Evaluation and assessment of the near field behind 3D printed-modulators for proton beam therapy

View through CrossRef
A 3D-printed modulator is an innovation developed for particle therapy delivery systems that enables a highly conformal and homogeneous dose distribution around the tumor within a very short irradiation time. In normal cases, the modulators are positioned far from the patient in order to avoid the field inhomogeneity resulting from the periodic structure of the modulators on patient’s skin. However, a smaller distance between the modulator and patient would provide a better dose conformation in the target volume. In this thesis, the FLUKA Monte Carlo simulation program was used to investigate the fluence distributions of protons penetrating through 3D-printed modulators and to determine the minimum distance at which the dose is homogeneous on patient’s skin. To implement the complex geometry of the 3Dprinted modulator in FLUKA, a dedicated FLUKA user routine with a shorter run time was developed. The sensitivity of the fluence ripple was also tested and found to be strongly dependent on the initial beam energy and the pin period of the modulator. The results of radiochromic film and dose measurements show a qualitatively good agreement with the FLUKA simulations. Furthermore, this thesis introduces the idea of the short distance of the modulator setup which could exploit strong dose inhomogeneities induced by the 3D-printed modulator for normal-tissue sparing. The minimum energy, which can utilize the advantages of the small distance setup without the interference of the dose inhomogeneity in the tumor, is 150 MeV for a tumor with a maximum width of 5 cm.
Office of Academic Resources, Chulalongkorn University
Title: Evaluation and assessment of the near field behind 3D printed-modulators for proton beam therapy
Description:
A 3D-printed modulator is an innovation developed for particle therapy delivery systems that enables a highly conformal and homogeneous dose distribution around the tumor within a very short irradiation time.
In normal cases, the modulators are positioned far from the patient in order to avoid the field inhomogeneity resulting from the periodic structure of the modulators on patient’s skin.
However, a smaller distance between the modulator and patient would provide a better dose conformation in the target volume.
In this thesis, the FLUKA Monte Carlo simulation program was used to investigate the fluence distributions of protons penetrating through 3D-printed modulators and to determine the minimum distance at which the dose is homogeneous on patient’s skin.
To implement the complex geometry of the 3Dprinted modulator in FLUKA, a dedicated FLUKA user routine with a shorter run time was developed.
The sensitivity of the fluence ripple was also tested and found to be strongly dependent on the initial beam energy and the pin period of the modulator.
The results of radiochromic film and dose measurements show a qualitatively good agreement with the FLUKA simulations.
Furthermore, this thesis introduces the idea of the short distance of the modulator setup which could exploit strong dose inhomogeneities induced by the 3D-printed modulator for normal-tissue sparing.
The minimum energy, which can utilize the advantages of the small distance setup without the interference of the dose inhomogeneity in the tumor, is 150 MeV for a tumor with a maximum width of 5 cm.

Related Results

SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
Purpose:To build the model of a spot scanning proton beam for the dose calculation of a synchrotron proton therapy accelerator, which is capable of accelerating protons from 50 up ...
SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities. It support...
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses...
E-Survey: Current Status of Proton Beam Therapy in USA
E-Survey: Current Status of Proton Beam Therapy in USA
The clinical use of Proton Beam Therapy (PBT) in the management of cancer is increasing worldwide. USA has high number of proton therapy centres and this number is growing with new...
Intense-Proton-Beam Transport through an Insulator Beam Guide
Intense-Proton-Beam Transport through an Insulator Beam Guide
In this paper we study intense-proton-beam transport through an insulator guide. In our previous papers [Jpn. J. Appl. Phys. 34 (1995) L520, Jpn. J. Appl. Phys. 35 (1996) L112...
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
Purpose: Intensity‐modulated proton therapy (IMPT) will improve the conformality of proton radiotherapy while preserving target homogeneity and low integral dose characteristics. I...
Radiobiology of proton therapy and its clinical implications
Radiobiology of proton therapy and its clinical implications
The chapter delves into the intricate relationship between proton therapy and its impact on biological systems, shaping the landscape of modern cancer treatment. Proton accelerator...
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
Focusing effect of channel target on ultra-intense laser-accelerated proton beam
In laser proton acceleration, the inevitable transverse divergence of proton beam restricts its applications in many fields. In this paper, a structured target with a properly wide...