Implant‐Abutment Interface: Biomechanical Study of Flat Top versus Conical

View through CrossRef
ABSTRACT Background: Overloading has been identified as a primary factor behind dental implant failure. The peak bone stresses normally appear in the marginal bone. The anchorage strength is maximized if the implant is given a design that minimizes the peak bone stress caused by a standardized load. Clinical studies have shown that it is possible to obtain a marginal bone level close to the crest of the implant. Different implant systems make use of different designs of the implant‐abutment interface. Different implant‐abutment interfaces imply that the functional load is distributed in different ways upon the implant. According to Saint‐Venant's principle, this will result in different stress patterns in the marginal bone when this reaches levels close to the implant crest. Purpose: One aim of the study was to theoretically investigate if a conical implant‐abutment interface gives rise to a changed stress pattern in the marginal bone, as compared to a flat top interface, for an axially loaded mandibular titanium implant, the neck of which is provided with retention elements giving effective interlocking with the bone. Further aims were to investigate if the way in which the axial load is distributed on the flat top and on the inner conus respectively affects the stress pattern in the marginal bone. The pertinent stress was considered to be the bone‐implant interfacial shear stress. It was assumed that the marginal bone reached the level of the implant–abutment interface. Method: The investigation was performed by means of axisymmetric finite element analysis. Results: The conical implant‐abutment interface of the type studied brought about a decrease in the peak bone‐implant interfacial shear stress as compared to the flat top interface of the type studied. This peak interfacial shear stress was located at the top marginal bone for the flat top implant‐abutment interface whereas it was located more apically in the bone for the conical implant‐abutment interface. The way in which the axial load was distributed on the flat top and on the inner conus respectively affected the peak interfacial shear stress level. Conclusion: The design of the implant‐abutment interface has a profound effect upon the stress state in the marginal bone when this reaches the level of this interface. The implant with the conical interface can theoretically resist a larger axial load than the implant with the flat top interface.
Title: Implant‐Abutment Interface: Biomechanical Study of Flat Top versus Conical
Description:
ABSTRACT Background: Overloading has been identified as a primary factor behind dental implant failure.
The peak bone stresses normally appear in the marginal bone.
The anchorage strength is maximized if the implant is given a design that minimizes the peak bone stress caused by a standardized load.
Clinical studies have shown that it is possible to obtain a marginal bone level close to the crest of the implant.
Different implant systems make use of different designs of the implant‐abutment interface.
Different implant‐abutment interfaces imply that the functional load is distributed in different ways upon the implant.
According to Saint‐Venant's principle, this will result in different stress patterns in the marginal bone when this reaches levels close to the implant crest.
Purpose: One aim of the study was to theoretically investigate if a conical implant‐abutment interface gives rise to a changed stress pattern in the marginal bone, as compared to a flat top interface, for an axially loaded mandibular titanium implant, the neck of which is provided with retention elements giving effective interlocking with the bone.
Further aims were to investigate if the way in which the axial load is distributed on the flat top and on the inner conus respectively affects the stress pattern in the marginal bone.
The pertinent stress was considered to be the bone‐implant interfacial shear stress.
It was assumed that the marginal bone reached the level of the implant–abutment interface.
Method: The investigation was performed by means of axisymmetric finite element analysis.
Results: The conical implant‐abutment interface of the type studied brought about a decrease in the peak bone‐implant interfacial shear stress as compared to the flat top interface of the type studied.
This peak interfacial shear stress was located at the top marginal bone for the flat top implant‐abutment interface whereas it was located more apically in the bone for the conical implant‐abutment interface.
The way in which the axial load was distributed on the flat top and on the inner conus respectively affected the peak interfacial shear stress level.
Conclusion: The design of the implant‐abutment interface has a profound effect upon the stress state in the marginal bone when this reaches the level of this interface.
The implant with the conical interface can theoretically resist a larger axial load than the implant with the flat top interface.

Related Results

REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
REUSE OF THE INTRAOSSEOUS PART OF THE DENTAL IMPLANT (THE CLINICAL CASE)
Introduction. Replacement of a small included defect in the frontal area with a fixed structure based on the implant is the optimal solution. It is not always required to remove th...
Factors Influencing Marginal Bone Loss around Dental Implants: A Narrative Review
Factors Influencing Marginal Bone Loss around Dental Implants: A Narrative Review
Implant supported dental prostheses are increasingly used in dental practice. The aim of this narrative review is to present the influence of transmucosal surface of prosthetic abu...
Soft tissue features of peri‐implant diseases and related treatment
Soft tissue features of peri‐implant diseases and related treatment
AbstractBackgroundThe need for soft tissue grafting at implant sites for preventing and treating peri‐implant diseases is a currently investigated and debated topic.PurposeThe aim ...
The Instability Analysis and Processing Scheme of ChangXin Viaduct Abutment
The Instability Analysis and Processing Scheme of ChangXin Viaduct Abutment
In order to solve the abutment instability problem of the ChangXin Viaduct in JiLin to YanJi highway, we had the detailed geology survey and exploration of the abutment and its bac...
Biomechanical characterisation of the bone-implant interface
Biomechanical characterisation of the bone-implant interface
Caractérisation biomécanique de l'interface os-implant. Les interventions chirurgicales impliquant des implants orthopédiques ou dentaires échouent encore souvent p...

Back to Top