ANALISIS PENAHAN RADIASI RUANGAN RADIOLOGI INTERVENSI CATHLAB SEBAGAI UPAYA PROTEKSI KESELAMATAN RADIASI DI RSUP. H. ADAM MALIK MEDAN
View through CrossRef
The exposure value received does not exceed the dose limit value (NBD) set by the regulation of the Nuclear Energy Supervisory Agency (BAPETEN) recommended by the National Council on Radiation Protection and Measurement (NCRP). This is the goal of radiation protection in achieving radiation safety for workers, patients and the public. To achieve this, it is necessary to evaluate the radiation-retaining structural walls of the interventional radiology room, X-ray fluoroscopy angiography, GE and Philips brands, Cathlab 1 and 2, respectively, at Adam Malik Hospital by comparing the data obtained from field calculations with the recommendations of regulators or theoretical. The results obtained are the thickness of the radiation retaining walls of the GE and Philips rooms by calculating the transmission factor and the thickness of the secondary radiation shield for each wall of the GE and Philips Cathlab rooms with a secondary air kerma value of 1 meter 3.8, the largest Xbarrier is 120.2 mm concrete, respectively. and 1.75mmPb and 81.13mmconcrete and 0.12mmPb. Using real calculations, the secondary air kerma value is 0.003 for the GE room, it is obtained that the concrete Xbarrier and Pb are minus values and the Philips air kerma value is 0.189, the Xbarrier is 22.1mm of concrete and 0.24mmPb. Meanwhile, the currently installed wall thickness is 28 cm of concrete and 0.5mmPb, with a total of 25 patients per week. On the other hand, exposure to environmental radiation, leakage of X-ray aircraft is also within safe limits. The conclusion is that the shield retaining installed both concrete and Pb on each wall of the X-ray inspection room of Cathlab 1 GE and Cathlab 2 Philips, Adam Malik General Hospital is in a safe condition for intervention, and radiation exposure measured in the environment as well as leakage of the light aircraft. -X is also within safe limits, so that the purpose of radiation shielding can be achieved, namely radiation exposure received by both officers, patients and the public does not exceed the specified dose limit value. Keywords: Radiation safety, dose limit value, radiation barrier, interventional radiology.
Badan Pengawas Tenaga Nuklir (BAPETEN)
Title: ANALISIS PENAHAN RADIASI RUANGAN RADIOLOGI INTERVENSI CATHLAB SEBAGAI UPAYA PROTEKSI KESELAMATAN RADIASI DI RSUP. H. ADAM MALIK MEDAN
Description:
The exposure value received does not exceed the dose limit value (NBD) set by the regulation of the Nuclear Energy Supervisory Agency (BAPETEN) recommended by the National Council on Radiation Protection and Measurement (NCRP).
This is the goal of radiation protection in achieving radiation safety for workers, patients and the public.
To achieve this, it is necessary to evaluate the radiation-retaining structural walls of the interventional radiology room, X-ray fluoroscopy angiography, GE and Philips brands, Cathlab 1 and 2, respectively, at Adam Malik Hospital by comparing the data obtained from field calculations with the recommendations of regulators or theoretical.
The results obtained are the thickness of the radiation retaining walls of the GE and Philips rooms by calculating the transmission factor and the thickness of the secondary radiation shield for each wall of the GE and Philips Cathlab rooms with a secondary air kerma value of 1 meter 3.
8, the largest Xbarrier is 120.
2 mm concrete, respectively.
and 1.
75mmPb and 81.
13mmconcrete and 0.
12mmPb.
Using real calculations, the secondary air kerma value is 0.
003 for the GE room, it is obtained that the concrete Xbarrier and Pb are minus values and the Philips air kerma value is 0.
189, the Xbarrier is 22.
1mm of concrete and 0.
24mmPb.
Meanwhile, the currently installed wall thickness is 28 cm of concrete and 0.
5mmPb, with a total of 25 patients per week.
On the other hand, exposure to environmental radiation, leakage of X-ray aircraft is also within safe limits.
The conclusion is that the shield retaining installed both concrete and Pb on each wall of the X-ray inspection room of Cathlab 1 GE and Cathlab 2 Philips, Adam Malik General Hospital is in a safe condition for intervention, and radiation exposure measured in the environment as well as leakage of the light aircraft.
-X is also within safe limits, so that the purpose of radiation shielding can be achieved, namely radiation exposure received by both officers, patients and the public does not exceed the specified dose limit value.
Keywords: Radiation safety, dose limit value, radiation barrier, interventional radiology.
Related Results
Analisis Keselamatan Radiasi pada Ruangan Konvensional di Unit Radiologi RSU. Kertha Usada Singaraja
Analisis Keselamatan Radiasi pada Ruangan Konvensional di Unit Radiologi RSU. Kertha Usada Singaraja
Keselamatan Radiasi Pengion pada Bidang Medik yang dianggap Keselamatan Radiasi merupakan tindakan yang dilakukan untuk melindungi pasien, pekerja, anggota masyarakat serta lingkun...
Analisis Proteksi Radiasi bagi Pekerja Radiasi di Instalasi Radiologi Rumah Sakit Universitas Hasanuddin Makassar
Analisis Proteksi Radiasi bagi Pekerja Radiasi di Instalasi Radiologi Rumah Sakit Universitas Hasanuddin Makassar
Proteksi radiasi adalah cabang ilmu pengetahuan yang berkaitan dengan teknik kesehatan lingkungan yaitu tentang proteksi yang perlu di berikan kepada seseorang atau sekelompok oran...
Evaluasi Penerapan Proteksi Radiasi pada Pekerja Radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri, dan RS UNAND
Evaluasi Penerapan Proteksi Radiasi pada Pekerja Radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri, dan RS UNAND
Telah dilakukan penelitian tentang penerapan proteksi radiasi pada pekerja radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri dan RS UNAND. Penelitian bertujuan untuk menge...
ANALISIS PENERAPAN KESELAMATAN RADIASI SINAR-X PADA PEKERJA RADIASI DI INSTALASI RADIOLOGI RUMAH SAKIT PEKANBARU MEDICAL CENTER (PMC) TAHUN 2020
ANALISIS PENERAPAN KESELAMATAN RADIASI SINAR-X PADA PEKERJA RADIASI DI INSTALASI RADIOLOGI RUMAH SAKIT PEKANBARU MEDICAL CENTER (PMC) TAHUN 2020
Keselamatan radiasi adalah tindakan yang dilakukan untuk melindungi pasien, pekerja, masyarakat, dan lingkungan hidup dari bahaya radiasi. Mengingat potensi bahaya radiasi yang bes...
Gambaran Peralatan Proteksi Radiasi di Intalasi Radiologi Rumah Sakit Khusus Bedah Hasta Husada
Gambaran Peralatan Proteksi Radiasi di Intalasi Radiologi Rumah Sakit Khusus Bedah Hasta Husada
Radiasi pengion adalah jenis energi yang dilepaskan oleh atom yang merambat dalam bentuk gelombang elektromagnetik atau partikel. Keselamatan terhadap radiasi pengion adalah salah ...
Kajian Aspek Keselamatan Kegiatan Pelimbahan Sumber Radiasi Co-60 Bekas Penggunaan di Iradiator Karet Alam (IRKA)
Kajian Aspek Keselamatan Kegiatan Pelimbahan Sumber Radiasi Co-60 Bekas Penggunaan di Iradiator Karet Alam (IRKA)
Iradiator Karet Alam (IRKA) milik Pusat Riset dan Teknologi Aplikasi Isotop dan Radiasi (PRTAIR – BRIN) digunakan sebagai fasilitas radiasi untuk penelitian dan layanan aplikasi is...
JARAK AMAN PETUGAS SAAT PEMERIKSAAN THORAX DI RUANG CRITICAL UNIT PEMERIKSAAN RSUD PROF DR MARGONO SOEKARJO
JARAK AMAN PETUGAS SAAT PEMERIKSAAN THORAX DI RUANG CRITICAL UNIT PEMERIKSAAN RSUD PROF DR MARGONO SOEKARJO
Pemeriksaan radiologi onsite atau beberapa sering disebut cito-bed adalah pemeriksaan yang dilakukan di ruangan non radiologi untuk kasus tertentu dengan menggunakan sinar-x mobile...
Kajian Penetapan Nilai Pembatas Dosis di Pusat Riset dan Teknologi Aplikasi Isotop dan Radiasi (PRTAIR) – Badan Riset dan Inovasi Nasional (BRIN)
Kajian Penetapan Nilai Pembatas Dosis di Pusat Riset dan Teknologi Aplikasi Isotop dan Radiasi (PRTAIR) – Badan Riset dan Inovasi Nasional (BRIN)
Pusat Riset dan Teknologi Aplikasi Isotop dan Radiasi (PRTAIR) melaksanakan beragam kegiatan penelitian, pengembangan, penerapan, dan pengkajian (litbangjirap) yang memanfaatkan is...