-times caliper, Pentacam HR, and IOL-Master 700. The measurement of ATA was 11.57 ± 0.51 mm, as done by UBM. The ICL vault was measured to be 400.97 ± 198.46 µm when examined with Pentacam HR three monthsafter the procedure. The linear regression analyses of ICL size and WTW of triple person-times caliper, ICL vault and WTW were (R = 0.703, p < 0.001; R = 0.0969, p < 0.001) respectively. The highest correlation was found between IOL-Master and Pentacam HR (r = 0.766, p = 0.000). The lowest correlation was found between UBM and Pentacam HR (r = 0.358, p = 0.002). Bland–Altman analysis showed that the 95% limits of agreement (LoA) were the triple person-times caliper and Pentacam HR (– 0.573, 0.298) and the triple person-times caliper and UBM (– 1.15, – 0.605). This indicated a strong agreement between the triple person-times caliper and Pentacam HR and a lack of agreement between the triple person-times caliper and UBM. Triple person-times caliper measurements offer excellent maneuverability, practicality, and reliable outcomes for determining ICL vaults. Measurements obtained using the triple-person caliper were less differece than those obtained using the Pentacam HR.
Title: The comparison of white-to-white via triple person-times caliper measuring and machine-measuring in V4c implantable collamer lens implantation
Description:
Abstract This study aimed to compare the differences and characteristics of white-to-white (WTW) values obtained before V4c implantation using triple person-times caliper, IOL-Master 700, Pentacam HR, and UBM, and to assess their correlation with vaulting.
A total of 930 myopia patients (1842 eyes) who were interested in undergoing ICL surgery were assessed before the procedure using various instruments.
The WTW measurements were obtained using a triple person-times caliper, Pentacam HR, and IOL-Master 700, whereas the angle-to-angle (ATA) measurements were obtained using UBM.
The size of the ICL was subsequently calculated using triple person-times caliper measurements.
The vault of the ICL was assessed using Pentacam HR three months after the surgery.
The WTW was determined to be 11.
30 ± 0.
29 mm, 11.
43 ± 0.
29 mm, and11.
86 ± 0.
38 mm, respectively, using the triple person-times caliper, Pentacam HR, and IOL-Master 700.
The measurement of ATA was 11.
57 ± 0.
51 mm, as done by UBM.
The ICL vault was measured to be 400.
97 ± 198.
46 µm when examined with Pentacam HR three monthsafter the procedure.
The linear regression analyses of ICL size and WTW of triple person-times caliper, ICL vault and WTW were (R = 0.
703, p < 0.
001; R = 0.
0969, p < 0.
001) respectively.
The highest correlation was found between IOL-Master and Pentacam HR (r = 0.
766, p = 0.
000).
The lowest correlation was found between UBM and Pentacam HR (r = 0.
358, p = 0.
002).
Bland–Altman analysis showed that the 95% limits of agreement (LoA) were the triple person-times caliper and Pentacam HR (– 0.
573, 0.
298) and the triple person-times caliper and UBM (– 1.
15, – 0.
605).
This indicated a strong agreement between the triple person-times caliper and Pentacam HR and a lack of agreement between the triple person-times caliper and UBM.
Triple person-times caliper measurements offer excellent maneuverability, practicality, and reliable outcomes for determining ICL vaults.
Measurements obtained using the triple-person caliper were less differece than those obtained using the Pentacam HR.

Related Results

Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
Hubungan Perilaku Pola Makan dengan Kejadian Anak Obesitas
<p><em><span style="font-size: 11.0pt; font-family: 'Times New Roman',serif; mso-fareast-font-family: 'Times New Roman'; mso-ansi-language: EN-US; mso-fareast-langua...
Implantable collamer lens with a central hole for residual refractive error correction after corneal refractive surgery
Implantable collamer lens with a central hole for residual refractive error correction after corneal refractive surgery
Abstract PURPOSE: To assess the visual and refractive outcomes of an implantable collamer lens with a central hole (ICL V4c) for residual refractive error correction after ...
Refractive outcomes after V4c Toric collamer lens implantation over 1y of follow-up
Refractive outcomes after V4c Toric collamer lens implantation over 1y of follow-up
AIM: To evaluate refractive outcomes and corneal astigmatism changes after Toric implantable collamer lens with a central port (V4c T-ICL) implantation over 1y of follow-up. METHOD...
Osoby niejednokrotnie przebywające w izbie wytrzeźwień
Osoby niejednokrotnie przebywające w izbie wytrzeźwień
In Poland we have at present in towns 29 detoxication centres with 1,226 beds; people found by the police in public places in a state of intoxication are more and more often taken ...

Back to Top