The fixation tines of the Micra™ leadless pacemaker are atraumatic to the tricuspid valve
View through CrossRef
AbstractBackgroundToday, there is no manufacturer‐supplied retrieval tool for the Micra™ pacemaker (Medtronic, Minneapolis, MN, USA); therefore, off‐the‐shelf catheters have been employed for retrievals. The proximal retrieval feature of the Micra™ can be snared and the device is then retracted from the myocardium, pulling the device through the tricuspid valve. This study characterizes the potential risks of Micra™ nitinol tine engagement with the tricuspid sub‐valvular apparatus.MethodsFresh human hearts nonviable for transplant (n = 10) were obtained from our regional organ procurement agency (LifeSource, Minneapolis, MN, USA). Micra™ fixation tines were affixed to a linear force transducer. Tines were then engaged in tricuspid chordae tendineae to conduct a constant velocity tensile test. Each test was run until tines disengaged from the chordae tendineae or until they released from the valve apparatus. Subsequently, biomechanical failure properties of the valve apparatus and isolated chordae tendineae were determined using a series of uniaxial tensile tests.ResultsThere were no chordal ruptures observed during our Micra™ tine extraction testing. Chordal failure required 15.0 times the force of extracting a single engaged tine, and 9.0 times the force of extracting two engaged tines. The uniaxial stresses required for isolated chordal failure averaged 17.4 N/mm2; failure strains exceeded 150% resting chordal length.ConclusionsThe forces required to rupture tricuspid chordae tendineae significantly exceeded the forces potentially imposed on the chordae during Micra™ device retrievals. We conclude that the fixation tines of the Micra™ device are unlikely to damage the tricuspid apparatus during either implant or retrieval.
Title: The fixation tines of the Micra™ leadless pacemaker are atraumatic to the tricuspid valve
Description:
AbstractBackgroundToday, there is no manufacturer‐supplied retrieval tool for the Micra™ pacemaker (Medtronic, Minneapolis, MN, USA); therefore, off‐the‐shelf catheters have been employed for retrievals.
The proximal retrieval feature of the Micra™ can be snared and the device is then retracted from the myocardium, pulling the device through the tricuspid valve.
This study characterizes the potential risks of Micra™ nitinol tine engagement with the tricuspid sub‐valvular apparatus.
MethodsFresh human hearts nonviable for transplant (n = 10) were obtained from our regional organ procurement agency (LifeSource, Minneapolis, MN, USA).
Micra™ fixation tines were affixed to a linear force transducer.
Tines were then engaged in tricuspid chordae tendineae to conduct a constant velocity tensile test.
Each test was run until tines disengaged from the chordae tendineae or until they released from the valve apparatus.
Subsequently, biomechanical failure properties of the valve apparatus and isolated chordae tendineae were determined using a series of uniaxial tensile tests.
ResultsThere were no chordal ruptures observed during our Micra™ tine extraction testing.
Chordal failure required 15.
0 times the force of extracting a single engaged tine, and 9.
0 times the force of extracting two engaged tines.
The uniaxial stresses required for isolated chordal failure averaged 17.
4 N/mm2; failure strains exceeded 150% resting chordal length.
ConclusionsThe forces required to rupture tricuspid chordae tendineae significantly exceeded the forces potentially imposed on the chordae during Micra™ device retrievals.
We conclude that the fixation tines of the Micra™ device are unlikely to damage the tricuspid apparatus during either implant or retrieval.
Related Results
The angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation
The angle of the tines before the pull and hold test predicts engagement of the tines in Micra leadless pacemaker implantation
AbstractBackgroundMicra leadless pacemaker is secured to the myocardium by engagement of at least 2/4 tines confirmed with pull and hold test. However, the pull and hold test is so...
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Aveir Leadless Pacemaker implantation in pediatric population: a case series
Abstract
Background
While the Medtronic Micra pacemaker provided a small device for leadless pacemaker implantation, the Aveir d...
Leadless Pacemaker tine fracture and dislocation: two case reports
Leadless Pacemaker tine fracture and dislocation: two case reports
Abstract
The rate of leadless pacemaker implantation is increasing worldwide. To date, there have been few reports of leadless pacemaker dislocation and extraction. This ar...
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
P185 FOLLOW–UP OF PATIENTS UNDERGOING POST–EXTRACTION LEADLESS PACEMAKER IMPLANTATION
Abstract
Introduction
The implantation of traditional stimulation devices involves numerous complications, in the short and long...
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Comparison of Heart Valve Circumference Examined Before and After 10% Formalin Fixation
Objective: To compare the heart valve circumference before and after 10% formalin fixation. Materials and Methods: The study analyzed 63 Thai human cadaveric hearts. Each heart val...
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
P425Single procedure pace and ablate. evaluation of efficacy and safety comparing three different vascular routes
Abstract
Funding Acknowledgements
No financial support for this study
Introduc...
Bacteraemia after leadless pacemaker implantation
Bacteraemia after leadless pacemaker implantation
Background: Transvenous permanent pacemaker related infection is a
severe condition associated with significant morbidity and mortality.
Leadless pacemakers may be more resistant t...
Redo isolated tricuspid valve replacement in a patient with isolated persistent left superior vena cava: a case report
Redo isolated tricuspid valve replacement in a patient with isolated persistent left superior vena cava: a case report
Abstract
Background
Redo isolated tricuspid valve surgery has high in-hospital mortality and morbidity and is a challenging procedure. We report a s...