Real-time CBCT imaging and motion tracking via a single arbitrarily-angled x-ray projection by a joint dynamic reconstruction and motion estimation (DREME) framework
View through CrossRef
Abstract
Objective. Real-time cone-beam computed tomography (CBCT) provides instantaneous visualization of patient anatomy for image guidance, motion tracking, and online treatment adaptation in radiotherapy. While many real-time imaging and motion tracking methods leveraged patient-specific prior information to alleviate under-sampling challenges and meet the temporal constraint (<500 ms), the prior information can be outdated and introduce biases, thus compromising the imaging and motion tracking accuracy. To address this challenge, we developed a framework dynamic reconstruction and motion estimation (DREME) for real-time CBCT imaging and motion estimation, without relying on patient-specific prior knowledge. Approach. DREME incorporates a deep learning-based real-time CBCT imaging and motion estimation method into a dynamic CBCT reconstruction framework. The reconstruction framework reconstructs a dynamic sequence of CBCTs in a data-driven manner from a standard pre-treatment scan, without requiring patient-specific prior knowledge. Meanwhile, a convolutional neural network-based motion encoder is jointly trained during the reconstruction to learn motion-related features relevant for real-time motion estimation, based on a single arbitrarily-angled x-ray projection. DREME was tested on digital phantom simulations and real patient studies. Main Results. DREME accurately solved 3D respiration-induced anatomical motion in real time (∼1.5 ms inference time for each x-ray projection). For the digital phantom studies, it achieved an average lung tumor center-of-mass localization error of 1.2 ± 0.9 mm (Mean ± SD). For the patient studies, it achieved a real-time tumor localization accuracy of 1.6 ± 1.6 mm in the projection domain. Significance. DREME achieves CBCT and volumetric motion estimation in real time from a single x-ray projection at arbitrary angles, paving the way for future clinical applications in intra-fractional motion management. In addition, it can be used for dose tracking and treatment assessment, when combined with real-time dose calculation.
Title: Real-time CBCT imaging and motion tracking via a single arbitrarily-angled x-ray projection by a joint dynamic reconstruction and motion estimation (DREME) framework
Description:
Abstract
Objective.
Real-time cone-beam computed tomography (CBCT) provides instantaneous visualization of patient anatomy for image guidance, motion tracking, and online treatment adaptation in radiotherapy.
While many real-time imaging and motion tracking methods leveraged patient-specific prior information to alleviate under-sampling challenges and meet the temporal constraint (<500 ms), the prior information can be outdated and introduce biases, thus compromising the imaging and motion tracking accuracy.
To address this challenge, we developed a framework dynamic reconstruction and motion estimation (DREME) for real-time CBCT imaging and motion estimation, without relying on patient-specific prior knowledge.
Approach.
DREME incorporates a deep learning-based real-time CBCT imaging and motion estimation method into a dynamic CBCT reconstruction framework.
The reconstruction framework reconstructs a dynamic sequence of CBCTs in a data-driven manner from a standard pre-treatment scan, without requiring patient-specific prior knowledge.
Meanwhile, a convolutional neural network-based motion encoder is jointly trained during the reconstruction to learn motion-related features relevant for real-time motion estimation, based on a single arbitrarily-angled x-ray projection.
DREME was tested on digital phantom simulations and real patient studies.
Main Results.
DREME accurately solved 3D respiration-induced anatomical motion in real time (∼1.
5 ms inference time for each x-ray projection).
For the digital phantom studies, it achieved an average lung tumor center-of-mass localization error of 1.
2 ± 0.
9 mm (Mean ± SD).
For the patient studies, it achieved a real-time tumor localization accuracy of 1.
6 ± 1.
6 mm in the projection domain.
Significance.
DREME achieves CBCT and volumetric motion estimation in real time from a single x-ray projection at arbitrary angles, paving the way for future clinical applications in intra-fractional motion management.
In addition, it can be used for dose tracking and treatment assessment, when combined with real-time dose calculation.
Related Results
Dynamic cone-beam CT reconstruction using spatial and temporal implicit neural representation learning (STINR)
Dynamic cone-beam CT reconstruction using spatial and temporal implicit neural representation learning (STINR)
Abstract
Objective
. Dynamic cone-beam CT (CBCT) imaging is highly desired in image-guided radiation therapy to provide v...
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Comparison between hybrid CBCT and CBCT in detection of vertical root fracture
Cone beam computerized tomography (CBCT) has been wildly used in many fields of dentistry, especially to investigate the vertical root fracture (VRF). CBCT systems were consisted o...
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging in relation to the mandibular third molar: tooth characteristics, modality choice, optimization, and absorbed dose
Radiographic imaging of the mandibular third molar (M3) is a valuable diagnostic tool. It provides information on tooth position, root morphology, and relations to surrounding anat...
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Barriers Toward Better CBCT Knowledge, Attitude, and Practice
Objective: This study attempts to evaluate the knowledge, attitude, and clinical practice of Iraqi dentists toward CBCT use, referral patterns, observed challenges, and the impacts...
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Evaluation of a novel CBCT conversion method implemented in a treatment planning system
Abstract
Background
To evaluate a novel CBCT conversion algorithm for dose calculation implemented in a research version ...
Abstract 014: Beveled‐Tip Aspiration Catheters Reduce Stoke Procedure Time And Cost For Large And Medium Vessel Occlusions
Abstract 014: Beveled‐Tip Aspiration Catheters Reduce Stoke Procedure Time And Cost For Large And Medium Vessel Occlusions
Introduction
Thrombectomy catheters with a unique angled‐tip design (ZoomTM Aspiration Catheters, Imperative Care Inc., Campbell, CA) have recently emerged as promising...
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Comparing Unet training with three different datasets to correct CBCT images for prostate radiotherapy dose calculations
Abstract
Image intensity correction is crucial to enable cone beam computed tomography (CBCT) based radiotherapy dose calculations. This study evaluated three dif...
Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy
Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy
Introduction
This retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy.
...