Refractive prediction by various intraocular lens formulas using optical biometry and effect of ocular parameters on their accuracy
View through CrossRef
Purpose:
To assess the prediction accuracy of intraocular lens (IOL) formulas and study the effect of axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), and lens thickness (LT) on the accuracy of formulas using optic biometry.
Methods:
This study was performed on 164 eyes of 164 patients who underwent uneventful cataract surgery. Ocular biometry values were measured using Lenstar-900, and intraocular lens (IOL) power was calculated using the SRK/T, SRK II, Hoffer Q, Holladay 2, and Barrett Universal II formulas. We evaluated the extent of bias within each formula for different ocular biometric measurements and explored the relationship between the prediction error and the ocular parameters by using various IOL formulas.
Results:
The summarization of refractive prediction error and absolute prediction error for each IOL formulation was performed after adjusting the mean refractive error to zero. The deviation in the error values was minimum for SRK/T (0.265) followed by Holladay 2 (0.327) and Barret (0.382). Further, SRK/T had the lowest median (0.15) and mean (0.198) absolute error as compared to other formulations. For the above formulations, 100% of the eyes were in the diopter range of ±1.0. It was observed that the overall distribution of error was closer to zero for SRK/T, followed by Holladay 2 and then Barrett.
Conclusion:
In summary, we found that accuracy was better in SRK/T formula. We achieved a better understanding of how each variable in the formulas is relatively weighed and the influencing factors in the refraction prediction.
Title: Refractive prediction by various intraocular lens formulas using optical biometry and effect of ocular parameters on their accuracy
Description:
Purpose:
To assess the prediction accuracy of intraocular lens (IOL) formulas and study the effect of axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), and lens thickness (LT) on the accuracy of formulas using optic biometry.
Methods:
This study was performed on 164 eyes of 164 patients who underwent uneventful cataract surgery.
Ocular biometry values were measured using Lenstar-900, and intraocular lens (IOL) power was calculated using the SRK/T, SRK II, Hoffer Q, Holladay 2, and Barrett Universal II formulas.
We evaluated the extent of bias within each formula for different ocular biometric measurements and explored the relationship between the prediction error and the ocular parameters by using various IOL formulas.
Results:
The summarization of refractive prediction error and absolute prediction error for each IOL formulation was performed after adjusting the mean refractive error to zero.
The deviation in the error values was minimum for SRK/T (0.
265) followed by Holladay 2 (0.
327) and Barret (0.
382).
Further, SRK/T had the lowest median (0.
15) and mean (0.
198) absolute error as compared to other formulations.
For the above formulations, 100% of the eyes were in the diopter range of ±1.
It was observed that the overall distribution of error was closer to zero for SRK/T, followed by Holladay 2 and then Barrett.
Conclusion:
In summary, we found that accuracy was better in SRK/T formula.
We achieved a better understanding of how each variable in the formulas is relatively weighed and the influencing factors in the refraction prediction.
Related Results
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Comparison of Ultrasound Biometry with Optical Biometry for Measurement of Axial Length and Calculation of Intraocular Lens Power in Patients Undergoing Routine Cataract Surgery
Objective: To compare the axial length and intraocular lens power calculated by using ultrasound biometry and optical biometry in patients with cataract
Methodology: It was prospec...
Comparison of four formulas for the intraocular lens power prediction
Comparison of four formulas for the intraocular lens power prediction
Abstract
Purpose
To assess the accuracy of four formulas for intraocular lens (IOL) power prediction in cataractous eyes.
METHODS
In this prospective study, 51 eyes of 3...
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Comparative Analysis of Refractive Outcomes Using Optical and Ultrasound Biometry in Phacoemulsification Cataract Surgery
Background: Accurate biometry is critical for optimal visual outcomes in cataract surgery. Both optical and ultrasound biometry are widely used, but their comparative effectiveness...
OptimEYEzing Emergency Skills: A Novel Model for Ocular Procedural Education for Emergency Medicine Residents
OptimEYEzing Emergency Skills: A Novel Model for Ocular Procedural Education for Emergency Medicine Residents
Audience: This model for ocular procedural education is designed to instruct emergency medicine residents of all levels of training. Introduction: Ocular complaints are a common pr...
Refractive Outcomes after Cataract Surgery
Refractive Outcomes after Cataract Surgery
A post-operative manifest refractive error as close as possible to target is key when performing cataract surgery with intraocular lens (IOL) implantation, given that residual asti...
Prediction and Estimation of Postoperative Refractive Error in Phacoemulsification: Using Ultrasound A-Scan and Intra Ocular Lens Master
Prediction and Estimation of Postoperative Refractive Error in Phacoemulsification: Using Ultrasound A-Scan and Intra Ocular Lens Master
Background: This study aims to predict and estimate the postoperative refractive outcome in participants undergoing phacoemulsification using IOL Master and A-scan biometry.Methods...
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Impact and Correction of an Anterior Phakic Intraocular Lens on Swept-Source Optical Coherence Tomography Biometry
Purpose:
To evaluate the impact of anterior chamber phakic intraocular lens (pIOL) on swept-source optical coherence tomography (SS-OCT) biometric measurements and IOL ...
Morphometric OCT parameters of the lens under accommodative stimulus. Report 1. Assessment of age-related changes
Morphometric OCT parameters of the lens under accommodative stimulus. Report 1. Assessment of age-related changes
Introduction. The structural approach to studying the mechanism of accommodation and its age-related changes focuses on analyzing morphometric parameters (size, shape, and position...