Traceable Calibration of Seismometers
View through CrossRef
At present, seismometers are not traceably calibrated. This means that their output sensitivity is not determined in a way that is traceable to the International Systems of Units (SI). The European research project 19ENV03 Infra-AUV, which is part of the EMPIR programme, develops methods and procedures to enable such traceable calibrations.In contrast to many other sensors, seismometers are operated stationary in their typical measurement application, i.e., they must not be moved after their deployment. Conventional calibration approaches which involve a laboratory calibration of the seismometer to be calibrated are therefore not feasible. For this reason, a new concept currently developed by different European partners within the Infra-AUV project proposes an on-site calibration scheme.For the on-site calibration, a reference seismometer is traceably calibrated to the SI in a laboratory. This reference is then used on-site to provide a secondary calibration of other seismometers, e.g. in a seismic station, using natural excitation sources [Schwardt et al., 2022, DOI: 10.1007/s10712-022-09713-4].The calibration of reference seismometers in the laboratory is carried out as a primary calibration. This means that the measured quantity (the velocity-proportional voltage output) is compared to a different quantity, in this case to a dynamic displacement measurement traced back to the units length and time, which can be measured very precisely by laser interferometry. In this calibration, the seismometer is excited with low-frequency mechanical vibrations generated by electrodynamic exciters. These calibrations must be performed for the horizontal and vertical axes. The frequency range of interest is from 20 Hz down to 0.01 Hz, depending on the seismometer under test. Either mono-frequency sinusoidal excitations of different frequencies are applied subsequently, or multiple frequencies are excited simultaneously using a multi-sine approach. The magnitudes and phases of both measured signals, the interferometric reference and the seismometer under test, are determined by using sine approximation algorithms or by applying a discrete Fourier transform (DFT).The results of the laboratory calibration, the transfer function of the reference seismometer, can then be derived from the ratios of the measured magnitudes and the differences of the phase angles for the different excitation frequencies. In addition, the associated measurement uncertainties are estimated and are part of the calibration result. Influences that may change the sensitivity of a seismometer, e.g., temperature effects, electromagnetic sensitivity, or ground stiffness need to be analysed and additionally taken into account for the uncertainty estimation.For the uncertainty of the on-site calibration, differences between the laboratory and the on-site environment also need to be taken into account. This includes, for example, aspects like typically different temperatures or different ground materials.
Title: Traceable Calibration of Seismometers
Description:
At present, seismometers are not traceably calibrated.
This means that their output sensitivity is not determined in a way that is traceable to the International Systems of Units (SI).
The European research project 19ENV03 Infra-AUV, which is part of the EMPIR programme, develops methods and procedures to enable such traceable calibrations.
In contrast to many other sensors, seismometers are operated stationary in their typical measurement application, i.
e.
, they must not be moved after their deployment.
Conventional calibration approaches which involve a laboratory calibration of the seismometer to be calibrated are therefore not feasible.
For this reason, a new concept currently developed by different European partners within the Infra-AUV project proposes an on-site calibration scheme.
For the on-site calibration, a reference seismometer is traceably calibrated to the SI in a laboratory.
This reference is then used on-site to provide a secondary calibration of other seismometers, e.
g.
in a seismic station, using natural excitation sources [Schwardt et al.
, 2022, DOI: 10.
1007/s10712-022-09713-4].
The calibration of reference seismometers in the laboratory is carried out as a primary calibration.
This means that the measured quantity (the velocity-proportional voltage output) is compared to a different quantity, in this case to a dynamic displacement measurement traced back to the units length and time, which can be measured very precisely by laser interferometry.
In this calibration, the seismometer is excited with low-frequency mechanical vibrations generated by electrodynamic exciters.
These calibrations must be performed for the horizontal and vertical axes.
The frequency range of interest is from 20 Hz down to 0.
01 Hz, depending on the seismometer under test.
Either mono-frequency sinusoidal excitations of different frequencies are applied subsequently, or multiple frequencies are excited simultaneously using a multi-sine approach.
The magnitudes and phases of both measured signals, the interferometric reference and the seismometer under test, are determined by using sine approximation algorithms or by applying a discrete Fourier transform (DFT).
The results of the laboratory calibration, the transfer function of the reference seismometer, can then be derived from the ratios of the measured magnitudes and the differences of the phase angles for the different excitation frequencies.
In addition, the associated measurement uncertainties are estimated and are part of the calibration result.
Influences that may change the sensitivity of a seismometer, e.
g.
, temperature effects, electromagnetic sensitivity, or ground stiffness need to be analysed and additionally taken into account for the uncertainty estimation.
For the uncertainty of the on-site calibration, differences between the laboratory and the on-site environment also need to be taken into account.
This includes, for example, aspects like typically different temperatures or different ground materials.
Related Results
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
Why do field seismometers need to be calibrated? Benefits through traceably calibrated seismometers from the laboratory to the field
Why do field seismometers need to be calibrated? Benefits through traceably calibrated seismometers from the laboratory to the field
As part of the EU-funded joint research project "Metrology for low-frequency sound and vibration - 19ENV03 Infra-AUV" laboratory calibration methods for seismometers in the low fre...
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-S...
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Aiming at solving the problems such as long calibration time, low precision, and complex operation in hand-eye calibration of welding robot, an automatic hand-eye calibration algor...
Binocular vision calibration method for a long-wavelength infrared camera and a visible spectrum camera with different resolutions
Binocular vision calibration method for a long-wavelength infrared camera and a visible spectrum camera with different resolutions
We present a calibration plate for the binocular vision system, which is composed of a long-wavelength infrared camera and a visible spectrum camera with different resolutions. The...
Applicability of self-consistency calibration method for polarimetric cloud radars
Applicability of self-consistency calibration method for polarimetric cloud radars
Cloud radars are powerful tools for investigating cloud formation, radiative processes, and cloud microphysics. In recent years, polarimetric cloud radars have become increasingly ...
A calibration method combining hand-eye calibration and TCP calibration
A calibration method combining hand-eye calibration and TCP calibration
Abstract
Robots' traditional Tool Center Point (TCP) calibration and hand-eye calibration are implemented independently, which is likely to generate large cumulative errors...
A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
Background:
To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as pe...