Telemetry changes over time in cochlear implant patients
View through CrossRef
Objectives:
This is a cross-sectional study conducted on cochlear implant recipients, to monitor changes in recorded impedance telemetry and evoked compound action potentials (ECAP) measured during surgery versus the same measures at post implant follow-up visits. An additional aim was to correlate the recorded ECAP measures with the patient's post-operative performance and to evaluate the incidence of abnormal intra-operative cochlear implant telemetry measurements and their changes at device activation.
Study design:
The present study comprised 44 subjects – 12 adults and 32 children of both genders – implanted at the Cochlear Implant Unit, Cairo University. ECAP thresholds and electrode impedance measures were collected intra-operatively, at initial stimulation, and at 9–12 months post initial stimulation. Measurements were compared at the three intervals and ECAP measures were correlated to the patient's post-operative performance in the form of aided sound field thresholds, speech detection and discrimination for adults and language assessment for children.
Results:
In the absence of electrical stimulation, there was an increase of impedance on all electrodes at the initial stimulation visit, which decreased at the 9–12 months follow-up. Open circuit findings were more prevalent than short circuits. Children had higher impedance values at post implant activation compared to adult patients. No correlation existed between ECAP thresholds and post-operative patient performance at the one-year evaluation.
Conclusion:
At the time of surgery, telemetry provides valuable information regarding the electrical output of the implant and the response of the auditory system to electrical stimulation; however, it is not a valuable predictor of post-operative performance.
Title: Telemetry changes over time in cochlear implant patients
Description:
Objectives:
This is a cross-sectional study conducted on cochlear implant recipients, to monitor changes in recorded impedance telemetry and evoked compound action potentials (ECAP) measured during surgery versus the same measures at post implant follow-up visits.
An additional aim was to correlate the recorded ECAP measures with the patient's post-operative performance and to evaluate the incidence of abnormal intra-operative cochlear implant telemetry measurements and their changes at device activation.
Study design:
The present study comprised 44 subjects – 12 adults and 32 children of both genders – implanted at the Cochlear Implant Unit, Cairo University.
ECAP thresholds and electrode impedance measures were collected intra-operatively, at initial stimulation, and at 9–12 months post initial stimulation.
Measurements were compared at the three intervals and ECAP measures were correlated to the patient's post-operative performance in the form of aided sound field thresholds, speech detection and discrimination for adults and language assessment for children.
Results:
In the absence of electrical stimulation, there was an increase of impedance on all electrodes at the initial stimulation visit, which decreased at the 9–12 months follow-up.
Open circuit findings were more prevalent than short circuits.
Children had higher impedance values at post implant activation compared to adult patients.
No correlation existed between ECAP thresholds and post-operative patient performance at the one-year evaluation.
Conclusion:
At the time of surgery, telemetry provides valuable information regarding the electrical output of the implant and the response of the auditory system to electrical stimulation; however, it is not a valuable predictor of post-operative performance.
Related Results
Abstract 173: Simple Guidance Improves Appropriate Telemetry Utilization
Abstract 173: Simple Guidance Improves Appropriate Telemetry Utilization
Introduction:
Cardiac telemetry is an in-hospital monitoring tool intended for ischemia surveillance, monitor QT-interval prolongation, and detect arrhythmias. It is a ...
The Effect of Cochlear Coverage on Auditory and Speech Performance in Cochlear Implant Patients
The Effect of Cochlear Coverage on Auditory and Speech Performance in Cochlear Implant Patients
Objective:
To determine the effect of cochlear coverage on audiological and speech parameters in patients with cochlear implants. Previous work has investigated the eff...
Abstract 11292: Reducing Telemetry Utilization in Hospitalized Comfort Care Patients
Abstract 11292: Reducing Telemetry Utilization in Hospitalized Comfort Care Patients
Introduction:
Telemetry overutilization increases costs and interventions that may result in patient harm, particularly at the end of life. ‘Comfort care’ (CC) patients...
The Therapeutic Dilemma of Cochlear Nerve Deficiency: Cochlear or Brainstem Implantation?
The Therapeutic Dilemma of Cochlear Nerve Deficiency: Cochlear or Brainstem Implantation?
ObjectiveTo compare the outcomes between 2 age‐matched cohorts of children with cochlear nerve deficiency: those receiving auditory brainstem implants (group A) or cochlear implant...
Guideline‐based intervention to reduce telemetry rates in a large tertiary centre
Guideline‐based intervention to reduce telemetry rates in a large tertiary centre
AbstractBackgroundInappropriate cardiac telemetry use is associated with reduced patient flow and increased healthcare costs.AimTo evaluate the outcomes of guideline‐based applicat...
Investigating the Electrical Properties of Different Cochlear Implants
Investigating the Electrical Properties of Different Cochlear Implants
Aim:
This study characterises and compares electrical properties and current spread across four different makes of cochlear implants with differing electrode designs us...
A novel scoring system for vestibular schwannomas to identify candidacy for cochlear implantation
A novel scoring system for vestibular schwannomas to identify candidacy for cochlear implantation
Objectives Here we present the audiometric outcomes of patients
undergoing vestibular schwannoma resection and cochlear implantation. We
additionally reviewed preoperative audiomet...
Large acute posttraumatic epidural hematoma centered beneath a cochlear implant: A case report
Large acute posttraumatic epidural hematoma centered beneath a cochlear implant: A case report
Background:
Cochlear implants provide substantial functional benefit to patients with severe sensorineural hearing loss. Management of acute intracranial hemorr...