Comparison of arterial and capillary blood glucose measured using a glucometer and blood gas analyzer, and its association with acuity of illness: A longitudinal cohort study

View through CrossRef
Objectives: Dysglycemia is one of the major complications in the critical care unit. Many factors interfere with the accuracy of blood glucose measurements among critically ill patients; therefore, results must be interpreted with caution. This longitudinal cohort study assessed the level of agreement between blood glucose values obtained from capillary and arterial blood using a glucometer over time and its association with the Acute Physiology and Chronic Health Evaluation-II score. Materials and Methods: Eighty-one participants were selected by convenience sampling method. Arterial and capillary blood blood glucose levels were measured at three points of time (i.e. on intensive care unit admission, at 24 hours, and 36 hours after cardiac surgery) using a Control-D glucometer and blood gas analyzer. Statistical analysis: Inferential analyses like two-way repeated measures analysis of variance, Wilcoxon’s signed rank test, intraclass correlation coefficient, and receiver operating characteristics analysis were used. Results: The study results revealed that there was a good agreement between the arterial and capillary blood sugar measured using a glucometer (intraclass correlation coefficient ranges from 0.95 to 0.98). The difference between the blood glucose levels measured using different methods was comparable (P > 0.05) during the admission at the range of hyperglycemia. However, at 24 and 36 h after cardiac surgery, the capillary blood significantly (P < 0.05) overestimated blood glucose level (by18–22 mg/dL and 30 mg/dL) as compared to arterial blood glucose measured using a glucometer and blood gas analyzer. Conclusions: Arterial blood in a glucometer can be used for measuring blood glucose levels in a critical care setting whenever an arterial access is available, or when a blood gas analyzer is not accessible.
Title: Comparison of arterial and capillary blood glucose measured using a glucometer and blood gas analyzer, and its association with acuity of illness: A longitudinal cohort study
Description:
Objectives: Dysglycemia is one of the major complications in the critical care unit.
Many factors interfere with the accuracy of blood glucose measurements among critically ill patients; therefore, results must be interpreted with caution.
This longitudinal cohort study assessed the level of agreement between blood glucose values obtained from capillary and arterial blood using a glucometer over time and its association with the Acute Physiology and Chronic Health Evaluation-II score.
Materials and Methods: Eighty-one participants were selected by convenience sampling method.
Arterial and capillary blood blood glucose levels were measured at three points of time (i.
e.
on intensive care unit admission, at 24 hours, and 36 hours after cardiac surgery) using a Control-D glucometer and blood gas analyzer.
Statistical analysis: Inferential analyses like two-way repeated measures analysis of variance, Wilcoxon’s signed rank test, intraclass correlation coefficient, and receiver operating characteristics analysis were used.
Results: The study results revealed that there was a good agreement between the arterial and capillary blood sugar measured using a glucometer (intraclass correlation coefficient ranges from 0.
95 to 0.
98).
The difference between the blood glucose levels measured using different methods was comparable (P > 0.
05) during the admission at the range of hyperglycemia.
However, at 24 and 36 h after cardiac surgery, the capillary blood significantly (P < 0.
05) overestimated blood glucose level (by18–22 mg/dL and 30 mg/dL) as compared to arterial blood glucose measured using a glucometer and blood gas analyzer.
Conclusions: Arterial blood in a glucometer can be used for measuring blood glucose levels in a critical care setting whenever an arterial access is available, or when a blood gas analyzer is not accessible.

Related Results

Comparison of CSF Glucose Values Obtained by Glucometer and Autoanalyzer
Comparison of CSF Glucose Values Obtained by Glucometer and Autoanalyzer
Bacterial meningitis is a disease with a high fatality rate and the chance of developing severe complications are more, which include hearing loss, seizures, limb weakness, and dif...
Blogging Illness: Recovering in Public
Blogging Illness: Recovering in Public
As a mode of open access public self-expression, blogs are one form of the unfolding massification of culture (Lovink). Though widely varied in content and style, they are characte...
Blood Glucose Measurement in Rabbits - Performance Evaluation of Portable Glucometers
Blood Glucose Measurement in Rabbits - Performance Evaluation of Portable Glucometers
Background: Blood glucose level measurement plays an important role in the diagnosis and prognosis of different clinical conditions in rabbits. Portable glucometers offer a quick a...
Effect of hematocrit on accuracy of two point-of-care glucometers for use in dogs
Effect of hematocrit on accuracy of two point-of-care glucometers for use in dogs
Abstract Objective—To determine the effect of Hct on blood glucose readings of dogs obtained by use of 2 point-of-care (POC) blood glucometers and a laboratory analyzer. Animals—18...
Cohort studies
Cohort studies
A cohort study is one in which the outcome (usually disease status) is ascertained for groups of individuals defined on the basis of their exposure. At the time exposure status is ...
A Comparative Study of Blood Glucose Level Measurement Between Glucometer and Semi Autoanalyser
A Comparative Study of Blood Glucose Level Measurement Between Glucometer and Semi Autoanalyser
Background; Glucometers are the excellent tools for self-monitoring of blood glucose (SMBG). Glucometer is widely used at hospitals and homes as a first line tool to get an idea ab...
Rapid detection of bacterial meningitis using a point-of-care glucometer
Rapid detection of bacterial meningitis using a point-of-care glucometer
Background In case of acute bacterial meningitis, a decision on the need for intensive care admission should be made within the first hour. The aim of this study was to...
The capillary fascicle in skeletal muscle: Structural and functional physiology of RBC distribution in capillary networks
The capillary fascicle in skeletal muscle: Structural and functional physiology of RBC distribution in capillary networks
Key points The capillary module, consisting of parallel capillaries from arteriole to venule, is classically considered as the building block of complex capillary networks. In skel...

Back to Top