Comparison of CSF Glucose Values Obtained by Glucometer and Autoanalyzer

View through CrossRef
Bacterial meningitis is a disease with a high fatality rate and the chance of developing severe complications are more, which include hearing loss, seizures, limb weakness, and difficulties with vision, speech, language, memory, and communication. During bacterial meningitis, cerebrospinal fluid(CSF) glucose level decreases owing to increased glycolysis, impaired glucose transport, or consumption by microorganisms. The prompt estimation of CSF glucose helps decide whether to use antibiotics in patients suspected of bacterial meningitis. This study is done to determine whether a point of care device(glucometer) can be used as an alternative to estimate CSF glucose level by finding the correlation between CSF glucose values obtained by glucometer and autoanalyzer. CSF samples that are clear and not contaminated with blood received by the Biochemistry laboratory, Osmania General Hospital, during the study period were analyzed in this cross-sectional study. Glucose level is estimated in CSF samples using an amperometric glucometer. Also, the glucose level is estimated in the CSF samples by autoanalyzer via hexokinase method, which is considered as the reference method. The CSF glucose values obtained by the two methods are compared by linear regression analysis. The mean value of CSF glucose measured by autoanalyzer is 59.26 ±29.02 mg/dl, and the mean value measured by glucometer is 65.01±30.56mg/dl. The percentage bias for CSF glucose measurement by glucometer is 9.60 percent. The correlation coefficient measured by linear regression analysis is 0.9846. The standard error is 5.35mg/dl. The glucometer could not detect CSF glucose values below 20 mg/dl (measured by autoanalyzer). The glucometer indicated a low value for 28 of 32 samples with glucose values below 20mg/dl (displayed as 'low'). For the remaining four samples, glucometer showed a higher mean value with a percentage bias of 53.2 percent. Thus, a glucometer can help to detect low CSF glucose values in bacterial meningitis. The comparison of CSF glucose estimation by glucometer with autoanalyzer (reference method) gives a correlation coefficient of 0.9846. Also, the glucometer detected values below 20 mg/dl as low values (displayed as 'low'). Therefore, a glucometer can be used for point-of-care estimation of CSF glucose so that initiation of treatment with antibiotics is not delayed.
Title: Comparison of CSF Glucose Values Obtained by Glucometer and Autoanalyzer
Description:
Bacterial meningitis is a disease with a high fatality rate and the chance of developing severe complications are more, which include hearing loss, seizures, limb weakness, and difficulties with vision, speech, language, memory, and communication.
During bacterial meningitis, cerebrospinal fluid(CSF) glucose level decreases owing to increased glycolysis, impaired glucose transport, or consumption by microorganisms.
The prompt estimation of CSF glucose helps decide whether to use antibiotics in patients suspected of bacterial meningitis.
This study is done to determine whether a point of care device(glucometer) can be used as an alternative to estimate CSF glucose level by finding the correlation between CSF glucose values obtained by glucometer and autoanalyzer.
CSF samples that are clear and not contaminated with blood received by the Biochemistry laboratory, Osmania General Hospital, during the study period were analyzed in this cross-sectional study.
Glucose level is estimated in CSF samples using an amperometric glucometer.
Also, the glucose level is estimated in the CSF samples by autoanalyzer via hexokinase method, which is considered as the reference method.
The CSF glucose values obtained by the two methods are compared by linear regression analysis.
The mean value of CSF glucose measured by autoanalyzer is 59.
26 ±29.
02 mg/dl, and the mean value measured by glucometer is 65.
01±30.
56mg/dl.
The percentage bias for CSF glucose measurement by glucometer is 9.
60 percent.
The correlation coefficient measured by linear regression analysis is 0.
9846.
The standard error is 5.
35mg/dl.
The glucometer could not detect CSF glucose values below 20 mg/dl (measured by autoanalyzer).
The glucometer indicated a low value for 28 of 32 samples with glucose values below 20mg/dl (displayed as 'low').
For the remaining four samples, glucometer showed a higher mean value with a percentage bias of 53.
2 percent.
Thus, a glucometer can help to detect low CSF glucose values in bacterial meningitis.
The comparison of CSF glucose estimation by glucometer with autoanalyzer (reference method) gives a correlation coefficient of 0.
9846.
Also, the glucometer detected values below 20 mg/dl as low values (displayed as 'low').
Therefore, a glucometer can be used for point-of-care estimation of CSF glucose so that initiation of treatment with antibiotics is not delayed.

Related Results

Rapid detection of bacterial meningitis using a point-of-care glucometer
Rapid detection of bacterial meningitis using a point-of-care glucometer
Background In case of acute bacterial meningitis, a decision on the need for intensive care admission should be made within the first hour. The aim of this study was to...
Intracranial Cerebrospinal Fluid Volume Evaluation in Healthy People and Hydrocephalus Patients using SPACE Sequence
Intracranial Cerebrospinal Fluid Volume Evaluation in Healthy People and Hydrocephalus Patients using SPACE Sequence
Introduction: Cerebrospinal Fluid (CSF) is produced mainly by the choroid plexus but with a substantial influence by the ependymal lining of the ventricles in the brain. Hydrocepha...
Association of CSF glucose concentration with neurosyphilis diagnosis
Association of CSF glucose concentration with neurosyphilis diagnosis
Abstract The most specific criterion for diagnosing neurosyphilis is a reactive CSF VDRL. Unfortunately, there are in Europe, including Poland small number of specia...
A Comparative Study of Blood Glucose Level Measurement Between Glucometer and Semi Autoanalyser
A Comparative Study of Blood Glucose Level Measurement Between Glucometer and Semi Autoanalyser
Background; Glucometers are the excellent tools for self-monitoring of blood glucose (SMBG). Glucometer is widely used at hospitals and homes as a first line tool to get an idea ab...
Differentiating neurosarcoidosis from multiple sclerosis using combined analysis of basic CSF parameters and MRZ reaction
Differentiating neurosarcoidosis from multiple sclerosis using combined analysis of basic CSF parameters and MRZ reaction
BackgroundNeurosarcodosis is one of the most frequent differential diagnoses of multiple sclerosis (MS) and requires central nervous system (CNS) biopsy to establish definite diagn...

Back to Top