HemoCue against Coulter LH-750 in the estimation of hemoglobin levels of blood donors in mobile collection settings: a comparative study

View through CrossRef
Background: The fast and outpatient setting for a determination of the hemoglobin (Hb) level is a well-recognized prerequisite to detect anemia in blood donors. This study aimed to evaluate the performance of the HemoCue methods (HemoCue B-Hb and HemoCue-301) against Coulter LH-750 as a reference method for Hb determination.Methods: This study was an experimental cross-sectional study. It includes 455 blood samples that were collected from volunteer blood donors between January 15, 2010 and February 15, 2011. The performance of the three methods and their comparisons were assessed using the analysis of coefficients of variation (CV), linear regression, and mean difference. Correlation coefficient and Bland–Altman plots were drawn to compare the two HemoCue measurements and the automated cell analyzer against each other and to evaluate their results. The Hb concentrations were compared using the concordance correlation coefficient.Results: The findings exhibited that the CV for the three methods Coulter LH-750, HemoCue B-Hb, and HemoCue-301 were 0.60%, 0.72%, and 0.92%, respectively. A statistically significant difference was observed between the means of the Hb measurements for the three methods (p<0.001). The HemoCue B-Hb and HemoCue-301 methods showed the best agreement, and the Coulter LH-750 method gave a lower Hb value compared with the two HemoCue methods. The results showed a positive correlation of HemoCue Hb results compared with the reference method.Conclusion: All three methods provide a good agreement for Hb determination. The new device HemoCue-301 was found to be more accurate compared with HemoCue B-Hb and Coulter LH-750.
Title: HemoCue against Coulter LH-750 in the estimation of hemoglobin levels of blood donors in mobile collection settings: a comparative study
Description:
Background: The fast and outpatient setting for a determination of the hemoglobin (Hb) level is a well-recognized prerequisite to detect anemia in blood donors.
This study aimed to evaluate the performance of the HemoCue methods (HemoCue B-Hb and HemoCue-301) against Coulter LH-750 as a reference method for Hb determination.
Methods: This study was an experimental cross-sectional study.
It includes 455 blood samples that were collected from volunteer blood donors between January 15, 2010 and February 15, 2011.
The performance of the three methods and their comparisons were assessed using the analysis of coefficients of variation (CV), linear regression, and mean difference.
Correlation coefficient and Bland–Altman plots were drawn to compare the two HemoCue measurements and the automated cell analyzer against each other and to evaluate their results.
The Hb concentrations were compared using the concordance correlation coefficient.
Results: The findings exhibited that the CV for the three methods Coulter LH-750, HemoCue B-Hb, and HemoCue-301 were 0.
60%, 0.
72%, and 0.
92%, respectively.
A statistically significant difference was observed between the means of the Hb measurements for the three methods (p<0.
001).
The HemoCue B-Hb and HemoCue-301 methods showed the best agreement, and the Coulter LH-750 method gave a lower Hb value compared with the two HemoCue methods.
The results showed a positive correlation of HemoCue Hb results compared with the reference method.
Conclusion: All three methods provide a good agreement for Hb determination.
The new device HemoCue-301 was found to be more accurate compared with HemoCue B-Hb and Coulter LH-750.

Related Results

[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED] Guardian Blood Balance –Feel the difference Guardian Blood Balance makes! v1
[RETRACTED]Guardian Blood Balance Reviews (Works Or Hoax) Does Guardian Botanicals Blood Balance AU Really Works? Read Updated Report! Diabetes and Hypertension is such a health p...
Coulter Method for the detection of Physiological parameters v1
Coulter Method for the detection of Physiological parameters v1
Instrument A Coulter counter [1] [2] is an apparatus for counting and sizing particles suspended in electrolytes. It is used for cells, bacteria, prokaryotic cells and virus parti...
Primerjalna književnost na prelomu tisočletja
Primerjalna književnost na prelomu tisočletja
In a comprehensive and at times critical manner, this volume seeks to shed light on the development of events in Western (i.e., European and North American) comparative literature ...
Capillary versus venous haemoglobin determination in the assessment of healthy blood donors
Capillary versus venous haemoglobin determination in the assessment of healthy blood donors
Background and ObjectivesTo determine the accuracy of fingerstick haemoglobin assessment in blood donors, the performance of a portable haemoglobinometer (HemoCue Hb 201+) was pros...
Follow-up of 12,559 Unrelated Donors of Peripheral Blood Stem Cells or Bone Marrow
Follow-up of 12,559 Unrelated Donors of Peripheral Blood Stem Cells or Bone Marrow
Abstract Abstract 365 Donor follow-up is indicated to detect potential long-term risks for allogeneic stem cell donors. We sent a follow-up questionna...
Serum Hepcidin Is an Early Predictor of Iron Depletion in Non-Anemic Blood Donors
Serum Hepcidin Is an Early Predictor of Iron Depletion in Non-Anemic Blood Donors
Introduction Iron deficiency is one of the major concerns in repeated whole blood donors. Although taking iron supplementation, frequent whole blood donors sometimes...
RHD Genotyping of Blood Donors May Avoid Anti-D Immunization.
RHD Genotyping of Blood Donors May Avoid Anti-D Immunization.
Abstract Antigen D compatible transfusion is standard practice in modern transfusion therapy, warranting proper antigen D typing in all blood donors. Blood group gen...

Back to Top