Binocular vision calibration method for a long-wavelength infrared camera and a visible spectrum camera with different resolutions
View through CrossRef
We present a calibration plate for the binocular vision system, which is composed of a long-wavelength infrared camera and a visible spectrum camera with different resolutions. The calibration plate mainly consists of a white low-temperature aluminum plate with 7×7 round through-holes, a black high-temperature stainless steel plate, and a heating plate. It can be captured by the long-wavelength infrared camera and visible spectrum camera simultaneously. In order to reduce the influence of thermal crosstalk on the edge and angle sharpness of the thermal image of the chessboard calibration plate, we use the round through-holes to replace the black-white squares in the chessboard calibration plate. Based on the fabricated calibration plate, we also propose a related calibration method. The proposed method can quickly detect the calibration plate by using the YOLO-V4 neural network. The affine transformation is performed to get the front view of the calibration plate, and a novel circular detection strategy based on arc level instead of pixel-level is adopted to detect the edges of the round through-holes in the calibration plate. The centers of round through-holes are detected, and the parameters of the cameras are calculated according to the coordinates of the centers in the image coordinate system. The simulation experiments and error analysis have been done to verify the centers detection method. The simulation results show that the error of center detection is always less than 1.4 (pixel). In order to further verify the performance of the calibration plate and the proposed calibration method, a binocular vision system based on long-wavelength infrared camera and visible spectrum camera is fabricated, and the verification experiments have been done. In experiments, our calibration plate and the proposed method are compared with the famous Zhang’s method. The calibration’s average overall mean errors of the visible spectrum camera and long-wavelength infrared camera are about 0.0126 (pixel) and 0.0238 (pixel), and they are respectively decreased by 78.13% and 81.93% compared with Zhang’s method. The re-projection error of the binocular vision system is about 0.548 (pixel), which is decreased by 24.52% compared with Zhang’s method. The average calibration time of the proposed method is about 0.26s.
Title: Binocular vision calibration method for a long-wavelength infrared camera and a visible spectrum camera with different resolutions
Description:
We present a calibration plate for the binocular vision system, which is composed of a long-wavelength infrared camera and a visible spectrum camera with different resolutions.
The calibration plate mainly consists of a white low-temperature aluminum plate with 7×7 round through-holes, a black high-temperature stainless steel plate, and a heating plate.
It can be captured by the long-wavelength infrared camera and visible spectrum camera simultaneously.
In order to reduce the influence of thermal crosstalk on the edge and angle sharpness of the thermal image of the chessboard calibration plate, we use the round through-holes to replace the black-white squares in the chessboard calibration plate.
Based on the fabricated calibration plate, we also propose a related calibration method.
The proposed method can quickly detect the calibration plate by using the YOLO-V4 neural network.
The affine transformation is performed to get the front view of the calibration plate, and a novel circular detection strategy based on arc level instead of pixel-level is adopted to detect the edges of the round through-holes in the calibration plate.
The centers of round through-holes are detected, and the parameters of the cameras are calculated according to the coordinates of the centers in the image coordinate system.
The simulation experiments and error analysis have been done to verify the centers detection method.
The simulation results show that the error of center detection is always less than 1.
4 (pixel).
In order to further verify the performance of the calibration plate and the proposed calibration method, a binocular vision system based on long-wavelength infrared camera and visible spectrum camera is fabricated, and the verification experiments have been done.
In experiments, our calibration plate and the proposed method are compared with the famous Zhang’s method.
The calibration’s average overall mean errors of the visible spectrum camera and long-wavelength infrared camera are about 0.
0126 (pixel) and 0.
0238 (pixel), and they are respectively decreased by 78.
13% and 81.
93% compared with Zhang’s method.
The re-projection error of the binocular vision system is about 0.
548 (pixel), which is decreased by 24.
52% compared with Zhang’s method.
The average calibration time of the proposed method is about 0.
26s.
Related Results
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
COMPARISON OF BINOCULAR SPECTACLE PRESCRIPTION VERSUS PATCHING THERAPY IN REFRACTIVE AMBLYOPIA PATIENTS
COMPARISON OF BINOCULAR SPECTACLE PRESCRIPTION VERSUS PATCHING THERAPY IN REFRACTIVE AMBLYOPIA PATIENTS
Background: Amblyopia is the leading cause of visual impairment in children, adolescents, and middle-aged adults, characterized by a loss of binocularity and a significant reductio...
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Design of high transparent infrared stealth thin films based on FTO/Ag/FTO structure
Multi-spectral compatible stealth materials have become an imperative development trend, especially visible and infrared compatible stealth materials have become the most important...
Decoding Visual Symptoms in Migraine: The Role of Binocular Vision Abnormalities: A Cross-sectional Study
Decoding Visual Symptoms in Migraine: The Role of Binocular Vision Abnormalities: A Cross-sectional Study
Introduction: Migraine is a chronic, neurovascular disorder characterised by recurrent attacks of headaches. Migraine is also associated with a variety of symptoms, including sensi...
Frequency and associated factors of accommodation and non-strabismic binocular vision dysfunction among medical university students
Frequency and associated factors of accommodation and non-strabismic binocular vision dysfunction among medical university students
AIM: To investigate the frequency and associated factors of accommodation and non-strabismic binocular vision dysfunction among medical university students.
METHODS: Totally 158 st...
A simple and precise calibration method for binocular vision
A simple and precise calibration method for binocular vision
Abstract
Binocular vision is an important part of machine vision measurement. Calibration accuracy is crucial for binocular vision. As for the determination of th...
Retinitis Pigmentosa
Retinitis Pigmentosa
In studying the cases with typical and atypical pigmentary degeneration of the retina we strived to analyse in the clinical material all ophthalmoscopic and ocular changes together...
Development of Binocular Vision
Development of Binocular Vision
AbstrakPenglihatan binokular secara harfiah berarti penglihatan dengan 2 mata dan dengan adanya penglihatan binokular, kita dapat melihat dunia dalam 3 dimensi meskipun bayangan ya...