SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
View through CrossRef
Purpose: To achieve real‐time image guided radiotherapy, a linac‐MR system is designed with the linac located on the symmetry axis of the MRI. Thus, an investigation was performed in order to determine the performance of an in‐line 6 MV linac in longitudinal magnetic fields. Method and Materials: Magnetic fringe fields from the PARAmed openMR™ were modeled and included in the simulation of an in‐line 6 MV linac. The electron gun simulation was modeled in OPERA‐3d/SCALA. The linac waveguide emulating the Varian 600C was modeled in COMSOL and the resulting dosimetry from the linac model was calculated using BEAMnrcMP 2007 and DOSXYZnrc. Results: With the linac on the symmetry axis of the openMR™ imager, the electron gun is expected to experience longitudinal magnetic fields ranging from 0.0022 to 0.011 T. At these field strengths, the rms emittance of the electron beam injected into the waveguide grew from 0.358 pi‐mm‐mrad at 0T to 3.255 pi‐mm‐mrad at 0.011 T. The increased emittance resulted from an increased beam diameter and an increasingly diverging beam. Upon injection, the larger emittance caused greater beam loss within the waveguide and a two orders of magnitude increase in the full‐width‐half‐maximum of the focal spot at the target. Despite the changes in the focal spot, the mean energy was altered by less than 0.5% and the calculated depth dose and profile distributions were identical to distributions at 0T. Magnetic fields up to 0.2T were also added to the linac simulation representing fringe fields from other MRIs. Both the injection current and target current reached a minimum at 0.06T, but increased slightly at higher field strengths due to increasing collimation in the electron gun anode. Conclusion: The effect of longitudinal magnetic fields on a simulated in‐ line 6 MV linac has been quantified enabling future development of a parallel linac‐MR system.
Title: SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
Description:
Purpose: To achieve real‐time image guided radiotherapy, a linac‐MR system is designed with the linac located on the symmetry axis of the MRI.
Thus, an investigation was performed in order to determine the performance of an in‐line 6 MV linac in longitudinal magnetic fields.
Method and Materials: Magnetic fringe fields from the PARAmed openMR™ were modeled and included in the simulation of an in‐line 6 MV linac.
The electron gun simulation was modeled in OPERA‐3d/SCALA.
The linac waveguide emulating the Varian 600C was modeled in COMSOL and the resulting dosimetry from the linac model was calculated using BEAMnrcMP 2007 and DOSXYZnrc.
Results: With the linac on the symmetry axis of the openMR™ imager, the electron gun is expected to experience longitudinal magnetic fields ranging from 0.
0022 to 0.
011 T.
At these field strengths, the rms emittance of the electron beam injected into the waveguide grew from 0.
358 pi‐mm‐mrad at 0T to 3.
255 pi‐mm‐mrad at 0.
011 T.
The increased emittance resulted from an increased beam diameter and an increasingly diverging beam.
Upon injection, the larger emittance caused greater beam loss within the waveguide and a two orders of magnitude increase in the full‐width‐half‐maximum of the focal spot at the target.
Despite the changes in the focal spot, the mean energy was altered by less than 0.
5% and the calculated depth dose and profile distributions were identical to distributions at 0T.
Magnetic fields up to 0.
2T were also added to the linac simulation representing fringe fields from other MRIs.
Both the injection current and target current reached a minimum at 0.
06T, but increased slightly at higher field strengths due to increasing collimation in the electron gun anode.
Conclusion: The effect of longitudinal magnetic fields on a simulated in‐ line 6 MV linac has been quantified enabling future development of a parallel linac‐MR system.
Related Results
Parametric modelling & digital manufacturing for better bra fit
Parametric modelling & digital manufacturing for better bra fit
<p>This research focuses on developing an overall system for a more accurate bra design and fit. The bra, consequently, becomes part of a life-long service system that is abl...
Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
Purpose:Linac‐magnetic resonance (MR) systems have been proposed in order to achieve real‐time image guided radiotherapy. The design of a new linac‐MR system with the in‐line 6 MV ...
Comparative Morphological Evaluation of Young Women’s Breast-Bra Reshaping by Different Bra Cups
Comparative Morphological Evaluation of Young Women’s Breast-Bra Reshaping by Different Bra Cups
Female breasts are regarded as a factor reflecting women’s morphological beauty. An appropriate bra can fulfill aesthetic needs, thus boosting self-esteem. This study proposed a me...
Sports Bra Pressure: Effect on Core Body Temperature and Comfort Sensation
Sports Bra Pressure: Effect on Core Body Temperature and Comfort Sensation
Background: Sports bras are engineer designed to enhance sports performance, which means that they need to provide an excellent fit, and offer adequate support and protection of th...
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
In this study, the authors synthesise and analyse the re-sults of the 2019-2020 statistical survey on infrastructure and human resources of radiotherapy centers in Viet-nam, thereb...
Abstract P4-13-15: A retrospective head-to-head comparison between TiLoop Bra/TiMesh and seragyn in 320 cases of reconstructive breast surgery
Abstract P4-13-15: A retrospective head-to-head comparison between TiLoop Bra/TiMesh and seragyn in 320 cases of reconstructive breast surgery
Abstract
Background: As an alternative option to ADMs (acellular dermal matrices) synthetic meshes are a cheaper and equally effective option in implanted breast rec...
Magnetic cloak made of NdFeB permanent magnetic material
Magnetic cloak made of NdFeB permanent magnetic material
In the past few years, the concept of an electromagnetic invisibility cloak has received much attention. Based on the pioneering theoretical work, invisibility cloaks have been gre...
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...