This study aimed to determine the repeatability and reproducibility of Orbscan for anterior and posterior best fit sphere (BFS), simulated keratometry (Sim-K), and central (CCT) and mid-peripheral (PCT) corneal thickness measurements in healthy eyes. Orbscan was performed in 40 healthy eyes (20 subjects) three consecutive times on each cornea during three visits scheduled over one week. Repeatability and reproducibility coefficients [Bland and Altman’s coefficient (BAC), coefficient of variation (CV) and intraclass correlation (ICC)] were calculated for Orbscan anterior and posterior BFS, Sim-K and corneal pachymetry (central, superior, inferior, nasal and temporal locations). Repeatability was calculated using three consecutive measurements during each visit, while reproducibility was calculated using the average of the measurements obtained at each visit. High repeatability was found for all Orbscan measurements (r2<0.01; P>0.05, two-way ANOVA) with BAC and CV <1% (except in PCT coefficients; from 0.97% to 1.67%) and ICC close to 0.98- 0.99 for all visits. High reproducibility was also found for all Orbscan measurements (r2<0.01; P>0.05 two-way ANOVA). BAC values were less than 1% for both BFS and Sim- K, and between 1.21 and 2.20% for corneal pachymetry. CV values were less than 1% (except in superior, nasal and temporal PCT, where they ranged from 1.06 to 1.30%). ICC was close to 0.98-0.99 for all measurements. The BAC of reproducibility was higher than the CV of reproducibility. PCT showed less repeatability and reproducibility than CCT. The Orbscan provides non-invasive, repeatable and reproducible measurements of anterior and posterior BFS, simulated keratometry. and central and mid-peripheral pachymetry in healthy eyes.
PAGEPress Publications
Title: Repeatability and reproducibility of Orbscan II
Description:
This study aimed to determine the repeatability and reproducibility of Orbscan for anterior and posterior best fit sphere (BFS), simulated keratometry (Sim-K), and central (CCT) and mid-peripheral (PCT) corneal thickness measurements in healthy eyes.
Orbscan was performed in 40 healthy eyes (20 subjects) three consecutive times on each cornea during three visits scheduled over one week.
Repeatability and reproducibility coefficients [Bland and Altman’s coefficient (BAC), coefficient of variation (CV) and intraclass correlation (ICC)] were calculated for Orbscan anterior and posterior BFS, Sim-K and corneal pachymetry (central, superior, inferior, nasal and temporal locations).
Repeatability was calculated using three consecutive measurements during each visit, while reproducibility was calculated using the average of the measurements obtained at each visit.
High repeatability was found for all Orbscan measurements (r2<0.
01; P>0.
05, two-way ANOVA) with BAC and CV <1% (except in PCT coefficients; from 0.
97% to 1.
67%) and ICC close to 0.
98- 0.
99 for all visits.
High reproducibility was also found for all Orbscan measurements (r2<0.
01; P>0.
05 two-way ANOVA).
BAC values were less than 1% for both BFS and Sim- K, and between 1.
21 and 2.
20% for corneal pachymetry.
CV values were less than 1% (except in superior, nasal and temporal PCT, where they ranged from 1.
06 to 1.
30%).
ICC was close to 0.
98-0.
99 for all measurements.
The BAC of reproducibility was higher than the CV of reproducibility.
PCT showed less repeatability and reproducibility than CCT.
The Orbscan provides non-invasive, repeatable and reproducible measurements of anterior and posterior BFS, simulated keratometry.
and central and mid-peripheral pachymetry in healthy eyes.
Related Results
Intrasubject Repeatability and Interdevice Agreement of Anterior Chamber Depth Measurements by Orbscan and Pentacam in Different Grades of Keratoconus
Intrasubject Repeatability and Interdevice Agreement of Anterior Chamber Depth Measurements by Orbscan and Pentacam in Different Grades of Keratoconus
Objectives:
To evaluate the repeatability of anterior chamber depth (ACD) measurements by Orbscan and Pentacam imaging devices in different grades of keratoconus.
...
Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Angle Kappa agreement between Scheimpflug tomography, combined placido Scheimpflug and combined slit scanning placido systems
Abstract
Purpose
To compare the measured or calculated angle Kappa using Oculus pentacam HR, Sirius and Orbscan III devices.
...
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Repeatability and Agreement of Horizontal Corneal Diameter Measurements Between Scanning-Slit Topography, Dual Rotating Scheimpflug Camera With Placido Disc Tomography, Placido Disc Topography, and Optical Coherence Tomography
Purpose:
The purpose of this study was to assess the repeatability and agreement of horizontal white-to-white (WTW) measurements using 4 different imaging modalities in...
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
Design for Accuracy and Repeatability for Robots Using Taguchi Methods
In this paper, we apply Taguchi Methods in the framework of concurrent engineering design for robot-based manufacturing sys tems to obtain higher accuracy and repeatability for rob...
Understanding reproducibility of bioinformatics workflows
Understanding reproducibility of bioinformatics workflows
Reproducibility is an essential factor in establishing the reliability of results reported in the scientific literature and ultimately in the knowledge that is generated. This know...
Repeatability and reproducibility of Keratograph 5M corneal topography
Repeatability and reproducibility of Keratograph 5M corneal topography
Background
Corneal topography is an important diagnostic tool and highly repeatable and reproducible topographic devices are essential in eye care practice. Placido ...
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
Uncertainty evaluation of measurement of orientation repeatability for industrial robots
PurposeThis paper aims to propose a reasonable method to evaluate uncertainty of measurement of industrial robots’ orientation repeatability and solve the non-linear problem existi...
Oxidative stress markers have low repeatability: A meta-analysis and simulation study with implications for measuring physiological condition and fitness
Oxidative stress markers have low repeatability: A meta-analysis and simulation study with implications for measuring physiological condition and fitness
Abstract
Markers of oxidative stress are widely used as indicators of physiological state, but their utility for ecological and evolutionary inference remains uncer...