Use of an active esophageal cooling device in zero-fluoroscopy settings without intracardiac echocardiography
View through CrossRef
Abstract
Background
Active esophageal cooling is increasingly being utilized during radiofrequency (RF) ablation to achieve pulmonary vein isolation (PVI) for the treatment of atrial fibrillation (AF). In addition to data showing decreases in severe esophageal injury with active esophageal cooling, placement of a commercially available cooling device can be identified on intracardiac echocardiography (ICE), allowing implementation in zero-fluoroscopy settings. In the case of procedures that do not have ICE available, an alternative method to determine proper placement of an esophageal cooling device is needed in zero-fluoroscopy settings..
Purpose
To describe a large single-center approach to visualizing an active esophageal cooling device into a 3D mapping system.
Methods
We reviewed data on patients treated with RF ablation for PVI procedures over the time frame 1/1/2020 to 12/31/21. Active esophageal cooling was phased in towards the end of 2020. For all cases, transseptal puncture was performed with the use of transesophageal echocardiography (TEE). After successful transseptal puncture, the TEE probe was removed, and either a single-sensor LET probe or an active esophageal cooling device was placed. For the active cooling device, the distal tip was cut, and an SL-1 (0.032 inch, 150 cm length) guidewire was placed through the central lumen of the cooling device (used for gastric suctioning and enteral feeding). The guidewire was pinned via a pin block to the 3D mapping system (EnSite, Abbott). Unipolar configuration was used to generate a 3D map, which was then visualized during device placement. Placement was confirmed after visualizing the guidewire tip on the 3D map passing below the coronary sinus (Figure 1).
Results
A total of 417 cases were performed over the study timeframe. The mean age of patients was 65±10 years, and 40% were female. A total of 5 complications were recorded (3 pericardial effusions, 1 pseudoaneurysm, and 1 air embolism). A total of 156 patients received LET monitoring, and 261 received active esophageal cooling. The mean procedure length was 109±24 minutes for cases utilizing LET monitoring, and 93±11 minutes for cases utilizing active esophageal cooling. Visualization of the active esophageal cooling device on the cardiac mapping system was possible in all cases.
Conclusions
We describe here the largest series to date utilizing active esophageal cooling in a zero-fluoroscopy, zero-ICE setting. With increasing efforts to reduce fluoroscopy, this approach allows pursuit of zero-fluoroscopy even in systems without the availability of ICE, while maintaining procedural speed and safety.
Funding Acknowledgement
Type of funding sources: Private company. Main funding source(s): Attune Medical
Oxford University Press (OUP)
Title: Use of an active esophageal cooling device in zero-fluoroscopy settings without intracardiac echocardiography
Description:
Abstract
Background
Active esophageal cooling is increasingly being utilized during radiofrequency (RF) ablation to achieve pulmonary vein isolation (PVI) for the treatment of atrial fibrillation (AF).
In addition to data showing decreases in severe esophageal injury with active esophageal cooling, placement of a commercially available cooling device can be identified on intracardiac echocardiography (ICE), allowing implementation in zero-fluoroscopy settings.
In the case of procedures that do not have ICE available, an alternative method to determine proper placement of an esophageal cooling device is needed in zero-fluoroscopy settings.
Purpose
To describe a large single-center approach to visualizing an active esophageal cooling device into a 3D mapping system.
Methods
We reviewed data on patients treated with RF ablation for PVI procedures over the time frame 1/1/2020 to 12/31/21.
Active esophageal cooling was phased in towards the end of 2020.
For all cases, transseptal puncture was performed with the use of transesophageal echocardiography (TEE).
After successful transseptal puncture, the TEE probe was removed, and either a single-sensor LET probe or an active esophageal cooling device was placed.
For the active cooling device, the distal tip was cut, and an SL-1 (0.
032 inch, 150 cm length) guidewire was placed through the central lumen of the cooling device (used for gastric suctioning and enteral feeding).
The guidewire was pinned via a pin block to the 3D mapping system (EnSite, Abbott).
Unipolar configuration was used to generate a 3D map, which was then visualized during device placement.
Placement was confirmed after visualizing the guidewire tip on the 3D map passing below the coronary sinus (Figure 1).
Results
A total of 417 cases were performed over the study timeframe.
The mean age of patients was 65±10 years, and 40% were female.
A total of 5 complications were recorded (3 pericardial effusions, 1 pseudoaneurysm, and 1 air embolism).
A total of 156 patients received LET monitoring, and 261 received active esophageal cooling.
The mean procedure length was 109±24 minutes for cases utilizing LET monitoring, and 93±11 minutes for cases utilizing active esophageal cooling.
Visualization of the active esophageal cooling device on the cardiac mapping system was possible in all cases.
Conclusions
We describe here the largest series to date utilizing active esophageal cooling in a zero-fluoroscopy, zero-ICE setting.
With increasing efforts to reduce fluoroscopy, this approach allows pursuit of zero-fluoroscopy even in systems without the availability of ICE, while maintaining procedural speed and safety.
Funding Acknowledgement
Type of funding sources: Private company.
Main funding source(s): Attune Medical.
Related Results
Significant procedure length reduction when utilizing active esophageal cooling – a six-year single center study
Significant procedure length reduction when utilizing active esophageal cooling – a six-year single center study
Abstract
Background
A number of factors can influence procedure duration of left atrial ablation for the treatment of atrial fib...
The analysis on Tiam2 for expression in esophageal carcinoma: A descriptive study
The analysis on Tiam2 for expression in esophageal carcinoma: A descriptive study
Rationale:
To investigate T lymphoma invasion and metastasis inducing factor 2 (Tiam2) protein for expression in esophageal carcinoma and relationship with clinical fea...
Reduction in procedure duration when utilizing active esophageal cooling in patients with persistent atrial fibrillation - update from a large hospital registry
Reduction in procedure duration when utilizing active esophageal cooling in patients with persistent atrial fibrillation - update from a large hospital registry
Abstract
Funding Acknowledgements
Type of funding sources: Public grant(s) – National budget only. Main funding source(s): Natio...
Arrhythmia recurrence reduction with an active esophageal cooling device during radiofrequency ablation
Arrhythmia recurrence reduction with an active esophageal cooling device during radiofrequency ablation
Abstract
Funding Acknowledgements
Type of funding sources: None.
Background
...
Application of Machine Learning Based Meta Models for Predicting Film Cooling Effectiveness in Gas Turbine Blades
Application of Machine Learning Based Meta Models for Predicting Film Cooling Effectiveness in Gas Turbine Blades
Abstract
In Large Gas Turbines, turbine components in particular blades and vanes operate at significantly high temperatures. As a result, cooling of these component...
022. TREATMENT EXPERIENCE OF ESOPHAGEAL FISTULA INDUCED BY NEOADJUVANT THERAPY IN ESOPHAGEAL SQUAMOUS CELL CARCINOMA: A SINGLE-CENTER EXPERIENCE
022. TREATMENT EXPERIENCE OF ESOPHAGEAL FISTULA INDUCED BY NEOADJUVANT THERAPY IN ESOPHAGEAL SQUAMOUS CELL CARCINOMA: A SINGLE-CENTER EXPERIENCE
Abstract
Background
Esophageal squamous cell carcinoma (ESCC) is typically treated with neoadjuvant therapy followed by esophage...
ÖZOFAGUS ATREZİLİ YENİDOĞANLARDA ANESTEZİK RİSKLER, MORBİDİTE VE MORTALİTE
ÖZOFAGUS ATREZİLİ YENİDOĞANLARDA ANESTEZİK RİSKLER, MORBİDİTE VE MORTALİTE
Giriş:
Özofagus atrezisi (ÖA), özofagusun proksimal ve distal bölümlerinin bağlantısının olmamasıdır ve 3000-4500 canlı
doğumda bir görülür. Özofagus atrezisi ve trakeoözofageal f...
O15 ESOPHAGEAL MOTILITY OF LUNG TRANSPLANT RECIPIENTS DIAGNOSED WITH SYSTEMIC CONNECTIVE TISSUE DISORDER: A SINGLE-CENTER EXPERIENCE
O15 ESOPHAGEAL MOTILITY OF LUNG TRANSPLANT RECIPIENTS DIAGNOSED WITH SYSTEMIC CONNECTIVE TISSUE DISORDER: A SINGLE-CENTER EXPERIENCE
Abstract
Aim
To examine the esophageal motility of lung transplant recipients diagnosed with systemic connective tissue disorder...