Sci‐Sat AM (2) Therapy‐08: Influence of time dose rate calibration on Siemens Virtual Wedge dosimetry
View through CrossRef
We have investigated the influence of time dose rate calibration on the dosimetry of Siemens Virtual Wedge (VW). The linear accelerator (LINAC) uses the ionization current generated in two dose monitoring ion chambers (MONITOR 1 and MONITOR 2) to dynamically control the three phases of VW delivery (initial gap, sweep portion, and open field). We intentionally offset the calibration of MONITOR 2, which controls the time dose rate delivered on the LINAC, by up to ± 3% and observed fractional changes in the wedge dose profile of approximately 10% for a 6MV beam with a 60° wedge at a field size of 20 × 20 cm2. We present an analytical expression to calculate the fractional change in wedge dosimetry as a function of time dose rate calibration of MONITOR 2. A comparison with measurement showed that the expression was accurate to within 1%. We will show how changes in the calibration of time dose rate (MONITOR 2) affect both the central axis wedge factor and wedge angle in order to assist in the diagnosis of wedge problems encountered in routine quality control measurements. The effects are most prominent at low beam energies, large wedge angles and large field sizes.
Title: Sci‐Sat AM (2) Therapy‐08: Influence of time dose rate calibration on Siemens Virtual Wedge dosimetry
Description:
We have investigated the influence of time dose rate calibration on the dosimetry of Siemens Virtual Wedge (VW).
The linear accelerator (LINAC) uses the ionization current generated in two dose monitoring ion chambers (MONITOR 1 and MONITOR 2) to dynamically control the three phases of VW delivery (initial gap, sweep portion, and open field).
We intentionally offset the calibration of MONITOR 2, which controls the time dose rate delivered on the LINAC, by up to ± 3% and observed fractional changes in the wedge dose profile of approximately 10% for a 6MV beam with a 60° wedge at a field size of 20 × 20 cm2.
We present an analytical expression to calculate the fractional change in wedge dosimetry as a function of time dose rate calibration of MONITOR 2.
A comparison with measurement showed that the expression was accurate to within 1%.
We will show how changes in the calibration of time dose rate (MONITOR 2) affect both the central axis wedge factor and wedge angle in order to assist in the diagnosis of wedge problems encountered in routine quality control measurements.
The effects are most prominent at low beam energies, large wedge angles and large field sizes.
Related Results
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
CLIMATE-2019 Program committee
CLIMATE-2019 Program committee
NOTITLE. Chairman
Mokhov Igor
RAS academecian, Dr. Sci., Professor
...
Influence of monitor chamber calibration on Virtual Wedge™ dosimetrya)
Influence of monitor chamber calibration on Virtual Wedge™ dosimetrya)
We have investigated the influence of the linear accelerator (LINAC) monitor chamber calibration on the dosimetry of Siemens Virtual Wedge™ (VW.) The doses delivered in the three p...
2 mils Au wire interchip wedge bond cratering study
2 mils Au wire interchip wedge bond cratering study
Au wire thermosonic wedge bonding is applied for die to die interconnect on accelerometer device. With the fragile bond pad structure of MEMS device, bond pad cratering or bond pad...
Macroeconomic and Social Precursors of Suicide Rates in the Philippines: A Quantitative Analysis (Preprint)
Macroeconomic and Social Precursors of Suicide Rates in the Philippines: A Quantitative Analysis (Preprint)
BACKGROUND
Suicide is a complex, serious and multifaceted public health issue that poses significant challenges to societies worldwide. In fact, it represen...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Developing the advanced dosimetry of linac beam radiotherapy
Developing the advanced dosimetry of linac beam radiotherapy
<p dir="ltr">Photon beams remains the most widely used modality for radiotherapy of cancer worldwide. Continued technological development has enabled increasingly conformal t...
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
Accelerating and improving radiochromic film calibration by utilizing the dose ratio in photon and proton beams
AbstractPurposeRadiochromic films are versatile 2D dosimeters with high‐resolution and near tissue equivalence. To assure high precision and accuracy, a time‐consuming calibration ...