SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System

View through CrossRef
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities. It supports three‐dimensional beam scanning with energy and intensity modulation. A series of measurements have been conducted to characterize the beam performance of the first installation of the system at the McLaren Proton Therapy Center in Flint, Michigan. These measurements were part of the technical commissioning of the system. Select measurements and results are presented.Methods:The Radiance 330 proton beam energy range is 70–250 MeV for treatment, and up to 330 MeV for proton tomography and radiography. Its 3‐D scanning capability, together with a small beam emittance and momentum spread, provides a highly efficient beam delivery. During the technical commissioning, treatment plans were created to deliver uniform maps at various energies to perform Gamma Index analysis. EBT3 Gafchromic films were irradiated using the Planned irradiation maps. Bragg Peak chamber was used to test the dynamic range during a scan in one layer for high (250 MeV) and Low (70 MeV) energies. The maximum and minimum range, range adjustment and modulation, distal dose falloff (80%–20%), pencil beam spot size, spot placement accuracy were also measured. The accuracy testing included acquiring images, image registration, receiving correction vectors and applying the corrections to the robotic patient positioner.Results:Gamma Index analysis of the Treatment Planning System (TPS) data vs. Measured data showed more than 90% of points within (3%, 3mm) for the maps created by the TPS. At Isocenter Beam Size (One sigma) < 3mm at highest energy (250 MeV) in air. Beam delivery was within 0.6 mm of the intended target at the entrance and the exit of the beam, through the phantom.Conclusion:The Radiance 330 Beam Performance Measurements have confirmed that the system operates as designed with excellent clinical performance specifications.Hovakim Nazaryan, Vahagn Nazaryan and Fuhua Wang are employees of ProTom International, Inc. who contributed to the development and completed the technical commissioning of the Radiance 330 proton therapy delivery system manufactured by ProTom International.
Title: SU‐E‐T‐470: Beam Performance of the Radiance 330 Proton Therapy System
Description:
Purpose:The ProTom Radiance 330 proton radiotherapy system is a fully functional, compact proton radiotherapy system that provides advanced proton delivery capabilities.
It supports three‐dimensional beam scanning with energy and intensity modulation.
A series of measurements have been conducted to characterize the beam performance of the first installation of the system at the McLaren Proton Therapy Center in Flint, Michigan.
These measurements were part of the technical commissioning of the system.
Select measurements and results are presented.
Methods:The Radiance 330 proton beam energy range is 70–250 MeV for treatment, and up to 330 MeV for proton tomography and radiography.
Its 3‐D scanning capability, together with a small beam emittance and momentum spread, provides a highly efficient beam delivery.
During the technical commissioning, treatment plans were created to deliver uniform maps at various energies to perform Gamma Index analysis.
EBT3 Gafchromic films were irradiated using the Planned irradiation maps.
Bragg Peak chamber was used to test the dynamic range during a scan in one layer for high (250 MeV) and Low (70 MeV) energies.
The maximum and minimum range, range adjustment and modulation, distal dose falloff (80%–20%), pencil beam spot size, spot placement accuracy were also measured.
The accuracy testing included acquiring images, image registration, receiving correction vectors and applying the corrections to the robotic patient positioner.
Results:Gamma Index analysis of the Treatment Planning System (TPS) data vs.
Measured data showed more than 90% of points within (3%, 3mm) for the maps created by the TPS.
At Isocenter Beam Size (One sigma) < 3mm at highest energy (250 MeV) in air.
Beam delivery was within 0.
6 mm of the intended target at the entrance and the exit of the beam, through the phantom.
Conclusion:The Radiance 330 Beam Performance Measurements have confirmed that the system operates as designed with excellent clinical performance specifications.
Hovakim Nazaryan, Vahagn Nazaryan and Fuhua Wang are employees of ProTom International, Inc.
who contributed to the development and completed the technical commissioning of the Radiance 330 proton therapy delivery system manufactured by ProTom International.

Related Results

Study on Dosimetry Technology of Absorbed Dose to Water for Proton
Study on Dosimetry Technology of Absorbed Dose to Water for Proton
In proton therapy, precise energy transfer to tissues is crucial for cancer treatment, as accurate dose delivery directly influences tumor control efficiency and minimizes radiatio...
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
SU‐E‐T‐47: A Monte Carlo Model of a Spot Scanning Proton Beam Based On a Synchrotron Proton Therapy Accelerator
Purpose:To build the model of a spot scanning proton beam for the dose calculation of a synchrotron proton therapy accelerator, which is capable of accelerating protons from 50 up ...
Transmittance and Radiance Computations for Rocket Engine Plume Environments
Transmittance and Radiance Computations for Rocket Engine Plume Environments
Rocket engine exhaust plume is generally thermal in character arising from changes in the internal energy of constituent molecules. Radiation from the plume is attenuated in its pa...
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses...
Computer simulation of an excess proton in aqueous systems
Computer simulation of an excess proton in aqueous systems
This thesis aims at studying the microscopic physical-chemical properties of an excess proton in aqueous systems. From bulk water environments to narrow hydrophobic channels constr...
E-Survey: Current Status of Proton Beam Therapy in USA
E-Survey: Current Status of Proton Beam Therapy in USA
The clinical use of Proton Beam Therapy (PBT) in the management of cancer is increasing worldwide. USA has high number of proton therapy centres and this number is growing with new...
Intense-Proton-Beam Transport through an Insulator Beam Guide
Intense-Proton-Beam Transport through an Insulator Beam Guide
In this paper we study intense-proton-beam transport through an insulator guide. In our previous papers [Jpn. J. Appl. Phys. 34 (1995) L520, Jpn. J. Appl. Phys. 35 (1996) L112...
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
TH‐E‐ValA‐02: Simultaneous Multi‐Pencil Fan‐Beam‐Based Intensity‐Modulated Proton Therapy.
Purpose: Intensity‐modulated proton therapy (IMPT) will improve the conformality of proton radiotherapy while preserving target homogeneity and low integral dose characteristics. I...

Back to Top