Correcting geometric image distortions in slice‐based 4D‐MRI on the MR‐linac

View through CrossRef
PurposeThe importance of four‐dimensional‐magnetic resonance imaging (4D‐MRI) is increasing in guiding online plan adaptation in thoracic and abdominal radiotherapy. Many 4D‐MRI sequences are based on multislice two‐dimensional (2D) acquisitions which provide contrast flexibility. Intrinsic to MRI, however, are machine‐ and subject‐related geometric image distortions. Full correction of slice‐based 4D‐MRIs acquired on the Unity MR‐linac (Elekta AB, Stockholm, Sweden) is challenging, since through‐plane corrections are currently not available for 2D sequences. In this study, we implement a full three‐dimensional 3D correction and quantify the geometric and dosimetric effects of machine‐related (residual) geometric image distortions.MethodsA commercial three‐dimensional (3D) geometric QA phantom (Philips, Best, the Netherlands) was used to quantify the effect of gradient nonlinearity (GNL) and static‐field inhomogeneity (B0I) on geometric accuracy. Additionally, the effectiveness of 2D (in‐plane, machine‐generic), 3D (machine‐generic), and in‐house developed 3D (machine‐specific) corrections was investigated. Corrections were based on deformable vector fields derived from spherical harmonics coefficients. Three patients with oligometastases in the liver were scanned with axial 4D‐MRIs on our MR‐linac (total: 10 imaging sessions). For each patient, a step‐and‐shoot IMRT plan (3 × 20 Gy) was created based on the simulation mid‐position (midP)‐CT. The 4D‐MRIs were then warped into a daily midP‐MRI and geometrically corrected. Next, the treatment plan was adapted according to the position offset of the tumor between midP‐CT and the 3D‐corrected midP‐MRIs. The midP‐CT was also deformably registered to the daily midP‐MRIs (different corrections applied) to quantify the dosimetric effects of (residual) geometric image distortions.ResultsUsing phantom data, median GNL distortions were 0.58 mm (no correction), 0.42–0.48 mm (2D), 0.34 mm (3D), and 0.34 mm (3D), measured over a diameter of spherical volume (DSV) of 200 mm. Median B0I distortions were 0.09 mm for the same DSV. For DSVs up to 500 mm, through‐plane corrections are necessary to keep the median residual GNL distortion below 1 mm. 3D and 3D corrections agreed within 0.15 mm. 2D‐corrected images featured uncorrected through‐plane distortions of up to 21.11 mm at a distance of 20–25 cm from the machine’s isocenter. Based on the 4D‐MRI patient scans, the average external body contour distortions were 3.1 mm (uncorrected) and 1.2 mm (2D‐corrected), with maximum local distortions of 9.5 mm in the uncorrected images. No (residual) distortions were visible for the metastases, which were all located within 10 cm of the machine’s isocenter. The interquartile range (IQR) of dose differences between planned and daily dose caused by variable patient setup, patient anatomy, and online plan adaptation was 1.37 Gy/Fx for the PTV D95%. When comparing dose on 3D‐corrected with uncorrected (2D‐corrected) images, the IQR was 0.61 (0.31) Gy/Fx.ConclusionsGNL is the main machine‐related source of image distortions on the Unity MR‐linac. For slice‐based 4D‐MRI, a full 3D correction can be applied after respiratory sorting to maximize spatial fidelity. The machine‐specific 3D correction did not substantially reduce residual geometric distortions compared to the machine‐generic 3D correction for our MR‐linac. In our patients, dosimetric variations in the target not related to geometric distortions were larger than those caused by geometric distortions.
Title: Correcting geometric image distortions in slice‐based 4D‐MRI on the MR‐linac
Description:
PurposeThe importance of four‐dimensional‐magnetic resonance imaging (4D‐MRI) is increasing in guiding online plan adaptation in thoracic and abdominal radiotherapy.
Many 4D‐MRI sequences are based on multislice two‐dimensional (2D) acquisitions which provide contrast flexibility.
Intrinsic to MRI, however, are machine‐ and subject‐related geometric image distortions.
Full correction of slice‐based 4D‐MRIs acquired on the Unity MR‐linac (Elekta AB, Stockholm, Sweden) is challenging, since through‐plane corrections are currently not available for 2D sequences.
In this study, we implement a full three‐dimensional 3D correction and quantify the geometric and dosimetric effects of machine‐related (residual) geometric image distortions.
MethodsA commercial three‐dimensional (3D) geometric QA phantom (Philips, Best, the Netherlands) was used to quantify the effect of gradient nonlinearity (GNL) and static‐field inhomogeneity (B0I) on geometric accuracy.
Additionally, the effectiveness of 2D (in‐plane, machine‐generic), 3D (machine‐generic), and in‐house developed 3D (machine‐specific) corrections was investigated.
Corrections were based on deformable vector fields derived from spherical harmonics coefficients.
Three patients with oligometastases in the liver were scanned with axial 4D‐MRIs on our MR‐linac (total: 10 imaging sessions).
For each patient, a step‐and‐shoot IMRT plan (3 × 20 Gy) was created based on the simulation mid‐position (midP)‐CT.
The 4D‐MRIs were then warped into a daily midP‐MRI and geometrically corrected.
Next, the treatment plan was adapted according to the position offset of the tumor between midP‐CT and the 3D‐corrected midP‐MRIs.
The midP‐CT was also deformably registered to the daily midP‐MRIs (different corrections applied) to quantify the dosimetric effects of (residual) geometric image distortions.
ResultsUsing phantom data, median GNL distortions were 0.
58 mm (no correction), 0.
42–0.
48 mm (2D), 0.
34 mm (3D), and 0.
34 mm (3D), measured over a diameter of spherical volume (DSV) of 200 mm.
Median B0I distortions were 0.
09 mm for the same DSV.
For DSVs up to 500 mm, through‐plane corrections are necessary to keep the median residual GNL distortion below 1 mm.
3D and 3D corrections agreed within 0.
15 mm.
2D‐corrected images featured uncorrected through‐plane distortions of up to 21.
11 mm at a distance of 20–25 cm from the machine’s isocenter.
Based on the 4D‐MRI patient scans, the average external body contour distortions were 3.
1 mm (uncorrected) and 1.
2 mm (2D‐corrected), with maximum local distortions of 9.
5 mm in the uncorrected images.
No (residual) distortions were visible for the metastases, which were all located within 10 cm of the machine’s isocenter.
The interquartile range (IQR) of dose differences between planned and daily dose caused by variable patient setup, patient anatomy, and online plan adaptation was 1.
37 Gy/Fx for the PTV D95%.
When comparing dose on 3D‐corrected with uncorrected (2D‐corrected) images, the IQR was 0.
61 (0.
31) Gy/Fx.
ConclusionsGNL is the main machine‐related source of image distortions on the Unity MR‐linac.
For slice‐based 4D‐MRI, a full 3D correction can be applied after respiratory sorting to maximize spatial fidelity.
The machine‐specific 3D correction did not substantially reduce residual geometric distortions compared to the machine‐generic 3D correction for our MR‐linac.
In our patients, dosimetric variations in the target not related to geometric distortions were larger than those caused by geometric distortions.

Related Results

Hydatid Disease of The Brain Parenchyma: A Systematic Review
Hydatid Disease of The Brain Parenchyma: A Systematic Review
Abstarct Introduction Isolated brain hydatid disease (BHD) is an extremely rare form of echinococcosis. A prompt and timely diagnosis is a crucial step in disease management. This ...
SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
Purpose: To achieve real‐time image guided radiotherapy, a linac‐MR system is designed with the linac located on the symmetry axis of the MRI. Thus, an investigation was performed ...
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
In this study, the authors synthesise and analyse the re-sults of the 2019-2020 statistical survey on infrastructure and human resources of radiotherapy centers in Viet-nam, thereb...
Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
Purpose:Linac‐magnetic resonance (MR) systems have been proposed in order to achieve real‐time image guided radiotherapy. The design of a new linac‐MR system with the in‐line 6 MV ...
Unusual Presentation of Alice in Wonderland Syndrome: A Case Report and Literature Review
Unusual Presentation of Alice in Wonderland Syndrome: A Case Report and Literature Review
Abstract Introduction Alice in Wonderland Syndrome (AIWS) is an uncommon and frequently overlooked neuropsychiatric condition, marked by brief episodes of altered visual and somato...
Incorporating multislice imaging into x‐ray CT polymer gel dosimetry
Incorporating multislice imaging into x‐ray CT polymer gel dosimetry
Purpose:To evaluate multislice computed tomography (CT) scanning for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD) and...
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...

Back to Top