Pengaruh Variasi Waktu dan Kecepatan Centrifuge terhadap Nilai Hematokrit Metode Makrohematokrit
View through CrossRef
Hematocrit is the volume percentage of red blood cells, which is an important parameter in the diagnosis of various health conditions. In this study, there were 9 samples with 3 repetitions. Each sample is then processed using a centrifuge with varying speeds and durations. The centrifuge speeds used included 3000, 3500, and 4000 RPM, while processing times were set at 10, 15, and 20 minutes. The results showed that there was a significant relationship between the speed and duration of the centrifuge and the hematocrit value obtained. In general, increasing centrifuge speed and processing duration tends to result in higher hematocrit values. However, at a certain rate and duration, further increases did not provide significant changes in hematocrit values, indicating the existence of a saturation point where optimal conditions have been reached. A centrifuge speed of 3000 RPM for 10 minutes was found to be the most efficient combination to obtain accurate hematocrit values without damaging red blood cells. This study provides important insights for clinical laboratories in determining optimal parameters for accurate and efficient hematocrit measurements. Proper setting of centrifuge speed and duration not only increases the accuracy of results but also reduces operational time and costs. Further studies are recommended to explore the effects of other factors such as temperature and type of centrifuge used on hematocrit values.
Title: Pengaruh Variasi Waktu dan Kecepatan Centrifuge terhadap Nilai Hematokrit Metode Makrohematokrit
Description:
Hematocrit is the volume percentage of red blood cells, which is an important parameter in the diagnosis of various health conditions.
In this study, there were 9 samples with 3 repetitions.
Each sample is then processed using a centrifuge with varying speeds and durations.
The centrifuge speeds used included 3000, 3500, and 4000 RPM, while processing times were set at 10, 15, and 20 minutes.
The results showed that there was a significant relationship between the speed and duration of the centrifuge and the hematocrit value obtained.
In general, increasing centrifuge speed and processing duration tends to result in higher hematocrit values.
However, at a certain rate and duration, further increases did not provide significant changes in hematocrit values, indicating the existence of a saturation point where optimal conditions have been reached.
A centrifuge speed of 3000 RPM for 10 minutes was found to be the most efficient combination to obtain accurate hematocrit values without damaging red blood cells.
This study provides important insights for clinical laboratories in determining optimal parameters for accurate and efficient hematocrit measurements.
Proper setting of centrifuge speed and duration not only increases the accuracy of results but also reduces operational time and costs.
Further studies are recommended to explore the effects of other factors such as temperature and type of centrifuge used on hematocrit values.
Related Results
MENENTUKAN WAKTU STANDAR PADA AKTIVITAS KERJA PRODUKSI SABLON MANUAL DI CV. DWIPUTRA IHWA
MENENTUKAN WAKTU STANDAR PADA AKTIVITAS KERJA PRODUKSI SABLON MANUAL DI CV. DWIPUTRA IHWA
Analisis waktu standar produksi merupakan salah satu analisis metoda kuantitatif yang dilakukan untuk mengukur waktu produksi dan bertujuan agar dapat memiliki waktu standar sebaga...
PERAN TATA KELOLA PERUSAHAAN DALAM MEMODERASI PENGARUH IMPLEMANTASI GREEN ACCOUNTING, CORPORATE SOCIAL RESPONSIBILITY DAN FIRM SIZE TERHADAP KINERJA KEUANGAN
PERAN TATA KELOLA PERUSAHAAN DALAM MEMODERASI PENGARUH IMPLEMANTASI GREEN ACCOUNTING, CORPORATE SOCIAL RESPONSIBILITY DAN FIRM SIZE TERHADAP KINERJA KEUANGAN
This study examines the role of corporate governance in moderating the influence of green accounting disclosure, corporate social responsibility (CSR), and firm size on the financi...
Lectin C gene analysis v1
Lectin C gene analysis v1
Mammalian Tissue Total RNA Purification Protocol by GeneJET RNA Purification Kit (Thermo Scientific, USA) Before starting: • Supplement the required amount of Lysis Buffer with β-...
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
Environmental Surveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) v2
EnvironmentalSurveillance Protocols for Highly Pathogenic Avian Influenza (HPAI) This comprehensive protocol suite enables systematic environmental surveillance for avian influenza...
Perbedaan Hasil Pemeriksaan Mikro Hematokrit Menggunakan Makrosentrifus Dengan Mikrosentrifus
Perbedaan Hasil Pemeriksaan Mikro Hematokrit Menggunakan Makrosentrifus Dengan Mikrosentrifus
Abstract: Hematocrit measurement is one of the most commonly special blood examination in laboratory to help in diagnosing various kind of diseases. Determination of hematocrit val...
HUBUNGAN KADAR HEMOGLOBIN DAN NILAI HEMATOKRIT PADA PEROKOK AKTIF
HUBUNGAN KADAR HEMOGLOBIN DAN NILAI HEMATOKRIT PADA PEROKOK AKTIF
Perokok aktif merupakan orang yang merokok secara aktif atau terus menerus. Rokok mengandung bahan beracun dan berefek candu. Adanya tar dan radikal bebas dari asap rokok dapat men...
Kecepatan rotasi dan angular
Kecepatan rotasi dan angular
Kecepatan rotasi adalah tentang seberapa lambat atau seberapa cepat suatu benda bergerak. Kita telah mendengar tentang kecepatan tetapi tahukah kita apa yang sedang kita bicarakan ...
Perbandingan Kadar Hemoglobin dan Hematokrit pada Early Onset dan Late Onset Preeclampsia di RSUD Dr. Moewardi Surakarta
Perbandingan Kadar Hemoglobin dan Hematokrit pada Early Onset dan Late Onset Preeclampsia di RSUD Dr. Moewardi Surakarta
Pendahuluan: Preeklampsia merupakan salah satu penyebab tingginya angka kematian ibu, terutama di negara berkembang. Pemeriksaan hematokrit dan hemoglobin adalah salah satu pemerik...