Influence of monitor chamber calibration on Virtual Wedge™ dosimetrya)
View through CrossRef
We have investigated the influence of the linear accelerator (LINAC) monitor chamber calibration on the dosimetry of Siemens Virtual Wedge™ (VW.) The doses delivered in the three phases of wedge delivery (initial gap, sweep portion, and open field) utilize the ionization current generated in two dose monitoring ion chambers (MONITOR 1 and MONITOR 2) in the LINAC to control the wedge delivery. We intentionally offset the calibration of each of these chambers by and observed up to a 13% change in the dose along the wedge profile for a beam at a field size of . If the calibration of one of the two dose monitoring chambers changed independently then the relative dose at points along the wedge profile were affected. Furthermore, the percentage change in dose varied across the wedge profile thereby affecting the wedge angle as well as the central axis wedge factor. We also present equations for calculating the change in dose at a position along the wedge profile as a function of monitor chamber calibration. A comparison with measurement showed that our theoretical predictions were accurate to within . The equations have proven useful tools in evaluating periodic drifts in VW dosimetry.
Title: Influence of monitor chamber calibration on Virtual Wedge™ dosimetrya)
Description:
We have investigated the influence of the linear accelerator (LINAC) monitor chamber calibration on the dosimetry of Siemens Virtual Wedge™ (VW.
) The doses delivered in the three phases of wedge delivery (initial gap, sweep portion, and open field) utilize the ionization current generated in two dose monitoring ion chambers (MONITOR 1 and MONITOR 2) in the LINAC to control the wedge delivery.
We intentionally offset the calibration of each of these chambers by and observed up to a 13% change in the dose along the wedge profile for a beam at a field size of .
If the calibration of one of the two dose monitoring chambers changed independently then the relative dose at points along the wedge profile were affected.
Furthermore, the percentage change in dose varied across the wedge profile thereby affecting the wedge angle as well as the central axis wedge factor.
We also present equations for calculating the change in dose at a position along the wedge profile as a function of monitor chamber calibration.
A comparison with measurement showed that our theoretical predictions were accurate to within .
The equations have proven useful tools in evaluating periodic drifts in VW dosimetry.
Related Results
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
(Invited) Strategies for Calibration Cost Reduction in Heterogeneous Chemical Sensor Arrays
Introduction
Heterogeneous gas sensor arrays coupled with machine learning algorithms have been proposed for a wide range of applications. However, i...
2 mils Au wire interchip wedge bond cratering study
2 mils Au wire interchip wedge bond cratering study
Au wire thermosonic wedge bonding is applied for die to die interconnect on accelerometer device. With the fragile bond pad structure of MEMS device, bond pad cratering or bond pad...
Sci‐Sat AM (2) Therapy‐08: Influence of time dose rate calibration on Siemens Virtual Wedge dosimetry
Sci‐Sat AM (2) Therapy‐08: Influence of time dose rate calibration on Siemens Virtual Wedge dosimetry
We have investigated the influence of time dose rate calibration on the dosimetry of Siemens Virtual Wedge (VW). The linear accelerator (LINAC) uses the ionization current generate...
Foreland basin systems
Foreland basin systems
A foreland basin system is defined as: (a) an elongate region of potential sediment accommodation that forms on continental crust between a contractional orogenic belt and the adja...
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
Radiometric cross-calibration of Sentinel-2B MSI with HY-1C SCS based on the near simultaneous imaging of common ground targets
To simplify the cross-calibration process and improve calibration frequency and accuracy, this paper proposes a cross-calibration method for the multispectral remote sensor Multi-S...
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Automatic Hand-Eye Calibration Method of Welding Robot Based on Linear Structured Light
Aiming at solving the problems such as long calibration time, low precision, and complex operation in hand-eye calibration of welding robot, an automatic hand-eye calibration algor...
EFEKTIFITAS PELATIHAN LABORATORIUM VIRTUAL SEBAGAI MEDIA PEMBELAJARAN BAGI GURU KIMIA
EFEKTIFITAS PELATIHAN LABORATORIUM VIRTUAL SEBAGAI MEDIA PEMBELAJARAN BAGI GURU KIMIA
EFFECTIVITY OF VIRTUAL LABORATORY TRAINING AS A LEARNING MEDIA FOR CHEMISTRY TEACHERSAchmad Lutfi, SukarminUniversitas Negeri Surabaya, Indonesia achmadlutfi@unesa.ac.idAbstractThe...
Thrust Wedge Dominated Multilayered Propagation Using Finite Strain Sandbox Modeling: Growth of the Western Xuefeng Fold-Thrust Belt, South China Block
Thrust Wedge Dominated Multilayered Propagation Using Finite Strain Sandbox Modeling: Growth of the Western Xuefeng Fold-Thrust Belt, South China Block
Foreland fold–thrust belts act as key engines of continental growth and reworking by accommodating crustal shortening and internal deformation during long-term continental evolutio...