Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
View through CrossRef
Purpose:Linac‐magnetic resonance (MR) systems have been proposed in order to achieve real‐time image guided radiotherapy. The design of a new linac‐MR system with the in‐line 6 MV linac generating x‐rays along the symmetry axis of an open MR imager is outlined. This new design allows for a greater MR field strength to achieve better quality images while reducing hot and cold spots in treatment planning. An investigation of linac's performance in the longitudinal fringe magnetic fields of the MR imager is given.Methods:The open MR imager fringe magnetic field was modeled using the analytic solution of the magnetic field generated from current carrying loops. The derived solution was matched to the magnetic fringe field isolines provided for a 0.5 T open MR imager through Monte Carlo optimization. The optimized field solution was then added to the previously validated 6 MV linac simulation to quantify linac's performance in the fringe magnetic field of a 0.5 T MR imager. To further the investigation, linac's performance in large fringe fields expected from other imagers was investigated through the addition of homogeneous longitudinal fields.Results:The Monte Carlo optimization of the analytic current loop solution provided good agreement with the magnetic fringe field isolines supplied by the manufacturer. The range of magnetic fields the linac is expected to experience when coupled to the 0.5 T MR imager was determined to be from 0.0022 to 0.011 T (as calculated at the electron gun cathode). The effect of the longitudinal magnetic field on the electron beam was observed to be only in the electron gun. The longitudinal field changed the electron gun optics, affecting beam characteristics, such as a slight increase in the injection current and beam diameter, and an increasingly nonlaminar transverse phase space. Although the target phase space showed little change in its energy spectrum from the altered injection phase space, a reduction in the target current and spatial distribution peak intensity was observed. Despite these changes, the target phase space had little effect on the depth dose curves or dose profiles calculated for a field at 1.5 cm depth. At longitudinal fields larger than 0.012 T, a drastic reduction in the injection current from the electron gun was observed due to a large fraction of electrons striking the anode. This further reduced the target current, which reached a minimum of at 0.06 T. A slow increase in the injection and target currents was observed at fields larger than 0.06 T due to greater beam collimation in the anode beam tube.Conclusions:In an effort to achieve higher quality images and a reduction in hot and cold spots in the treatment plan, a parallel configuration linac‐MR system is presented. The longitudinal magnetic fields of the MR imager caused large beam losses within the electron gun. These losses may be eliminated through a redesign of the electron gun optics incorporating a longitudinal magnetic field, or through magnetic shielding, which has already been proven successful for the transverse configuration.
Title: Effect of longitudinal magnetic fields on a simulated in‐line 6 MV linac
Description:
Purpose:Linac‐magnetic resonance (MR) systems have been proposed in order to achieve real‐time image guided radiotherapy.
The design of a new linac‐MR system with the in‐line 6 MV linac generating x‐rays along the symmetry axis of an open MR imager is outlined.
This new design allows for a greater MR field strength to achieve better quality images while reducing hot and cold spots in treatment planning.
An investigation of linac's performance in the longitudinal fringe magnetic fields of the MR imager is given.
Methods:The open MR imager fringe magnetic field was modeled using the analytic solution of the magnetic field generated from current carrying loops.
The derived solution was matched to the magnetic fringe field isolines provided for a 0.
5 T open MR imager through Monte Carlo optimization.
The optimized field solution was then added to the previously validated 6 MV linac simulation to quantify linac's performance in the fringe magnetic field of a 0.
5 T MR imager.
To further the investigation, linac's performance in large fringe fields expected from other imagers was investigated through the addition of homogeneous longitudinal fields.
Results:The Monte Carlo optimization of the analytic current loop solution provided good agreement with the magnetic fringe field isolines supplied by the manufacturer.
The range of magnetic fields the linac is expected to experience when coupled to the 0.
5 T MR imager was determined to be from 0.
0022 to 0.
011 T (as calculated at the electron gun cathode).
The effect of the longitudinal magnetic field on the electron beam was observed to be only in the electron gun.
The longitudinal field changed the electron gun optics, affecting beam characteristics, such as a slight increase in the injection current and beam diameter, and an increasingly nonlaminar transverse phase space.
Although the target phase space showed little change in its energy spectrum from the altered injection phase space, a reduction in the target current and spatial distribution peak intensity was observed.
Despite these changes, the target phase space had little effect on the depth dose curves or dose profiles calculated for a field at 1.
5 cm depth.
At longitudinal fields larger than 0.
012 T, a drastic reduction in the injection current from the electron gun was observed due to a large fraction of electrons striking the anode.
This further reduced the target current, which reached a minimum of at 0.
06 T.
A slow increase in the injection and target currents was observed at fields larger than 0.
06 T due to greater beam collimation in the anode beam tube.
Conclusions:In an effort to achieve higher quality images and a reduction in hot and cold spots in the treatment plan, a parallel configuration linac‐MR system is presented.
The longitudinal magnetic fields of the MR imager caused large beam losses within the electron gun.
These losses may be eliminated through a redesign of the electron gun optics incorporating a longitudinal magnetic field, or through magnetic shielding, which has already been proven successful for the transverse configuration.
Related Results
SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
SU‐C‐BRA‐03: Investigation of the Performance of a Simulated In‐Line 6 MV Linac in Longitudinal Magnetic Fields
Purpose: To achieve real‐time image guided radiotherapy, a linac‐MR system is designed with the linac located on the symmetry axis of the MRI. Thus, an investigation was performed ...
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
Analysis of radiotherapy technology infrastructure and human resources in Vietnam
In this study, the authors synthesise and analyse the re-sults of the 2019-2020 statistical survey on infrastructure and human resources of radiotherapy centers in Viet-nam, thereb...
Analysis of magnetohydrodynamic drag character for hypersonic vehicles
Analysis of magnetohydrodynamic drag character for hypersonic vehicles
In hypersonic flight, a very high temperature area can form ahead of the nose of aerocraft due to the shock aerodynamic heating, which leads to air weakly ionized. Many researchers...
Magnetic cloak made of NdFeB permanent magnetic material
Magnetic cloak made of NdFeB permanent magnetic material
In the past few years, the concept of an electromagnetic invisibility cloak has received much attention. Based on the pioneering theoretical work, invisibility cloaks have been gre...
Automation of the Monte Carlo simulation of medical linear accelerators
Automation of the Monte Carlo simulation of medical linear accelerators
The main result of this thesis is a software system, called PRIMO, which simulates clinical linear accelerators and the subsequent dose distributions using the Monte Carlo method. ...
Development of a 6 MeV electron beam energy Linac for fruit sterilization
Development of a 6 MeV electron beam energy Linac for fruit sterilization
Abstract
Treatment of fresh fruits, meat, and food with ionizing radiation has been frequently chosen to destroy harmful pathogenic microorganisms and oriental fruit...
Multifunctional linear induction motor with longitudinal-transverce magnetic flux for magnetic-levitational transport
Multifunctional linear induction motor with longitudinal-transverce magnetic flux for magnetic-levitational transport
Background: Traction linear induction motors (LIM) at the current stage of human society development are the most promising for high-speed magnetic-levitation transport (MLT) and a...
Localized nanoscale induction by single domain magnetic particles
Localized nanoscale induction by single domain magnetic particles
AbstractSingle domain magnetic nanoparticles are increasingly investigated as actuators of biological and chemical processes that respond to externally applied magnetic fields. Alt...