Optimisasi Dosis dan Kualitas Citra CT-Scan untuk Variasi Pitch sebagai Upaya Proteksi Radiasi di Rumah Sakit Umum Pusat Adam Malik Medan
View through CrossRef
Optimization is an effort to ensure patient radiation safety and is the main action in overcoming concerns about CT-Scan radiation exposure. This led to the emergence of various measures to reduce the dose. This study aims to obtain a minimal dose with a high-quality image. Optimization efforts were carried out by the radiology team at Adam Malik Hospital Medan using a 16 slice GE CT-Scan and a water phantom with a diameter of 16 and 32 cm and an image quality questionnaire form. Collected data by observing the head, chest, and abdomen CT-Scan in adult patients (≥15 years). The data taken is the value of CTDI vol and DLP for a year. Then a water phantom scan was carried out with the head protocol using pitch parameters 0.562 and 0.938. The chest and abdomen use pitches of 1.375 and 1.75. The results obtained were evaluated and applied to patients, then filled in the image quality questionnaire scores. The results of CTDI_vol and DLP values with 16 and 32 cm water phantom scans showed a decrease in the dose value; for pitch 0.938, it was 1.6% lower than pitch 0.562, and pitch 1.75 was 1.2% lower, compared to pitch 1.375. For CT head examination using a pitch of 0.963, the CTDI_vol value was 1.5%, and DLP was 2%. For chest using a pitch of 1.75, CTDI_vol values were 1.3% and DLP 2%, while abdominal examination with a pitch of 1.75 obtained CTDI_vol values 1.8% and DLP 1.4%. From these three results, the CTDI_vol and DLP values were higher than the national DRL values. The value obtained is higher than the national DRL due to differences in the phantom test protocol with clinical implementation and the lack of accuracy in using other parameters. Changes in scan parameters are not comprehensive. Obtained a score of 3 in the questionnaire form stating that the radiology doctor can still interpret the image. This study concluded that it could make optimization efforts by changing the pitch parameter by paying attention to other parameters without reducing the quality of the image interpreted by the radiologist.
Badan Pengawas Tenaga Nuklir (BAPETEN)
Title: Optimisasi Dosis dan Kualitas Citra CT-Scan untuk Variasi Pitch sebagai Upaya Proteksi Radiasi di Rumah Sakit Umum Pusat Adam Malik Medan
Description:
Optimization is an effort to ensure patient radiation safety and is the main action in overcoming concerns about CT-Scan radiation exposure.
This led to the emergence of various measures to reduce the dose.
This study aims to obtain a minimal dose with a high-quality image.
Optimization efforts were carried out by the radiology team at Adam Malik Hospital Medan using a 16 slice GE CT-Scan and a water phantom with a diameter of 16 and 32 cm and an image quality questionnaire form.
Collected data by observing the head, chest, and abdomen CT-Scan in adult patients (≥15 years).
The data taken is the value of CTDI vol and DLP for a year.
Then a water phantom scan was carried out with the head protocol using pitch parameters 0.
562 and 0.
938.
The chest and abdomen use pitches of 1.
375 and 1.
75.
The results obtained were evaluated and applied to patients, then filled in the image quality questionnaire scores.
The results of CTDI_vol and DLP values with 16 and 32 cm water phantom scans showed a decrease in the dose value; for pitch 0.
938, it was 1.
6% lower than pitch 0.
562, and pitch 1.
75 was 1.
2% lower, compared to pitch 1.
375.
For CT head examination using a pitch of 0.
963, the CTDI_vol value was 1.
5%, and DLP was 2%.
For chest using a pitch of 1.
75, CTDI_vol values were 1.
3% and DLP 2%, while abdominal examination with a pitch of 1.
75 obtained CTDI_vol values 1.
8% and DLP 1.
4%.
From these three results, the CTDI_vol and DLP values were higher than the national DRL values.
The value obtained is higher than the national DRL due to differences in the phantom test protocol with clinical implementation and the lack of accuracy in using other parameters.
Changes in scan parameters are not comprehensive.
Obtained a score of 3 in the questionnaire form stating that the radiology doctor can still interpret the image.
This study concluded that it could make optimization efforts by changing the pitch parameter by paying attention to other parameters without reducing the quality of the image interpreted by the radiologist.
Related Results
Analisis Proteksi Radiasi bagi Pekerja Radiasi di Instalasi Radiologi Rumah Sakit Universitas Hasanuddin Makassar
Analisis Proteksi Radiasi bagi Pekerja Radiasi di Instalasi Radiologi Rumah Sakit Universitas Hasanuddin Makassar
Proteksi radiasi adalah cabang ilmu pengetahuan yang berkaitan dengan teknik kesehatan lingkungan yaitu tentang proteksi yang perlu di berikan kepada seseorang atau sekelompok oran...
STUDI ANALISIS AUDIT PAPARAN MEDIK SEBAGAI UPAYA OPTIMISASI PADA CT HEAD DI RUMAH SAKIT UMUM ELISABET MEDAN
STUDI ANALISIS AUDIT PAPARAN MEDIK SEBAGAI UPAYA OPTIMISASI PADA CT HEAD DI RUMAH SAKIT UMUM ELISABET MEDAN
Studi analisis audit paparan medik sebagai upaya optimisasi pada CT HEAD di Rumah Sakit Umum Elisabet Medan. Audit dosis dilakukan pada pemeriksaan CT scan melalui metrik dosis, un...
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi Dosis Radiasi Teknik Penyinaran 3D-CRT pada Pasien Kanker Payudara Menggunakan Film EBT3 di Rumah Sakit UNAND
Verifikasi dosis radiasi teknik penyinaran Three-Dimensional Conformal Radiation Therapy  (3D-CRT) pada pasien kanker payudara telah dilakukan menggunakan film EBT3. Verifikasi do...
OPTIMISASI PARAMETER SCAN PADA CT SCAN KEPALA NONKONTRAS DI INSTALASI RADIOLOGI RSUD PROF DR MARGONO SOEKARJO
OPTIMISASI PARAMETER SCAN PADA CT SCAN KEPALA NONKONTRAS DI INSTALASI RADIOLOGI RSUD PROF DR MARGONO SOEKARJO
Faktor eksposi yang sangat mempengaruhi dosis radiasi dan kualitas citra pada pemeriksaan CT Scan diantaranya tegangan (kV), kuat arus (mA) dan waktu pemindaian citra (s). Tegangan...
Pengantar Sistem Proteksi Radiasi
Pengantar Sistem Proteksi Radiasi
Selain membawa manfaat yang sangat besar, penggunaan tenaga nuklir diketahui memiliki efek yang berbahaya bagi kesehatan manusia. Efek radiasi yang dihasilkan dari tenaga nuklir da...
Evaluasi Penerapan Proteksi Radiasi pada Pekerja Radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri, dan RS UNAND
Evaluasi Penerapan Proteksi Radiasi pada Pekerja Radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri, dan RS UNAND
Telah dilakukan penelitian tentang penerapan proteksi radiasi pada pekerja radiasi di Instalasi Radiologi RS Naili DBS, RS Selaguri dan RS UNAND. Penelitian bertujuan untuk menge...
Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Penerapan Optimisasi Paparan Medik CT-Scan Melalui Tingkat Panduan Diagnostik di RSUP dr. Soeradji Tirtonegoro
Latar belakang: Pemanfaatan CT-Scan di RSUP dr. Soeradji Tirtonegoro menunjukkan adanya peningkatan yang signifikan, pemanfaatan tersebut harus dilakukan pengawasan untuk menjamin ...
Gambaran Peralatan Proteksi Radiasi di Intalasi Radiologi Rumah Sakit Khusus Bedah Hasta Husada
Gambaran Peralatan Proteksi Radiasi di Intalasi Radiologi Rumah Sakit Khusus Bedah Hasta Husada
Radiasi pengion adalah jenis energi yang dilepaskan oleh atom yang merambat dalam bentuk gelombang elektromagnetik atau partikel. Keselamatan terhadap radiasi pengion adalah salah ...