Calibration HPGe detector using IAEA-U source for CBRNe

View through CrossRef
AbstractDetecting and preventing the illicit movement of radioactive materials within a country is crucial, requiring the identification of radiologic anomalies against the normal radiation background. High-purity germanium (HPGe) detectors, known for their precision and sensitivity, have become popular for analyzing radioactive materials in CBRNe scenarios. This study focused on calibrating an HPGe detector for CBRNe applications, using gamma-ray spectra from an IAEA-U reference source. Energy calibration involved identifying peaks in the spectra and creating a calibration curve for energy and channel number data. Efficiency calibration, determined using the known activity of the source, revealed a linear relationship between energy and detector response. Over four years, systematic efficiency calibrations showed a deviation of only 3% well below the recommended limit of 5%. These findings underscore the reliability of the system as a reference spectrometry method for accurate detection of radioactive materials. Graphical abstract
Title: Calibration HPGe detector using IAEA-U source for CBRNe
Description:
AbstractDetecting and preventing the illicit movement of radioactive materials within a country is crucial, requiring the identification of radiologic anomalies against the normal radiation background.
High-purity germanium (HPGe) detectors, known for their precision and sensitivity, have become popular for analyzing radioactive materials in CBRNe scenarios.
This study focused on calibrating an HPGe detector for CBRNe applications, using gamma-ray spectra from an IAEA-U reference source.
Energy calibration involved identifying peaks in the spectra and creating a calibration curve for energy and channel number data.
Efficiency calibration, determined using the known activity of the source, revealed a linear relationship between energy and detector response.
Over four years, systematic efficiency calibrations showed a deviation of only 3% well below the recommended limit of 5%.
These findings underscore the reliability of the system as a reference spectrometry method for accurate detection of radioactive materials.
Graphical abstract.

Related Results

STRATEGY OF THE ARMY NUBIKA ZENI COMPANY TO FACE THE COVID-19 PANDEMIC
STRATEGY OF THE ARMY NUBIKA ZENI COMPANY TO FACE THE COVID-19 PANDEMIC
Abstract-The threat of pandemics and epidemics of dangerous and deadly diseases is very likely to occur in Indonesia. The spread of Covid-19 that is out of control and exceeds the ...
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
Long-term Evaluation of HPGe Calibration for Environmental Radioactivity Assessment Using IAEA-U and IAEA-Th Sources
Long-term Evaluation of HPGe Calibration for Environmental Radioactivity Assessment Using IAEA-U and IAEA-Th Sources
The investigation of the dynamics between environmental radioactivity and its implications for human health stands as a fundamental pursuit in contemporary scientific research. Emp...
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Conceptual design report of the MPD Cosmic Ray Detector (MCORD)
Abstract This report presents a concept of constructing a detector dedicated for detection of muons observed during measurements carried out at the MPD (Multi-Pu...
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
Characterization of a novel HgCdTe focal plane array for ground and space astronomy through innovative infrared setups
(English) Nowadays, mercury-cadmium-telluride (MCT) short-wave infrared (SWIR) detectors are widely used in cutting-edge space missions and ground-based telescopes. They take adva...
Study on the virtual source calibration technology based on the volume of radioactive gas source
Study on the virtual source calibration technology based on the volume of radioactive gas source
The calibration methods for the radioactive Kr and Xe gases produce the key data for the judgment of nuclear fission reaction, whose accurate measurement has always been a difficul...
Neutron holography simulation based on different sample rotations
Neutron holography simulation based on different sample rotations
Neutron holography is a new imaging technique based on the recording of the interference pattern of two coherent waves emitted by the same source, which allows observing the spatia...
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-S...

Back to Top