Computed Dosimeter Dose Index on a 16-Slice Computed Tomography Scanner
View through CrossRef
A computed tomography dose index can be used to quantify the radiation dose received during a CT scan and it is an indicator of the radiation dose to the polymetaylenmetaAcrylate (PMMA) standardized phantom. The objective of this study was 2-fold. The first was to measure the computed tomography (CT) radiation dose for the head and body polymetaylelenmetaAcrylate (PMMA) phantoms and to determine the accuracy of the CT radiation dose parameter displayed on the CT scanner console; these were measured in this investigation and compared with the dose displayed on the CT scanner console. The dose was calculated using the formalism described in the American Association of Physics in Medicine (AAPM) Report 96. The second was to compare the dosimetric results of the head and body polymetaylelenmetaAcrylate (PMMA) phantoms with dose reference levels published in international journals, as well as to measure the central cumulative dose (DL
′
(0)), as recommended by the American Association of Physics in Medicine (AAPM) report 111. This is a new, cutting-edge methodology for estimating the CT radiation dosage provided by the abdomen, thorax, and head of a PMMA phantom. We used a Philips Big Bore CT scanner with 16 slices. A CT dosimeter head phantom with a diameter of 16 cm, a CT dosimeter body phantom with a diameter of 32 cm, a 100 mm pencil chamber (PC), and a 20 mm short chamber (SC) were employed. These were coupled to an electrometer and a dosimetric readout device. The measured volume computed tomography dose index (CTDIvol) values were in good agreement with the CT radiation dose displayed on the corresponding CT scanner console. The percentage disagreement was less than 10%, with a maximal difference of 1.7% and 5.5% for the body and head phantom, respectively. The central cumulative dose (DL (0)) measurements (for L
′
= 100 mm) also matched nominal or the corresponding computed tomography dose index (CT) scanner console volume computed tomography dose index (CTDIvol) values. In this case, the agreement is always below 3% for abdomen scans and 1.0% for head examinations. This result implies that the radiation dose supplied by the 16-slice computed tomography (CT) system was in good agreement with the international dose reference level and we observed something different.
Title: Computed Dosimeter Dose Index on a 16-Slice Computed Tomography Scanner
Description:
A computed tomography dose index can be used to quantify the radiation dose received during a CT scan and it is an indicator of the radiation dose to the polymetaylenmetaAcrylate (PMMA) standardized phantom.
The objective of this study was 2-fold.
The first was to measure the computed tomography (CT) radiation dose for the head and body polymetaylelenmetaAcrylate (PMMA) phantoms and to determine the accuracy of the CT radiation dose parameter displayed on the CT scanner console; these were measured in this investigation and compared with the dose displayed on the CT scanner console.
The dose was calculated using the formalism described in the American Association of Physics in Medicine (AAPM) Report 96.
The second was to compare the dosimetric results of the head and body polymetaylelenmetaAcrylate (PMMA) phantoms with dose reference levels published in international journals, as well as to measure the central cumulative dose (DL
′
(0)), as recommended by the American Association of Physics in Medicine (AAPM) report 111.
This is a new, cutting-edge methodology for estimating the CT radiation dosage provided by the abdomen, thorax, and head of a PMMA phantom.
We used a Philips Big Bore CT scanner with 16 slices.
A CT dosimeter head phantom with a diameter of 16 cm, a CT dosimeter body phantom with a diameter of 32 cm, a 100 mm pencil chamber (PC), and a 20 mm short chamber (SC) were employed.
These were coupled to an electrometer and a dosimetric readout device.
The measured volume computed tomography dose index (CTDIvol) values were in good agreement with the CT radiation dose displayed on the corresponding CT scanner console.
The percentage disagreement was less than 10%, with a maximal difference of 1.
7% and 5.
5% for the body and head phantom, respectively.
The central cumulative dose (DL (0)) measurements (for L
′
= 100 mm) also matched nominal or the corresponding computed tomography dose index (CT) scanner console volume computed tomography dose index (CTDIvol) values.
In this case, the agreement is always below 3% for abdomen scans and 1.
0% for head examinations.
This result implies that the radiation dose supplied by the 16-slice computed tomography (CT) system was in good agreement with the international dose reference level and we observed something different.
Related Results
Incorporating multislice imaging into x‐ray CT polymer gel dosimetry
Incorporating multislice imaging into x‐ray CT polymer gel dosimetry
Purpose:To evaluate multislice computed tomography (CT) scanning for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD) and...
Development of end-to-end test phantom for stereotactic radiosurgery using alanine dosimeter
Development of end-to-end test phantom for stereotactic radiosurgery using alanine dosimeter
Alanine dosimeters are generally used in high-dose dosimetry, while accurate and precise dose delivery verification procedures are important in stereotactic radiosurgery (SRS). Hen...
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Investigation of alanine dosimeter for output measurement in radiotherapy: Results from multicenter in Thailand
Background: Alanine dosimeters are generally used in high-dose industrial applications (kGy). Later, research into employing alanine as a dosimeter in radiotherapy (1-20 Gy) has in...
SU‐FF‐I‐89: Effects of EPI Slice Angle, Slice Thickness and Phase Encoding Direction On FMRI Sensitivity in Areas with Susceptibility Artifact
SU‐FF‐I‐89: Effects of EPI Slice Angle, Slice Thickness and Phase Encoding Direction On FMRI Sensitivity in Areas with Susceptibility Artifact
Purpose: A key problem of using EPI for BOLD fMRI is the field inhomogeneities near air/tissue interfaces, or the susceptibility artifact. Therefore, fMRI studies in brain areas su...
Does bore size matter?—A comparison of the subjective perception of patient comfort during low field (0.55 Tesla) and standard (1.5 Tesla) MRI imaging
Does bore size matter?—A comparison of the subjective perception of patient comfort during low field (0.55 Tesla) and standard (1.5 Tesla) MRI imaging
The purpose of the present study was to evaluate the subjectively perceived patient comfort during magnetic resonance imaging (MRI) examinations and to assess potential differences...
Q&A with Sami Iskander
Q&A with Sami Iskander
Oil prices, profits, and capital spending are on the rise, and future hydrocarbon demand appears robust. Do you expect the industry to enjoy a sustained boom?
All in...
Image Quality and Radiation Dose Assessment for the Clinical Applied 16 Slice CT Scanner using PMMA phantom and Quality Assurance phantom
Image Quality and Radiation Dose Assessment for the Clinical Applied 16 Slice CT Scanner using PMMA phantom and Quality Assurance phantom
When compared to other radiological examinations, computed tomography has a significantly higher radiation exposure and employs ionizing radiation to make an image of a slice or cu...
Surface and build-up region dose measurement in head and neck region using radio - photoluminescent glass dosimeter
Surface and build-up region dose measurement in head and neck region using radio - photoluminescent glass dosimeter
The use of commercially available radio-photoluminescent glass dosimeter (RPLGD) for radiation measurement has increased in recent years. It has been shown to have superior dosimet...